Selenia Full Field Digital Mammography System with Software… recallZ-1546-2008 · Lorad, A Hologic, Inc. · initiated Feb 28, 2008

Terminated All recalls posted in May 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Lorad, A Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.

FDA root cause category: Software design

What the company is doing

Consignees were notified by an Urgent Recall Notice sent on 3/21/08. The letter informed users that representatives were dispatched to all locations and they have corrected the magnification factor problem. The letter instructed users to re-evaluate any cases read using CAD from the period of installation until the date of system correction. The letter also instructed users that only images read with CAD markers are affected. The letter requests users to return a confirmation of notification. For additional information, contact 1-877-371-4372.

Recall details

Recall number
Z-1546-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 28, 2008
Posted by the FDA
May 28, 2008
Recalling firm
Lorad, A Hologic, Inc.
Quantity
73 units
Product code
MUE
FDA submissions
P010025
Distribution
Worldwide Distribution including USA and countries of Canada and the Republic of Korea.

Product description

Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810

Code information (models, lots, serial numbers)

Serial Numbers: 28401060607, 28401060617, 28401060619, 28401060641, 28401060645, 28401071345, 28401071361, 28401071363, 28401082760, 28402050317, 28402071431, 28403060709, 28403071548, 28403071601, 28404060755, 28405060792, 28405060813, 28405071807, 28405071813, 28406071833, 28407050431, 28407060899, 28407071951, 28407072014, 28408050465, 28408072100, 28408072106, 28408072130, 28409030075, 28409030076, 28409050497, 28410072360, 28411061182, 28411061271, 28411072479, 28412050577, 28412050589, 28412061309, 28412061319, 28412061333, 28412061336, 28412061339, 28412072602, 29408072073, 28407072007S, 28409030072RF and 28409030077RM.

Other devices under the same FDA submission

5 devices cleared under P010025 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.