LATITUDE Patient Management System, Models 6441*, 6442*… recallZ-0333-2012 · Boston Scientific CRM Corp · initiated Oct 12, 2011
Terminated All recalls posted in December 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Current Boston Scientific defibrillator systems measure shock lead impedances up to 200 ohms, and export results to the programmer and LATITUDE monitoring system via the Daily Measurements feature. A measured shock lead impedance that is greater than 125 ohms or less than 20 ohms will generate an alert message on programmer screens and a Red Alert within LATITUDE (if activated). However, the mea
FDA root cause category: Software design
What the company is doing
Boston Scientific sent an "IMPORTANT DIAGNOSTIC INFORMATION" letter dated October 2011 to all affected customers. The letter described the product, problem and actions to be taken by the customers.
Recall details
- Recall number
- Z-0333-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 12, 2011
- Posted by the FDA
- Dec 1, 2011
- Recalling firm
- Boston Scientific CRM Corp
- Quantity
- 25,300 (14,700 US; 10,600 OUS)
- Product code
- LWS
- FDA submissions
- P910077
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, NETHERLANDS, NORWAY, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM, CANADA, AUSTRALIA, NEW ZEALAND, CYPRUS, CZECH REPUBLIC, GREECE, GUADELOUPE, HUNGARY, BANGLADESH, CHINA, COCOS ISLAND, HONG KONG, INDIA, INDONESIA, KOREA, MALAYSIA, PHILIPPINES, SINGAPORE, SOUTH KOREA, SRI LANKA, THAILAND, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SOUTH AFRICA, NEW CALEDONIA, POLAND, PORTUGAL, QATAR, REUNION,IRAN, ISRAEL, JORDAN, KUWAIT, LATVIA, LEBANON, LUXEMBOURG, MONACO, MOROCCO, JAPAN, ARGENTINA, ARUBA, BAHAMAS, BARBADOS, BELIZE, BERMUDA, BRAZIL, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, DOMINICAN REPUBLIC, ECUADOR, EL SALVADOR, FALKLAND ISLANDS, HAITI, HONDURAS, JAMAICA, MEXICO, PANAMA, PERU, TRINIDAD & Tobago, URUGUAY, VENEZUELA, AND NETHERLANDS ANTILLES.
Product description
LATITUDE Patient Management System, Models 6441*, 6442*, 6465*, 6488 (USA). There are two software applications: the programmer software and the LATITUDE remote monitoring system software. The Model 2868 Software Application is loaded onto the Model 3120 programmers. These programmers are used during in-office follow-ups. The software version affected domestically is version 1.05 for COGNIS and TELIGEN and internationally is version 2.04 for COGNIS, TELIGEN, and PUNCTUA, ENERGEN, and INCEPTA. The PUNCTUA, ENERGEN, and INCEPTA are not currently approved in the US The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Guidant or Boston Scientific pulse generator and transfer data to a central database. The LATITUDE Patient Management System software is web-based and therefore at any point in time, there is only one (1) released version for each model number. The current version is 7.1. Both the programmer software and the LATITUDE remote monitoring system software similarly display the shock lead impedance measurements.
Code information (models, lots, serial numbers)
Models 6441*, 6442*, 6465*, 6488 (USA).
Other devices under the same FDA submission
34 devices cleared under P910077 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 death65 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls9 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls4 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 report
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 report
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- LATITUDE® Communicator 6290Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components142 reports
- LATITUDE™ Programming System 3300Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components141 reports
- LATITUDE™ Programming System 3300Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components72 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.