Medical device recalls: October 2016
The FDA posted 60 medical device recall records in October 2016, from 43 recall events by 38 companies. 10 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Zimmer Biomet, Inc. · 4 events, 5 records
- Teleflex Medical · 2 events, 3 records, 1 Class I
- Randox Laboratories Ltd. · 2 events, 2 records
- Synergetics Inc · 1 event, 4 records
- Toshiba American Medical Systems Inc · 1 event, 4 records
- Trumpf Medical Systems, Inc. · 1 event, 4 records
- St Jude Medical Inc. · 1 event, 3 records, 3 Class I
- Vascular Solutions, Inc. · 1 event, 3 records, 3 Class I
FDA root cause
- Nonconforming Material/Component · 10
- Device Design · 10
- Process control · 7
- Software design · 6
- Under Investigation by firm · 5
- Other · 5
- Mixed-up of materials/components · 4
- Use error · 2
Class I recalls (10)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0115-2017 · St Jude Medical Inc. · posted Oct 21, 2016 · Terminated
St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit.
- Z-0116-2017 · St Jude Medical Inc. · posted Oct 21, 2016 · Terminated
St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit.
- Z-0117-2017 · St Jude Medical Inc. · posted Oct 21, 2016 · Terminated
St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit.
- Z-0053-2017 · Vascular Solutions, Inc. · posted Oct 20, 2016 · Terminated
Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.
- Vascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. ↗Class IZ-0054-2017 · Vascular Solutions, Inc. · posted Oct 20, 2016 · Terminated
Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.
- Vascular Solutions Twin-Pass .023" Dual Access Catheter, model 5230. Sterilized with Ethylene Oxide.Class IZ-0055-2017 · Vascular Solutions, Inc. · posted Oct 20, 2016 · Terminated
Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.
- Z-0044-2017 · Teleflex Medical · posted Oct 18, 2016 · Terminated
The connector may disconnect from the tracheostomy tube during use.
- Z-0005-2017 · Heartware, Inc. · posted Oct 17, 2016 · Terminated
Loose connector ports in controller body.
- Z-0004-2017 · Leonhard Lang Medizintechnik GmbH · posted Oct 14, 2016 · Terminated
There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a…
- 0.2 Micron Filter, 50 mmClass IZ-0007-2017 · Baxter Corporation Englewood · posted Oct 6, 2016 · Terminated
Baxter Healthcare Corporation is issuing a voluntary product recall for all unexpired lots of the 50mm 0.2 Micron Filter (Product Code: H93835) due to the potential for the filter membrane layer to be missing and for particulate matter to be present.
Class II recalls (49)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0326-2017 · Hill-Rom, Inc. · posted Oct 31, 2016 · Terminated
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load…
- Z-0322-2017 · Medtronic Navigation, Inc. · posted Oct 28, 2016 · Terminated
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls…
- Z-0323-2017 · Medtronic Navigation, Inc. · posted Oct 28, 2016 · Terminated
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls…
- Z-0114-2017 · Sequent Medical Inc · posted Oct 27, 2016 · Terminated
Sequent Medical is recalling the VIA 27 Microcatheter because it may not adequately re-sheath the interventional device into the catheter as designed.
- Z-0315-2017 · Intuitive Surgical, Inc. · posted Oct 26, 2016 · Terminated
Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in unexpected master movement and potential instrument tip movement under certain circumstances
- Z-0318-2017 · Pentax of America Inc · posted Oct 26, 2016 · Terminated
PENTAX Medical is initiating this field action to provide customers that have purchased the affected endoscopes with the most recent Operation and Reprocessing IFUs.
- 4205 Journey Rolling Walker ↗Class IIZ-0304-2017 · Nova Ortho-Med Inc · posted Oct 24, 2016 · Terminated
Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.
- FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip (SPSA-20002590) ↗Class IIZ-0301-2017 · Lumenis Limited · posted Oct 23, 2016 · Terminated
The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a potential for cross-contamination.
- Z-0299-2017 · Keystone Dental Inc · posted Oct 21, 2016 · Terminated
Small diameter (SD) cover screw was packaged with the regular diameter implant and do not match the implant diameter (RD)
- RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L)Class IIZ-0290-2017 · RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) · posted Oct 20, 2016 · Terminated
RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on the rods. The small cracks on the rods used in implants…
- Z-0295-2017 · Ethicon Endo-Surgery Inc · posted Oct 20, 2016 · Terminated
Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing…
- Z-0296-2017 · Ethicon Endo-Surgery Inc · posted Oct 20, 2016 · Terminated
Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing…
- Z-0138-2017 · Olympus Corporation of the Americas · posted Oct 19, 2016 · Terminated
New warning requiring Users to conduct duodenoscope precleaning and manual cleaning even when using an Automated Endoscope Reprocessor (AER) that has instructions that may indicate a user could forego certain steps in precleaning and manual cleaning of the…
- Z-0289-2017 · US Endoscopy Group Inc · posted Oct 19, 2016 · Terminated
US Endoscopy is conducting a voluntary product recall of one lot of the Vari-Safe Injection Needle as it was identified that the lot was incorrectly distributed. The lot was assembled with internal catheter components which differ from the original…
- Z-0111-2017 · Zimmer Biomet, Inc. · posted Oct 18, 2016 · Terminated
Contents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes that could occur as a result of this…
- Z-0112-2017 · Stryker Instruments Div. of Stryker Corporation · posted Oct 18, 2016 · Terminated
Stryker Instruments is voluntarily recalling the Sagittal Blade 18.0X.97X90MM due to the potential for a sterile barrier breach in the product's packaging.
- Z-0119-2017 · Synergetics Inc · posted Oct 18, 2016 · Terminated
Vitrectomy cutters have been found not to be operating within the correct pressure pulse per specification.
- Z-0120-2017 · Synergetics Inc · posted Oct 18, 2016 · Terminated
Vitrectomy cutters have been found not to be operating within the correct pressure pulse per specification.
- Z-0121-2017 · Synergetics Inc · posted Oct 18, 2016 · Terminated
Vitrectomy cutters have been found not to be operating within the correct pressure pulse per specification.
- Z-0122-2017 · Synergetics Inc · posted Oct 18, 2016 · Terminated
Vitrectomy cutters have been found not to be operating within the correct pressure pulse per specification.
- Randox Lipase (Colorimetric) ↗Class IIZ-0123-2017 · Randox Laboratories Ltd. · posted Oct 18, 2016 · Terminated
According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place…
- Z-0088-2017 · SenTec AG · posted Oct 17, 2016 · Terminated
Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical…
- Z-0093-2017 · Tearscience, Inc · posted Oct 17, 2016 · Terminated
Complaints related to an error condition occurring during patient treatment where the FipiFlow System automatically stopped treatment after detecting a problem with the Activator.
- Z-0083-2017 · Zimmer Biomet, Inc. · posted Oct 14, 2016 · Terminated
Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new…
- Phoenix Recon Targeting Arm ↗Class IIZ-0084-2017 · Zimmer Biomet, Inc. · posted Oct 14, 2016 · Terminated
Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new…
- Z-0087-2017 · Zimmer Biomet, Inc. · posted Oct 14, 2016 · Terminated
The G-PIN 2.8BL 3.2PL 60 CM 2PK is intended to contain one (1) 467265-01 Ball Tip Guide Pin and one (1) 467255-02 Plain Tip Guide Pin. The G-PIN 2.8BL 3.2PL 60 CM 2PK may have been packaged incorrectly either with two (2) 467265-01 Ball Tip Guide Pins and…
- Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). ↗Class IIZ-0077-2017 · Varian Medical Systems, Inc. · posted Oct 13, 2016 · Terminated
Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on the metal head ring. There was no report of serious injury due to this issue
- Z-0080-2017 · Contour Fabricators Inc · posted Oct 13, 2016 · Terminated
Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no…
- Z-0067-2017 · Brathwaites Oliver Medical · posted Oct 11, 2016 · Terminated
The Introes Pocket Bougie an endotracheal tube introducer size 14 Fr (4.7mm) x 60 cm may break during intubation.
- Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 ↗Class IIZ-0070-2017 · Medtronic Navigation, Inc.-Littleton · posted Oct 11, 2016 · Terminated
O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images
- Z-0071-2017 · GE OEC Medical Systems, Inc · posted Oct 11, 2016 · Terminated
GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm moving up or down…
- Z-0051-2017 · Argon Medical Devices, Inc · posted Oct 7, 2016 · Terminated
The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath
- Z-0056-2017 · Natus Medical Incorporated · posted Oct 7, 2016 · Terminated
NatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several complaints of wheels bending out of alignment.
- Z-0058-2017 · Teleflex Medical · posted Oct 7, 2016 · Terminated
The product is being recalled due to a labeling error. The Gripper Graspers are labeled as Johans Graspers and the Johans Graspers are labeled as Gripper Graspers.
- Z-0059-2017 · Teleflex Medical · posted Oct 7, 2016 · Terminated
The product is being recalled due to a labeling error. The Gripper Graspers are labeled as Johans Graspers and the Johans Graspers are labeled as Gripper Graspers.
- Z-0060-2017 · Stryker Howmedica Osteonics Corp. · posted Oct 7, 2016 · Terminated
Stryker Orthopaedics has received reports of the thread length protruding past the dome of the acetabular trial or implant for the Trident Universal Impactor/Positioner.
- Z-0046-2017 · Toshiba American Medical Systems Inc · posted Oct 6, 2016 · Terminated
It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not…
- Z-0047-2017 · Toshiba American Medical Systems Inc · posted Oct 6, 2016 · Terminated
It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not…
- Z-0048-2017 · Toshiba American Medical Systems Inc · posted Oct 6, 2016 · Terminated
It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not…
- Z-0049-2017 · Toshiba American Medical Systems Inc · posted Oct 6, 2016 · Terminated
It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not…
- IgE Calibrator Series ↗Class IIZ-0050-2017 · Randox Laboratories Ltd. · posted Oct 6, 2016 · Terminated
According to the firm, the IgE Calibrator Series IE2492 (lot 387-392IE) value assignment has now been aligned to the 3rd International Standard for human serum IgE 11/234. Turbidimetric (Non-IFCC Cal.) Quality Control Targets and Ranges now apply. The…
- Z-0040-2017 · Trumpf Medical Systems, Inc. · posted Oct 5, 2016 · Terminated
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
- Z-0041-2017 · Trumpf Medical Systems, Inc. · posted Oct 5, 2016 · Terminated
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
- Z-0042-2017 · Trumpf Medical Systems, Inc. · posted Oct 5, 2016 · Terminated
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
- Z-0043-2017 · Trumpf Medical Systems, Inc. · posted Oct 5, 2016 · Terminated
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
- ROI-A Anterior Delivery Device. Orthopedic manual surgical instrument. Model number: SI-ROIA-0023 ↗Class IIZ-0045-2017 · LDR Spine USA, Inc. · posted Oct 5, 2016 · Terminated
The ROI-A Anterior Delivery Device T-Handle, part number SI-ROIA-0023 has experienced binding of the rotation of the t-handle. There have been no reports of patient injury
- Z-0039-2017 · Siemens Healthcare Diagnostics, Inc. · posted Oct 4, 2016 · Terminated
Siemens Healthcare Diagnostics determined that the Dimension Vista Calcium (CA) Flex reagent cartridge lot # 16060BB may produce erroneously low or high results from specific well sets.
- Z-0003-2017 · Zimmer Biomet, Inc. · posted Oct 3, 2016 · Terminated
The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be…
- Z-0008-2017 · Microgenics Corporation · posted Oct 3, 2016 · Terminated
A reagent stability issue (still under investigation) causes less separation in the calibration curve, which affects results across the entire range. For controls, maximum bias of +/- 35% for low levels, +/- 33% for mid-range levels, and +/- 47% for high…
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0057-2017 · Circulatory Technology Inc · posted Oct 7, 2016 · Terminated
Twisted or wrinkled balloon components which are considered a cosmetic defect by the firm.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.