HeartWare Ventricular Assist System Controller recallZ-0005-2017 · Heartware, Inc. · initiated Jun 8, 2016
Class I Terminated All recalls posted in October 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heartware, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Loose connector ports in controller body.
FDA root cause category: Device Design
What the company is doing
HeartWare sent an Urgent Medical Device Correction letter dated June 8, 2016, to all affected customers, HeartWare advised clinicians to remind patients to follow unit instructions and water avoidance, and to regularly schedule appointments to check the integrity of the unit and to be cautious with connectors when inspecting connectors. Clinicians were also instructed to send back the enclosed "Acknowledgement Form" within 30 days and to forward the message to necessary individuals within their organization. Customers with questions were instructed to contact their local HeartWare representative. For questions regarding this recall call 305-364-1575.
Recall details
- Recall number
- Z-0005-2017
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jun 8, 2016
- Posted by the FDA
- Oct 17, 2016
- Recalling firm
- Heartware, Inc.
- Quantity
- 4235 Worldwide, 7272 units in U.S.
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI and countries of Australia, Austria, Belgium, Canada, France, Germany, Lebanon, Malaysia, Sweden, Switzerland, & United Kingdom.
Product description
HeartWare Ventricular Assist System Controller Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HVAD is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter
Code information (models, lots, serial numbers)
All Heartware Controllers manufactured after September 2014. US Model No: 1400, 1401XX, 1403US. International Model No: 1400, 1401XX, 1407XX.
Other devices under the same FDA submission
33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page