TRUEtrack TRUEtrack Blood Glucose Meter - Free Meter Only (REF A4088-85)Recalls & reported problems
Trividia Health, Inc. · HealthWise Pharmacy TRUEtrack Blood Glucose Meter - Free Meter Only · Class 2
All models of TRUEtrack (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00021292009281), received between Aug 21, 2017 and Apr 10, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K032657); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Low Test Results2
- High Test Results1
Patient problems
- No Known Impact Or Consequence To Patient2
- Dizziness1
- Headache1
- Shaking/Tremors1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2018Low Test Results; Low Test Results
CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM FASTING BACK TO BACK BLOOD TESTS PERFORMED DURING THE CALL OF 58 AND 88 MG/DL USING TRUETRACK METER. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 - 100 MG/DL. AT THE TIME OF THE CALL ON (B)(6) 2018, THE CUSTOMER REPORTED SYMPTOMS OF HEADACHE, DIZZY AND SWEATY; MEDICAL ATTENTION WAS NOT NEEDED AT TIME OF CALL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING ROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 05/21/2020 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE/TIME NOT SET; UNABLE TO VERIFY MORE THAN TWO MEMORY RESULTS): RESULT 1 :69 MG/DL DATE: (B)(6) 2018 TIME: 11:55 AM FASTING. RESULT 2 :88 MG/DL DATE: (B)(6) 2018 TIME: 02:25 PM FASTING. CUSTOMER CALLED STATING HER BLOOD SUGAR IS READING LOWER THAN NORMAL ON THIS METER. CUSTOMERS REGULAR FASTING RANGE IS 80-100 MG/DL. CUSTOMER STATES HER METER READ 69 MG/DL FASTING. CUSTOMER IS FEELING SYMPTOMS OF LOW BLOOD SUGAR. UNABLE TO VERIFY MORE THAN 2 RESULTS ON THE METERS MEMORY. WHILE DOING THE BACK TO BACK TEST, THE
Patient problems: Dizziness; Headache; Shaking/Tremors
MAUDE report 7412436 ↗MalfunctionAug 24, 2017High Test Results; High Test Results
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 302, 416, 312 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 110 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6)2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER; CUSTOMER STATED SHE DOES NOT HAVE ANY STRIPS LEFT. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/17/2018 AND OPEN VIAL DATE WAS UNDISCLOSED. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: A 302MG/DL (B)(6)/2017 10:17 AM FASTING:YES. A 416MG/DL (B)(6)2017 02:53 PM FASTING:YES. A 312MG/DL (B)(6)2017 01:42 PM FASTING:YES. A 348MG/DL (B)(6)2017 11:13 AM FASTING:YES. A 214MG/DL (B)(6)2017 12:11 PM FASTING:YES. MEMORY CONCERNS:CUSTOMER IS CONCERNED WITH THE RESULTS OF 302, 416, 312, 348 AND 214 MG/DL IN THE METERS MEMORY. CUSTOMER STORES STRIPS IN BEDROOM HOWEVER CUSTOMER ADVISED CCR THAT SHE DOES NOT HAVE ANY STRIPS LEFT AND SHE CANNOT AFFORD TO
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6816353 ↗MalfunctionAug 21, 2017Low Test Results; Low Test Results
CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. SISTER IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH ALL TESTS RESULTS FROM THE METER MEMORY. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 135 - 175 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 140 MG/DL AND 159 MG/DL USING TRUETRACK METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/28/2018 AND OPEN VIAL DATE AT TIME OF CALL IS TWO MONTHS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 77MG/DL, (B)(6) 2017, 09:39 PM, FASTING: YES; 73MG/DL, (B)(6) 2017, 09:30 PM, FASTING: YES; 30MG/DL, (B)(6) 2017, 09:12 PM, FASTING: YES; 81MG/DL, (B)(6) 2017, 10:30 AM, FASTING: YES; 68MG/DL, (B)(6) 2017, 10:24 AM, FASTING: YES. MEMORY CONCERNS: CUSTOMER IS CONCERN WITH ALL THESE RESULTS. (B)(6) (SISTER) CALLING STATES THAT THE METER IS NOT GIVING CORRECT RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6807009 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032657. One submission often covers many models, so check the product and lot details in each record.
- Z-2881-2016 · Jun 28, 2016 · Open, ClassifiedTRUEread Blood Glucose Test Strips (Glucose Oxidase) are used with TRUEresult and TRUE2go meters. TRUEread Blood Glucose Test Strips are available in vials with 10 ct., 25 ct., 50
- Z-0910-2014 · Oct 2, 2013 · TerminatedMooreMedical Supporting Health & Care...MooreBrand Blood Glucose Monitor for Self-Testing & Multi-Patient Use TRUEtrack. Manufactured by: NIPRO DIAGNOSTICS, Fort Lauderdale, FL 333
- Z-1140-2013 · Feb 25, 2013 · TerminatedDuane Reade TRUEtrack monitor kit, UPC# 6 39194 01679 7 Part # A4006-81 The TRUEtrack Blood Glucose System is intended for the quantitative determination of glucose in human whole
Other Glucose Oxidase, Glucose devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 022-895LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall7 reports
- OneTouch Ultra Blue 020-994LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall2 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Cholestech LDX™ 97991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page