Skip to content

Posey 8309WLRecalls & reported problems

Posey Products, LLC · SITTER ON CUE PRO WIRELESS CHAIR PAD, TAP-2-PAIR · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00190676007653), received between Nov 12, 2024 and Jun 9, 2026 — classified as 1 death, 6 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Audible Alarm6
  • Human-Device Interface Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • Ecchymosis1
  • Laceration(s)1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • DeathJun 9, 2026No Audible Alarm

    ON 5/13/2026, A COMPLAINT WAS REPORTED THAT A PATIENT WAS FOUND OUT OF A CHAIR AFTER AMBULATING TO THE BATHROOM AND HAD REMOVED A PICC LINE, RESULTING IN FLUID EXPOSURE. THE PATIENT WAS ALSO ATTEMPTING TO INSERT DENTURES. THE POSEY ALARM WAS ACTIVE (GREEN LIGHT ON) BUT DID NOT ALERT WHEN THE PATIENT EXITED THE CHAIR. STAFF CONFIRMED THE ALARM REMAINED ACTIVE DESPITE THE EVENT. THE PATIENT WAS LATER TRANSFERRED TO HOSPICE AND SUBSEQUENTLY PASSED AWAY.

    Patient problems: Insufficient Information

    MAUDE report 25443106 ↗
  • MalfunctionMay 26, 2026No Audible Alarm

    PATIENT WAS PLACED ON CHAIR ALARM WITH POSEY SENSOR PAD IN PLACE AS THE PATIENT WAS A HIGH FALL RISK AND THEN PATIENT SLID OUT OF THE CHAIR AND NOTHING ALARMED AND EVERYTHING WAS PLUGGED IN. NO PATIENT INJURY AND NO VISBALE SIGNS OF DAMAGED ON ALARM OR SENSOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25278734 ↗
  • MalfunctionMar 11, 2026Human-Device Interface Problem

    CUSTOMER REPORTED A PATIENT FALL WITH A BRUISE ON THE BACK OF THE HEAD. CUSTOMER ADVISED NO DAMAGE TO THE PRODUCTS, ISSUE RESULTED FROM HUMAN ERROR. RETURN DUE TO PATIENT INCIDENT. ITEM 8645WL, 8309WL.

    Patient problems: Ecchymosis

    MAUDE report 24572103 ↗
  • MalfunctionAug 20, 2025No Audible Alarm; No Audible Alarm

    CUSTOMER REPORTED, ACCORDING TO THE RN, THE PATIENT SAT ON THE CHAIR ALARM, HEARD THE BEEP, AND GOT UP FOR A LITTLE SO THE NURSE COULD ADJUST THE CHAIR PAD, AND SAT BACK DOWN. WHEN THE PATIENT GOT UP , THE CHAIR ALARM DID NOT ALARM. PATIENT INJURY, PATIENT SUSTAINED MINOR INJURY TO RIGHT KNEE. CHAIR SENSOR NOT AVAILABLE, AS THIS HAS BEEN DISPOSED OF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850093 ↗
  • MalfunctionJan 23, 2025No Audible Alarm; No Audible Alarm

    CUSTOMER REPORTING A COMPLAINT ON PRODUCT 8309WL LOT# 4304T148. CUSTOMER STATE THAT THEY RECENTLY HAD A PATIENT INCIDENT ON (B)(6) 2024. THE PATIENT WAS IN THEIR CHAIR WITH OUR 8309WL PAIRED TO THE ALARM. THE PATIENT GOT UP AND FELL. ALERTED STAFF BY YELLING. THE ALARM DID NOT SOUND WHEN THE PATIENT STOOD UP. A FOLEY WAS PULLED OUT AND A LACERATION OCCURED. STAFF RE-PAIRED THE SENSOR & ALARM AND IT STILL DIDN'T SOUND ALARM. STAFF TRIED A NEW SENSOR, THE NEW SENSOR WORKED CORRECTLY PAIRED/SOUNDED. CUSTOMER HAS ORIGINAL PAD STILL TO RETURN FOR EVALUATION.

    Patient problems: Laceration(s)

    MAUDE report 21223603 ↗
  • MalfunctionNov 21, 2024No Audible Alarm; No Audible Alarm

    CUSTOMER REPORTING A COMPLAINT ON PRODUCTS 8645WL & 8309WL. CUSTOMER IS REPORTING A FALL INCIDENT INVOLVING A PATIENT. CUSTOMER DID NOT SPECIFY IF PATIENT WAS INJURED AS A RESULT. CUSTOMER STATES THAT THE FALL TOOK PLACE DUE TO THE ALARM NOT SOUNDING. THE NURSE THAT FOUND THE THE PATIENT SAID THAT THE UNIT AND PAD NEEDED TO BE PAIRED AGAIN.

    Patient problems: Insufficient Information

    MAUDE report 20750034 ↗
  • MalfunctionNov 12, 2024No Audible Alarm; No Audible Alarm

    CUSTOMER REPORTED THAT THEY HAVE HAD PATIENT FALL INCIDENT DUE TO A NEW BATCH OF SENSOR PADS LOT: 4218T065. THEY HAD PURCHASED THESE ON (B)(6). THIS IS A NEW ISSUE FOR THEM. TM ALERTED TO THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20665769 ↗

Other Monitor, Bed Patient devices

All Monitor, Bed Patient devices →

Sources and updates