AVVIGO™+ H7492493120I0Recalls & reported problems
Boston Scientific Corporation · Multi-Modality Guidance System · Class 2
All models of AVVIGO+ (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00191506033163), received between Nov 26, 2024 and May 7, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Complete Loss of Power” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 5, 2024Class IIOpen, Classified
Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
Z-0100-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Complete Loss of Power1
- Connection Problem1
- Incorrect Measurement1
- Material Deformation1
- Nonstandard Device1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2026Incorrect Measurement; Output Problem
IT WAS REPORTED THAT INACCURATE MEASUREMENT OCCURRED. AN AVVIGO PLUS MULTI-MODALITY GUIDANCE SYSTEM WAS SELECTED. DURING THE PROCEDURE, THE MINIMAL LUMEN AREA (MLA) MEASUREMENTS WERE INACCURATE. IT WAS ALSO NOTED THAT THE FRACTIONAL FLOW RESERVE (FFR) AND DIASTOLIC HYPEREMIA-FREE RATIO (DFR) VALUES WERE ABLE TO EQUALIZE BUT DISAPPEARED DURING PULLBACK. THE PROCEDURE WAS COMPLETED AND THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25116545 ↗MalfunctionFeb 3, 2026Nonstandard Device; Complete Loss of Power
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. AN AVVIGO MULTI-MODALITY GUIDANCE SYSTEM WAS SELECTED FOR USE. DURING THE CASE THE SYSTEM WOULD NOT "GO LIVE". THE PROCEDURE WAS CANCELED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254519 ↗MalfunctionNov 20, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE TABLE SIDE CONTROLLER WAS SPARKING. OUTSIDE OF A PROCEDURE, IT WAS NOTED THAT THE AVVIGO PLUS SYSTEM TABLE SIDE CONTROLLER HAD VISIBLE SPARKS COMING OFF OF IT. THE HOSPITAL STAFF TURNED OFF POWER TO THE SYSTEM, AND THERE WAS NO INJURY CAUSED BY THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23611477 ↗MalfunctionNov 26, 2024Connection Problem; Connection Problem
IT WAS REPORTED THAT THE OPTICROSS CATHETER WAS MISRECOGNIZED. AN OPTICROSS 6HD AND AVVIGO PLUS SYSTEM WERE SELECTED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION. DURING THE PROCEDURE, THE OPTICROSS 6HD WAS BEING IDENTIFIED AS AN OPTICROSS 18 WHILE IT WAS PLUGGED IN TO THE AVVIGO PLUS MDU. THIS RESULTED IN THE IMAGES NOT BEING TO SCALE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20789547 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page