AVVIGO™+ H7492493120I0Recalls & reported problems

Boston Scientific Corporation · Multi-Modality Guidance System · Class 2

All models of AVVIGO+ (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00191506033163), received between Nov 26, 2024 and May 7, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 5, 2024Class IIOpen, Classified

    Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.

    Z-0100-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Complete Loss of Power1
  • Connection Problem1
  • Incorrect Measurement1
  • Material Deformation1
  • Nonstandard Device1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026Incorrect Measurement; Output Problem

    IT WAS REPORTED THAT INACCURATE MEASUREMENT OCCURRED. AN AVVIGO PLUS MULTI-MODALITY GUIDANCE SYSTEM WAS SELECTED. DURING THE PROCEDURE, THE MINIMAL LUMEN AREA (MLA) MEASUREMENTS WERE INACCURATE. IT WAS ALSO NOTED THAT THE FRACTIONAL FLOW RESERVE (FFR) AND DIASTOLIC HYPEREMIA-FREE RATIO (DFR) VALUES WERE ABLE TO EQUALIZE BUT DISAPPEARED DURING PULLBACK. THE PROCEDURE WAS COMPLETED AND THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25116545 ↗
  • MalfunctionFeb 3, 2026Nonstandard Device; Complete Loss of Power

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. AN AVVIGO MULTI-MODALITY GUIDANCE SYSTEM WAS SELECTED FOR USE. DURING THE CASE THE SYSTEM WOULD NOT "GO LIVE". THE PROCEDURE WAS CANCELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24254519 ↗
  • MalfunctionNov 20, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TABLE SIDE CONTROLLER WAS SPARKING. OUTSIDE OF A PROCEDURE, IT WAS NOTED THAT THE AVVIGO PLUS SYSTEM TABLE SIDE CONTROLLER HAD VISIBLE SPARKS COMING OFF OF IT. THE HOSPITAL STAFF TURNED OFF POWER TO THE SYSTEM, AND THERE WAS NO INJURY CAUSED BY THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23611477 ↗
  • MalfunctionNov 26, 2024Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE OPTICROSS CATHETER WAS MISRECOGNIZED. AN OPTICROSS 6HD AND AVVIGO PLUS SYSTEM WERE SELECTED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION. DURING THE PROCEDURE, THE OPTICROSS 6HD WAS BEING IDENTIFIED AS AN OPTICROSS 18 WHILE IT WAS PLUGGED IN TO THE AVVIGO PLUS MDU. THIS RESULTED IN THE IMAGES NOT BEING TO SCALE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20789547 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →