Skip to content

Change Healthcare Cardiology Hemo 15.1Recalls & reported problems

Change Healthcare Canada Company · Class 2

All models of Change Healthcare Cardiology Hemo (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
4

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 17540262100129), received between Jul 26, 2024 and Jul 26, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).

4 FDA recall records name this device's identifier, 4 still open.

8 other recalls cover devices cleared under the same FDA submission (K131497); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 22, 2025Class IIOpen, Classified

    Due to complaints, software update may cause software to unexpectedly shutdown.

    Z-2072-2025 · Root cause: Software design · Recall details · FDA record ↗

  2. Dec 2, 2024Class IIOpen, Classified

    Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s).

    Z-0958-2025 · Root cause: Software design · Recall details · FDA record ↗

  3. Dec 2, 2024Class IIOpen, Classified

    Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s).

    Z-0959-2025 · Root cause: Software design · Recall details · FDA record ↗

  4. Aug 21, 2024Class IIOpen, Classified

    A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.

    Z-0126-2025 · Root cause: Use error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Human-Device Interface Problem1
  • Inadequate User Interface1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 26, 2024Human-Device Interface Problem; Inadequate User Interface; Human-Device Interface Problem; Inadequate User Interface

    CHANGE HEALTHCARE CARDIOLOGY HEMODYNAMICS PROVIDES HEMODYNAMIC CALCULATIONS AND FORMULAS. THE SETTINGS, SUCH AS UNIT OF MEASUREMENT (UOM), FOR THESE FORMULAS AND CALCULATIONS ARE CONFIGURED IN THE CHANGE HEALTHCARE MANAGEMENT CONSOLE (CHMC) AND ARE DESIGNATED BY THE SITE ACCORDING TO FACILITY STANDARDS. IN THIS EVENT, THE UOM FOR THE HEMOGLOBIN VALUE, USED AS PART OF THE HEMODYNAMICS FORMULAS, WAS CONFIGURED IN THE CHMC AS MMOL/L (MILLIMOLES/LITER). HOWEVER, THE CLINICAL STAFF CONTINUED TO MANUALLY ENTER THE HEMOGLOBIN VALUE AS G/DL (GRAMS/DECILITER) WITHOUT BEING AWARE OF THE CHANGED CONFIGURATION FOR THE UOM. NO ADVERSE HEATH CONSEQUENCES RESULTING FROM THIS ISSUE WERE REPORTED TO CHANGE HEALTHCARE.

    Patient problems: Insufficient Information

    MAUDE report 19844346 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131497. One submission often covers many models, so check the product and lot details in each record.

  • Z-2071-2025 · May 22, 2025 · Open, Classified
    Change Healthcare Cardiology Hemo Software
  • Z-0957-2025 · Dec 2, 2024 · Open, Classified
    McKesson Cardiology Hemo software
  • Z-0128-2025 · Aug 21, 2024 · Open, Classified
    McKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
  • Z-1490-2019 · Mar 25, 2019 · Terminated
    McKesson Cardiology Hemo
  • Z-1489-2019 · Mar 25, 2019 · Terminated
    Horizon Cardiology Hemo
  • Z-2968-2018 · Mar 12, 2018 · Terminated
    McKesson Cardiology Hemo, Release SW version 13.0. Used for physiological monitoring, image and data processing. McKesson Cardiology Hemo is intended for complete physiological/h
  • Z-0042-2015 · Aug 27, 2014 · Terminated
    McKesson Cardiology" Hemo intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
  • Z-0093-2015 · Jun 25, 2014 · Terminated
    McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, an

Other Computer, Diagnostic, Programmable devices

All Computer, Diagnostic, Programmable devices →