3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 17540262100105), received between Jan 8, 2026 and Jan 8, 2026 — classified as 1 death.
The most frequently reported specific device problem is “Blocked Connection” (1 report).
3 FDA recall records name this device's identifier, 3 still open.
7 other recalls cover devices cleared under the same FDA submission (K131497); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 22, 2025Class IIOpen, Classified
Due to complaints, software update may cause software to unexpectedly shutdown.
Z-2071-2025 · Root cause: Software design · Recall details · FDA record ↗
- Dec 2, 2024Class IIOpen, Classified
Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s).
Z-0958-2025 · Root cause: Software design · Recall details · FDA record ↗
- Aug 21, 2024Class IIOpen, Classified
A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
Z-0126-2025 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Blocked Connection1
Patient problems
- Myocardial Infarction1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathJan 8, 2026Blocked Connection
CHANGE HEALTHCARE (CHC) IS FILING THIS REPORT BECAUSE A PATIENT DEATH OCCURRED AT A CUSTOMER FACILITY; CHC'S INVESTIGATION, CONSISTENT WITH THE TREATING CLINICIAN'S ACCOUNT, DETERMINED THAT THE CHC CARDIOLOGY HEMO SYSTEM DID NOT CAUSE OR CONTRIBUTE TO THE ADVERSE EVENT. EVENT SUMMARY: ON DEC 10TH 2025, DURING A SALES CALL WITH A U.S. CUSTOMER (B)(6), IT WAS REVEALED THAT AN ADVERSE EVENT OCCURRED ON (B)(6) WHILE THE HEMO SYSTEM WAS NOT RESPONDING. AT THE REQUEST OF CHC, THE CUSTOMER PROVIDED ADDITIONAL DETAILS AND THEIR INTERPRETATION OF THE CASE. THE PATIENT WAS A 75-YEAR-OLD WITH SEVERE CARDIAC HISTORY AND ARRIVED AT THE CUSTOMER FACILITY FOR A CORONARY INTERVENTION PROCEDURE. UPON ARRIVAL, THE PATIENT'S CONDITION DETERIORATED RAPIDLY. AN EMERGENCY HEART ATTACK WAS ACTIVATED; THE PATIENT WAS INTUBATED AND WAS ACTIVELY BEING RESUSCITATED. THE CUSTOMER INDICATED THAT THE PATIENT INFORMATION WAS NOT TRANSFERRING TO THE REAL TIME (RT) MONITOR, AND FOR APPROXIMATELY 40 MINUTES, VITAL SIGNS AND CARDIAC RHYTHM WERE UNAVAILABLE ON THE RT MONITOR. CHANGE HEALTHCARE CUSTOMER SUPPORT WAS CONTACTED, AND THE CHC REPRESENTATIVE WALKED THE CUSTOMER THROUGH REMOVING A FIREWALL THAT WAS DISRUPTIN
Patient problems: Myocardial Infarction
MAUDE report 24010403 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131497. One submission often covers many models, so check the product and lot details in each record.
- Z-0957-2025 · Dec 2, 2024 · Open, ClassifiedMcKesson Cardiology Hemo software
- Z-0128-2025 · Aug 21, 2024 · Open, ClassifiedMcKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
- Z-1490-2019 · Mar 25, 2019 · TerminatedMcKesson Cardiology Hemo
- Z-1489-2019 · Mar 25, 2019 · TerminatedHorizon Cardiology Hemo
- Z-2968-2018 · Mar 12, 2018 · TerminatedMcKesson Cardiology Hemo, Release SW version 13.0. Used for physiological monitoring, image and data processing. McKesson Cardiology Hemo is intended for complete physiological/h
- Z-0042-2015 · Aug 27, 2014 · TerminatedMcKesson Cardiology" Hemo intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
- Z-0093-2015 · Jun 25, 2014 · TerminatedMcKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, an
Other Computer, Diagnostic, Programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
- CARESCAPE CENTRAL STATION MAI800Ge Medical Systems Information Technologies, Inc. · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page