OPAL HDx™ M004OPLINSTALL2EN0Recalls & reported problems
Boston Scientific Corporation · Mapping System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00191506055370), received between Nov 11, 2025 and May 12, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Misassembled During Manufacturing /Shipping1
- Failure to Read Input Signal1
- Signal Artifact/Noise1
- Wrong Label1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 12, 2026Signal Artifact/Noise
THE PROCEDURE WAS CANCELLED. IT WAS REPORTED THAT DURING A PROCEDURE, A OPAL HDX WAS SELECTED FOR USE. THE SYSTEM WAS PRODUCING EXCESSIVE NOISE ON BOTH THE OPAL SYSTEM AND THE RECORDING SYSTEM DURING RF DELIVERY WHEN CONNECTED TO AN INTELLAGEN GENERATOR. THE NOISE WAS ONLY OBSERVED ON THE ECG SIGNAL DURING RF DELIVERY. NO NOISE WAS REPORTED OUTSIDE RF APPLICATION. TWO GROUNDING PATCHES WERE UTILIZED DURING THE PROCEDURE. THERE WERE NO ERROR CODES OR ALARMS REPORTED ON THE OPAL SYSTEM OR THE INTELLAGEN GENERATOR. EPPS RECOMMENDED EVALUATION OF ALL SYSTEM CONNECTIONS AND GROUNDING PATCHES AS PART OF TROUBLESHOOTING. THE PHYSICIAN DECIDED, PARTIALLY BASED ON THE ISSUE WITH NOISE ON THE ECG LEADS WHILE ABLATING TO END THE PROCEDURE EARLY WITHOUT KNOWING IF IT WAS SUCCESSFUL. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25150141 ↗MalfunctionNov 18, 2025Device Misassembled During Manufacturing /Shipping; Wrong Label; Device Misassembled During Manufacturing /Shipping; Wrong Label
IT WAS REPORTED THAT THE ISOLATION TRANSFORMER PACKAGED WITH THE OPAL HDX MAPPING SYSTEM INSTALL BUNDLE WAS MISLABELED AS THE UNITED STATES MODEL BUT WAS ACTUALLY THE EUROPEAN MODEL. THIS WAS AN OUT-OF-BOX FAILURE, SO NO PATIENT WAS INVOLVED. A REPLACEMENT TRANSFORMER WAS SHIPPED TO THE CUSTOMER AND INSTALLED. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23581092 ↗MalfunctionNov 11, 2025Failure to Read Input Signal; Failure to Read Input Signal
IT WAS REPORTED THAT DURING A PROCEDURE AN OPAL HDX MAPPING SYS INSTALL BUNDLE WAS SELECTED FOR USE. HOWEVER, A BADHDR ERROR OCCURRED. THE PHYSICIAN DECIDED TO TERMINATE THE THERAPY EARLY SINCE NO MAPPING NAVIGATION COULD BE PROVIDED FOR THE REQUESTED OPAL CASE. NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS EXPECTED TO BE RETURNED FOR LABORATORY ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23519485 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page