Alaris 8015 (UDI 10885403801549)Recalls & reported problems
Carefusion 303, Inc. · Alaris PC Unit · Class 2
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Alaris 8015 PC unit has 13,014 adverse event reports, including 6 deaths and 32 injuries. Many describe signal reading failures, corroded or broken parts, and early battery drain, and one injury report describes a heparin infusion running faster than planned. There are 3 open recalls, covering a bent connector, unvalidated syringe labeling, and network connection failures. Reports fell from 1,314 in the prior 12 months to 297 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 13,022
- Deaths reported
- 6
- Injuries reported
- 32
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 13,022 adverse event reports that list this exact device (DI 10885403801549), received between Mar 2, 2022 and Sep 23, 2026 — classified as 6 deaths, 32 injuries, 12,984 malfunctions.
The most frequently reported specific device problem is “Failure to Read Input Signal” (5,126 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Aug 15, 2024Open, Classified
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Z-3138-2024 · Root cause: Process control · FDA record ↗
- Sep 15, 2023Open, Classified
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Z-0186-2024 · Root cause: Under Investigation by firm · FDA record ↗
- Jan 24, 2023Open, Classified
Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed
Z-1192-2023 · Root cause: Software Design Change · FDA record ↗
Most reported problems
Device problems
- Failure to Read Input Signal5,126
- Corroded3,815
- Break2,477
- Contamination601
- Premature Discharge of Battery511
- Physical Resistance/Sticking276
- Crack252
- Electrical /Electronic Property Problem225
Patient problems
- No Clinical Signs, Symptoms or Conditions12,918
- Insufficient Information75
- Low Blood Pressure/ Hypotension21
- High Blood Pressure/ Hypertension3
- Bradycardia2
- Hypoxia2
- Convulsion/Seizure1
- Cough1
Recent adverse event reports
The 25 most recent of 13,022 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Battery Problem
It was reported that the device had failed battery conditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870760 ↗MalfunctionSep 22, 2026Inaccurate Delivery
It was reported that the patient was receiving a continuous bivalirudin infusion that was ordered to run at 0.158 mg/kg/hr. The infusion was started at the correct rate 0.158 mg/kg/h (12.6 mL/hr) via interoperability. It was later discovered running at the incorrect rate of 0.1593 mg/kg/hr (12.7 mL/hr) for approximately two hours starting at 5:41AM. There was patient involvement and no harm.Additional information provided 02SEP2026: Per review of the event logs, it appears that the clinician inadvertently pressed the "Up Arrow" key prior to pressing start at 05:41:17. This increased the rate from 12.6 mL/hr to 12.7 mL/hr.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50850121 ↗MalfunctionSep 15, 2026Inadequate or Insufficient Training
It was reported that the 8015 Defective battery. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50768175 ↗MalfunctionSep 8, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50672945 ↗MalfunctionSep 3, 2026Device Alarm System
It was reported that the device had error 255-20-22 during battery reconditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50593002 ↗MalfunctionSep 3, 2026Device Alarm System
It was reported that the device had error 255-20-22 during battery reconditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50589599 ↗MalfunctionSep 3, 2026Device Alarm System
It was reported that the device had error 255-20-22 during battery reconditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50589594 ↗MalfunctionSep 2, 2026Device Alarm System
It was reported that the device had error code 133.6090. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50568103 ↗MalfunctionAug 19, 2026Battery Problem
Patient problems: Insufficient Information
MAUDE report 50315365 ↗MalfunctionAug 14, 2026Break; Contamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50251017 ↗MalfunctionAug 14, 2026Contamination; Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50251011 ↗InjuryAug 14, 2026Inaccurate Delivery
It was reported that there was an event where a patient was receiving a continuous intravenous heparin infusion experienced an unintended rapid infusion event. The heparin infusion was ordered and programmed to run at 1,150 units/hour in accordance with the established heparin protocol.A new heparin bag was initiated at 1214 hours following independent dose and rate verification by two registered nurses using the patient¿s most recent PTT result. The infusion pump was programmed to deliver heparin at 1,150 units/hour.During the scheduled infusion check at 1315 hours, nursing staff noted the heparin bag was nearly empty despite the prescribed infusion rate. The pump alarmed at this time, and the infusion was immediately stopped. A second nurse independently reviewed the pump programming and confirmed that the programmed settings were correct.The patient was assessed immediately following identification of the event. No signs or symptoms of bleeding were present, and vital signs remained stable. The responsible medical team, charge nurses, and unit manager were notified. The heparin infusion was held, and urgent coagulation studies were obtained. Protamine 100 mg IV was administered
Patient problems: Insufficient Information
MAUDE report 50249313 ↗MalfunctionAug 13, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50233234 ↗MalfunctionAug 10, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50192968 ↗MalfunctionAug 7, 2026Device Alarm System; Circuit Failure; Corroded; Application Program Problem; Computer Operating System Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50169534 ↗MalfunctionJul 30, 2026Battery Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50082969 ↗MalfunctionJul 29, 2026Device Alarm System
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50068946 ↗MalfunctionJul 29, 2026Device Alarm System; Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50067059 ↗MalfunctionJul 27, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50022072 ↗MalfunctionJul 27, 2026Premature Discharge of Battery
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50021934 ↗MalfunctionJul 6, 2026Break; Display or Visual Feedback Problem
IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25716023 ↗MalfunctionJul 2, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25689685 ↗MalfunctionJul 2, 2026Break; Contamination
IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED FOR THE REPORTED FAILURE AS IT WAS OBSERVED BY THE MANUFACTURER DURING SERVICING ACTIVITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25689240 ↗MalfunctionJul 2, 2026Battery Problem
IT WAS REPORTED THAT THE DEVICE HAD WILL NOT CHARGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25687836 ↗MalfunctionJul 1, 2026Contamination
IT WAS REPORTED THAT THE DEVICE HAD IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25676562 ↗
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
- CADD 21-6400-51Icu Medical, Inc. · Pump, infusion7 deaths7,543 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.