Alaris 8015 (UDI 10885403801549)Recalls & reported problems
Carefusion 303, Inc. · Alaris PC Unit · Class 2
All models of BD Alaris Infusion System (29) · Alaris (25)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Alaris 8015 PC unit has 13,014 adverse event reports, including 6 deaths and 32 injuries. Many describe signal reading failures, corroded or broken parts, and early battery drain, and one injury report describes a heparin infusion running faster than planned. There are 3 open recalls, covering a bent connector, unvalidated syringe labeling, and network connection failures. Reports fell from 1,314 in the prior 12 months to 297 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 13,013
- Deaths reported
- 6
- Injuries reported
- 32
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 13,013 adverse event reports that list this exact device (DI 10885403801549), received between Mar 2, 2022 and Sep 23, 2026 — classified as 6 deaths, 32 injuries, 12,975 malfunctions.
The most frequently reported specific device problem is “Failure to Read Input Signal” (5,126 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 15, 2024Class IIOpen, Classified
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Z-3138-2024 · Root cause: Process control · Recall details · FDA record ↗
- Sep 15, 2023Class IOpen, Classified
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Z-0186-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jan 24, 2023Class IIOpen, Classified
Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed
Z-1192-2023 · Root cause: Software Design Change · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Read Input Signal5,126
- Corroded3,810
- Break2,474
- Contamination599
- Premature Discharge of Battery509
- Physical Resistance/Sticking276
- Crack251
- Electrical /Electronic Property Problem225
Patient problems
- No Clinical Signs, Symptoms or Conditions12,912
- Insufficient Information72
- Low Blood Pressure/ Hypotension21
- High Blood Pressure/ Hypertension3
- Bradycardia2
- Hypoxia2
- Convulsion/Seizure1
- Cough1
Recent adverse event reports
The 25 most recent of 13,013 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Battery Problem
It was reported that the device had failed battery conditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870760 ↗MalfunctionSep 22, 2026Use of Device Problem; Inaccurate Delivery
It was reported that the patient was receiving a continuous bivalirudin infusion that was ordered to run at 0.158 mg/kg/hr. The infusion was started at the correct rate 0.158 mg/kg/h (12.6 mL/hr) via interoperability. It was later discovered running at the incorrect rate of 0.1593 mg/kg/hr (12.7 mL/hr) for approximately two hours starting at 5:41AM. There was patient involvement and no harm.Additional information provided 02SEP2026: Per review of the event logs, it appears that the clinician inadvertently pressed the "Up Arrow" key prior to pressing start at 05:41:17. This increased the rate from 12.6 mL/hr to 12.7 mL/hr.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50850121 ↗MalfunctionSep 15, 2026Inadequate or Insufficient Training
It was reported that the 8015 Defective battery. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50768175 ↗MalfunctionSep 8, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50672945 ↗MalfunctionSep 3, 2026Device Alarm System
It was reported that the device had error 255-20-22 during battery reconditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50593002 ↗MalfunctionSep 3, 2026Device Alarm System
It was reported that the device had error 255-20-22 during battery reconditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50589599 ↗MalfunctionSep 3, 2026Device Alarm System
It was reported that the device had error 255-20-22 during battery reconditioning. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50589594 ↗MalfunctionSep 2, 2026Device Alarm System
It was reported that the device had error code 133.6090. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50568103 ↗MalfunctionAug 19, 2026Battery Problem
Patient problems: Insufficient Information
MAUDE report 50315365 ↗MalfunctionAug 14, 2026Break; Contamination
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no patient involvement associated for the reported failure as it was observed by the manufacturer during servicing activity.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50251017 ↗MalfunctionAug 14, 2026Contamination; Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50251011 ↗InjuryAug 14, 2026Premature Discharge of Battery; Contamination; Corroded; Use of Device Problem; Inaccurate Delivery
It was reported that there was an event where a patient was receiving a continuous intravenous heparin infusion experienced an unintended rapid infusion event. The heparin infusion was ordered and programmed to run at 1,150 units/hour in accordance with the established heparin protocol.\\A new heparin bag was initiated at 1214 hours following independent dose and rate verification by two registered nurses using the patient¿s most recent PTT result. The infusion pump was programmed to deliver heparin at 1,150 units/hour.\\During the scheduled infusion check at 1315 hours, nursing staff noted the heparin bag was nearly empty despite the prescribed infusion rate. The pump alarmed at this time, and the infusion was immediately stopped. A second nurse independently reviewed the pump programming and confirmed that the programmed settings were correct.\\The patient was assessed immediately following identification of the event. No signs or symptoms of bleeding were present, and vital signs remained stable. The responsible medical team, charge nurses, and unit manager were notified. The heparin infusion was held, and urgent coagulation studies were obtained. Protamine 100 mg IV was adminis
Patient problems: Insufficient Information
MAUDE report 50249313 ↗MalfunctionAug 13, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50233234 ↗MalfunctionAug 10, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50192968 ↗MalfunctionAug 7, 2026Device Alarm System; Circuit Failure; Corroded; Application Program Problem; Computer Operating System Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50169534 ↗MalfunctionJul 30, 2026Battery Problem
It was reported that the device had PCU that is not charging and it showed a malfunction code 120.4610.0. The battery logs show it toggling from AC to DC power repeatedly in the same second as well. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50082969 ↗MalfunctionJul 29, 2026Device Alarm System
It was reported that the device displayed error code 255-202-2. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50068946 ↗MalfunctionJul 29, 2026Device Alarm System; Corroded
It was reported that the device had error code 110.6021 upon bootup. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50067059 ↗MalfunctionJul 27, 2026Electrical /Electronic Property Problem
It was reported that the device had It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50022072 ↗MalfunctionJul 27, 2026Premature Discharge of Battery
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50021934 ↗MalfunctionJul 6, 2026Break; Display or Visual Feedback Problem
IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25716023 ↗MalfunctionJul 2, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25689685 ↗MalfunctionJul 2, 2026Break; Contamination
IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED FOR THE REPORTED FAILURE AS IT WAS OBSERVED BY THE MANUFACTURER DURING SERVICING ACTIVITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25689240 ↗MalfunctionJul 2, 2026Battery Problem
IT WAS REPORTED THAT THE DEVICE HAD WILL NOT CHARGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25687836 ↗MalfunctionJul 1, 2026Contamination
IT WAS REPORTED THAT THE DEVICE HAD IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25676562 ↗
FDA clearance and classification
The FDA regulates product code FRN (pump, infusion) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Alaris 8015
Has the Alaris 8015 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-3138-2024 (Class II), began on Aug 15, 2024: Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Alaris 8015?
The FDA's MAUDE database lists 13,013 adverse event reports for this device, received between Mar 2, 2022 and Sep 23, 2026. In those reports, the most common specific device problems are “Failure to Read Input Signal” (5,126), “Corroded” (3,810) and “Break” (2,474); the most common patient problems are “Low Blood Pressure/ Hypotension” (21), “High Blood Pressure/ Hypertension” (3) and “Bradycardia” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alaris 8015?
6 reports are classified as deaths, 32 as injuries and 12,975 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Alaris 8015 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Alaris 8015, and is it still sold?
The FDA's device database lists Carefusion 303, Inc. as the company responsible for it (brand Alaris, model 8015). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,412 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,079 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths50,284 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,947 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,545 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,232 reports
- CADD 21-6400-51Icu Medical, Inc. · Pump, infusion7 deaths7,543 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page