BD® Alaris™ Pump Module, Model 8100Recalls & reported problems

Carefusion 303, Inc. · BD Alaris™ Pump Module, Model 8100 v12.1.2 · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a BD Alaris 8100 pump module. Its reports mostly describe sensing failures, breaks, cracks, corrosion and error codes, and many state no patient was involved. There are 49,742 reports, including 33 deaths and 144 injuries. Five recalls are open: hospital network connection problems, damage from drops that can cause over- or under-infusion, flow accuracy, a manual instruction error and bent connectors.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50,284
Deaths reported
30
Injuries reported
143
Recalls naming it
5

Summary

The FDA's MAUDE database contains 50,284 adverse event reports that list this exact device (DI 10885403810046), received between Aug 9, 2021 and Sep 30, 2026 — classified as 30 deaths, 143 injuries, 50,103 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (11,217 reports).

5 FDA recall records name this device's identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 584 reportsJan 2024: 742 reportsFeb 2024: 2,121 reportsMar 2024: 849 reportsApr 2024: 652 reportsMay 2024: 442 reportsJun 2024: 504 reportsJul 2024: 554 reportsAug 2024: 515 reportsSep 2024: 611 reportsOct 2024: 833 reportsNov 2024: 821 reportsDec 2024: 617 reportsJan 2025: 1,002 reportsFeb 2025: 942 reportsMar 2025: 1,047 reportsApr 2025: 889 reportsMay 2025: 1,267 reportsJun 2025: 1,382 reportsJul 2025: 1,623 reportsAug 2025: 1,114 reportsSep 2025: 1,140 reportsOct 2025: 1,633 reportsNov 2025: 1,488 reportsDec 2025: 1,381 reportsJan 2026: 1,383 reportsFeb 2026: 1,464 reportsMar 2026: 1,769 reportsApr 2026: 1,223 reportsMay 2026: 809 reportsJun 2026: 1,178 reportsJul 2026: 1,361 reportsAug 2026: 1,260 reportsSep 2026: 1,539 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262,121
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 6, 2025Class IIOpen, Classified

    Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network

    Z-1217-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Oct 17, 2025Class IOpen, Classified

    If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

    Z-0430-2026 · Root cause: Device Design · Recall details · FDA record ↗

  3. Jul 8, 2025Class IOpen, Classified

    Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

    Z-2179-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Jan 23, 2025Class IIOpen, Classified

    Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.

    Z-1231-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3142-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem11,217
  • Break8,763
  • Display Difficult to Read7,791
  • Corroded7,056
  • Failure to Sense6,140
  • Device Alarm System5,738
  • Crack5,722
  • Contamination4,818

Patient problems

  • No Clinical Signs, Symptoms or Conditions49,631
  • Insufficient Information645
  • Low Blood Pressure/ Hypotension57
  • High Blood Pressure/ Hypertension18
  • Hyperglycemia13
  • Hypoglycemia12
  • Bradycardia8
  • Tachycardia7

Recent adverse event reports

The 25 most recent of 50,284 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that the device had failed patient side occlusion. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950578 ↗
  • MalfunctionSep 30, 2026Contamination; Corroded; Use of Device Problem

    It was reported an over infusion of Magnesium. On (b)(6)2026 at 0925 Magnesium Sulfate 50ml was to infuse at a rate of 33.3ml/hr; however, at approximately 1045 the "entire bag was found to be empty". This was a one-time order for Magnesium replacement. No other medications were infusing at the time of the event. The Magnesium was programmed manually using guardrails library. This did not occur at shift change. There was patient involvement and no harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947222 ↗
  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946196 ↗
  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946192 ↗
  • MalfunctionSep 30, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946191 ↗
  • MalfunctionSep 30, 2026Failure to Sense

    It was reported that the device had error code 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946180 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Failure to Sense

    It was reported that the device had error code 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946152 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Break; Crack; Device Sensing Problem

    It was reported that the device displayed multiple error codes 241.4140, 242.4010 and 242.4030. Device needs new right IUI. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946150 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Failure to Sense

    It was reported that the device had error code 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945874 ↗
  • MalfunctionSep 30, 2026Failure to Sense

    It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945187 ↗
  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945186 ↗
  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that the device had Will not turn on. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945184 ↗
  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945179 ↗
  • MalfunctionSep 30, 2026Device Sensing Problem

    It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945178 ↗
  • MalfunctionSep 30, 2026Failure to Sense

    It was reported that the device displayed channel error code 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945166 ↗
  • MalfunctionSep 30, 2026Failure to Sense

    It was reported that the device had error code 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945163 ↗
  • MalfunctionSep 30, 2026Break; Failure to Sense

    It was reported that the device had air in line issue. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945151 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Crack

    It was reported that the device had channel error 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945150 ↗
  • MalfunctionSep 30, 2026Crack

    It was reported that the device had error message safety clamp open even when closed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945148 ↗
  • MalfunctionSep 30, 2026Circuit Failure

    It was reported that the device had Unit will not power on either side of the PCU. May be due to IUI issue. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945147 ↗
  • MalfunctionSep 30, 2026Break

    It was reported that the device had broken door latch. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945146 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Failure to Sense; Device Sensing Problem

    It was reported that the device displayed channel error 242.4030. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945145 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Failure to Sense

    It was reported that the device displayed channel error code 242.4030, when door was opened. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945143 ↗
  • MalfunctionSep 30, 2026Device Alarm System; Device Sensing Problem

    It was reported that the device had occlusion error at startup. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945142 ↗
  • MalfunctionSep 30, 2026Calibration Problem; Device Sensing Problem

    It was reported that the device had failed pressure transducer testing and rate calibration testing. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945141 ↗

FDA clearance and classification

The FDA regulates product code FRN (pump, infusion) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the BD® Alaris™ Pump Module, Model 8100

Has the BD® Alaris™ Pump Module, Model 8100 been recalled?

Yes. 5 FDA recall records name this device's identifier: 5 still open. The most recent, Z-1217-2026 (Class II), began on Nov 6, 2025: Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the BD® Alaris™ Pump Module, Model 8100?

The FDA's MAUDE database lists 50,284 adverse event reports for this device, received between Aug 9, 2021 and Sep 30, 2026. In those reports, the most common specific device problems are “Device Sensing Problem” (11,217), “Break” (8,763) and “Display Difficult to Read” (7,791); the most common patient problems are “Low Blood Pressure/ Hypotension” (57), “High Blood Pressure/ Hypertension” (18) and “Hyperglycemia” (13). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD® Alaris™ Pump Module, Model 8100?

30 reports are classified as deaths, 143 as injuries and 50,103 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BD® Alaris™ Pump Module, Model 8100 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the BD® Alaris™ Pump Module, Model 8100, and is it still sold?

The FDA's device database lists Carefusion 303, Inc. as the company responsible for it (brand BD® Alaris™ Pump Module, Model 8100, model 8100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Pump, infusion devices