Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00193982175191), received between May 19, 2026 and Aug 26, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Migration1

Patient problems

  • Nonunion/Delayed-union Bone Fracture2
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Fracture

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 50459946 ↗
  • MalfunctionMay 19, 2026Fracture; Migration

    IT WAS REPORTED THAT THE PATIENT HAD ORIGINAL SURGERY FOR A LEFT DISTAL FEMUR FRACTURE (B)(6) 2025. THE PATIENT WAS HAVING A REVISION SURGERY DONE DUE TO A NONUNION. THE DISTAL LOCKING SCREWS IN THE PLATE SHEARED OFF AND THE CANCELLOUS SCREW BACKED OUT. THERE WERE ALSO 4 LAG SCREWS IN THE PATIENT AND 2 OF THE WERE BROKEN AS WELL. THE SURGEON DID NOT REMOVE ANY OF THE LAG SCREWS AS HE DID NOT FEEL IT WOULD HAVE ANY ADVERSE EFFECT TO THE PATIENT. THE SURGEON TOOK OUT THE PLATE AND ALL THE SCREWS FOR THE PLATE, USED THE SYNTHES RIA AND THEN PUT IN ONE OF OUR 12X360 RETROGRADE NAIL AND LEFT RETROGRADE LOCKING WASHER PLATE. DURING TODAY'S SURGERY SOME SAMPLES WERE SENT TO PATHOLOGY AND ONE OF THE SAMPLES DID COME BACK WITH A NEUTROPHILS RANGE OF 5-10 WHICH INDICATES A POSSIBLE INFECTION. AFTER THE ORIGINAL SURGERY ON (B)(6) 2025, PATIENT DID NOT RETURN FOR PART 2 OF MASQUELET.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection

    MAUDE report 25221674 ↗

Other Plate, fixation, bone devices