Skip to content

Discovery MI (UDI 00195278098603)Recalls & reported problems

GE Medical Systems, L.L.C. · PET DMI Gen2 Image System Asm, No Detectors Installed, Common Weld Gantry, Manufacturing Make · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K161574, K211846); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jun 2, 2022Open, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K161574, K211846. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2019 · Feb 12, 2019 · Open, Classified
    Discovery Ml Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activi
  • Z-2580-2018 · May 24, 2018 · Terminated
    Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring
  • Z-2193-2017 · May 19, 2017 · Terminated
    GE Healthcare Discovery Ml X-ray system

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page