Skip to content

RT Refresh PackageRecalls & reported problems

GE Hangwei Medical Systems Company Ltd. · Designated medical device · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K132410); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Nov 5, 2024Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0613-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K132410 on Jan 17, 2014 (decision: Substantially Equivalent). Submitted by Ge Medical Systems, LLC as “Discovery Ct 590 Rt, Optima Ct 580, Optima Ct 580 Rt, Optima Ct 580 W”. The FDA review took 168 days. FDA record ↗

The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K132410. One submission often covers many models, so check the product and lot details in each record.

  • Z-2581-2018 · May 24, 2018 · Terminated
    Discovery RT labeled as: a. MID BJG; b. HVY BJG
  • Z-2184-2017 · May 19, 2017 · Terminated
    GE Healthcare Optima CT580; Discovery RT, CT590 RT X-ray system

Common questions about the RT Refresh Package

Has the RT Refresh Package been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0613-2025 (Class II), began on Nov 5, 2024: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the RT Refresh Package FDA approved?

It was cleared through 510(k) K132410 on Jan 17, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the RT Refresh Package, and is it still sold?

The FDA's device database lists GE Hangwei Medical Systems Company Ltd. as the company responsible for it (brand RT, model Refresh Package). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, X-Ray, Tomography, Computed devices

All System, X-Ray, Tomography, Computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page