Skip to content

ZENIX 8116-12-41-01300-00-USARecalls & reported problems

Zoll Medical Corporation · ZENIX, HOSPITAL MID LEVEL + VENTILATION, CONFIGURATION WITH MNC · Class 3

All models of ZENIX (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00198843000315), received between Apr 13, 2026 and Apr 16, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Grounding Malfunction” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 4 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Grounding Malfunction4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE FAILED ELECTRICAL SAFETY TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24899049 ↗
  • MalfunctionApr 15, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE FAILED ELECTRICAL SAFETY TEST FOR GROUND RESISTANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24884703 ↗
  • MalfunctionApr 15, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE FAILED ELECTRICAL SAFETY TEST FOR GROUND RESISTANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24884628 ↗
  • MalfunctionApr 13, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING INCOMING INSPECTION, THE DEVICE FAILED ELECTRICAL SAFETY TEST FOR GROUND RESISTANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24858326 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page