Skip to content

R SERIES 30001005201310012Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN,ALS,MAN-1,3/5 LD,STD ECG,AC PWR,SPO2,ETCO2,PACE,NIBP,CF/USB,R SERIES,DMST · Class 3

Quick summary

This Zoll R Series is a defibrillator with ECG and pacing features. Of its 1,711 reports, many describe the device failing to power up, showing no display, being unable to read the patient's heart signal, and failing to deliver shocks or pacing. The reports include 10 deaths and 25 injuries, and some mention burns. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,720
Deaths reported
10
Injuries reported
25
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,720 adverse event reports that list this exact device (DI 00847946017521), received between Mar 13, 2017 and Sep 30, 2026 — classified as 10 deaths, 25 injuries, 1,685 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (332 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 22 reportsDec 2023: 0 reportsJan 2024: 19 reportsFeb 2024: 21 reportsMar 2024: 23 reportsApr 2024: 16 reportsMay 2024: 21 reportsJun 2024: 25 reportsJul 2024: 18 reportsAug 2024: 11 reportsSep 2024: 24 reportsOct 2024: 18 reportsNov 2024: 13 reportsDec 2024: 15 reportsJan 2025: 22 reportsFeb 2025: 21 reportsMar 2025: 23 reportsApr 2025: 20 reportsMay 2025: 26 reportsJun 2025: 15 reportsJul 2025: 13 reportsAug 2025: 18 reportsSep 2025: 16 reportsOct 2025: 22 reportsNov 2025: 20 reportsDec 2025: 27 reportsJan 2026: 9 reportsFeb 2026: 19 reportsMar 2026: 29 reportsApr 2026: 31 reportsMay 2026: 24 reportsJun 2026: 9 reportsJul 2026: 15 reportsAug 2026: 17 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202631
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unable to Obtain Readings332
  • Failure to Power Up288
  • No Display/Image242
  • Defibrillation/Stimulation Problem225
  • Pacing Problem129
  • Failure to Deliver Shock/Stimulation125
  • Failure of Device to Self-Test97
  • Grounding Malfunction97

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,219
  • No Patient Involvement313
  • Insufficient Information93
  • No Information36
  • No Consequences Or Impact To Patient23
  • Burn(s)17
  • Death5
  • Necrosis4

Recent adverse event reports

The 25 most recent of 1,720 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Grounding Malfunction; Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951903 ↗
  • MalfunctionSep 29, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934520 ↗
  • MalfunctionSep 24, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB DISABLED", "DEFIB FAULT 72" AND "DEFIB FAULT 76" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883662 ↗
  • MalfunctionSep 23, 2026No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY SHOWED HORIZONTAL LINES AND WAS NOT LEGIBLE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872043 ↗
  • MalfunctionSep 23, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "ECG DISABLED" AND "ECG FAULT 6" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867959 ↗
  • MalfunctionSep 22, 2026Pacing Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "PACER FAULT 115" MESSAGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855448 ↗
  • MalfunctionSep 21, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS.COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841466 ↗
  • MalfunctionSep 18, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50828616 ↗
  • MalfunctionSep 18, 2026Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED TO DISCHARGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50823903 ↗
  • MalfunctionSep 18, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "ERROR FFFF" MESSAGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50817059 ↗
  • MalfunctionSep 15, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 50772311 ↗
  • MalfunctionSep 14, 2026Defibrillation/Stimulation Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAYED A "DEFIB DISABLED" AND "DEFIB FAULT 76" MESSAGES.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50753834 ↗
  • MalfunctionSep 11, 2026Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED TO DISCHARGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735596 ↗
  • MalfunctionSep 8, 2026Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669091 ↗
  • MalfunctionSep 4, 2026Unable to Obtain Readings

    Patient problems: Insufficient Information

    MAUDE report 50620146 ↗
  • MalfunctionSep 4, 2026Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ERROR FFFF" MESSAGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50612599 ↗
  • MalfunctionSep 4, 2026Unable to Obtain Readings; Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50612521 ↗
  • MalfunctionSep 3, 2026Pacing Problem; Defibrillation/Stimulation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50592128 ↗
  • MalfunctionSep 2, 2026Pacing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50567536 ↗
  • MalfunctionAug 31, 2026Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50520179 ↗
  • MalfunctionAug 26, 2026Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE FAILED TO DISCHARGE.COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT.COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50452494 ↗
  • MalfunctionAug 25, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50429499 ↗
  • MalfunctionAug 24, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50412481 ↗
  • MalfunctionAug 20, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50345725 ↗
  • MalfunctionAug 17, 2026Failure to Deliver Shock/Stimulation

    Patient problems: Insufficient Information

    MAUDE report 50286893 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.