LIFEPAK CR2 99512 (REF 99512-001261)Recalls & reported problems
Physio-Control, Inc. · CR2BAAAABFHBBBCDBCDBAAAALPCR2 CONFIGURED UNIT · Class 3
All models of LIFEPAK CR2 (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Physio-Control LIFEPAK CR2 is an automated external defibrillator that delivers shocks to restart the heart. The 1,502 reports most often describe a part coming loose or missing, such as the lid magnet, so the unit may not power on automatically when the lid is opened. Other reports describe the unit failing to power up, failing to deliver shocks, and having no audible prompts. The reports list 8 deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,502
- Deaths reported
- 8
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,502 adverse event reports that list this exact device (DI 00883873824139), received between Mar 10, 2020 and Apr 2, 2026 — classified as 8 deaths, 1 injury, 1,493 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1,232 reports).
2 other recalls cover devices cleared under the same FDA submission (P170018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1,232
- Failure to Power Up134
- Electrical /Electronic Property Problem33
- Audible Prompt/Feedback Problem22
- No Audible Prompt/Feedback22
- Failure to Sense21
- Component Missing19
- Failure to Deliver Shock/Stimulation13
Patient problems
- No Clinical Signs, Symptoms or Conditions1,487
- No Patient Involvement20
- Cardiac Arrest9
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Death1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 1,502 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 2, 2026Failure to Deliver Energy
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE DOES NOT DELIVER DEFIBRILLATION ENERGY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24766322 ↗MalfunctionDec 29, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE WAS UNABLE TO DELIVER A SHOCK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23916790 ↗MalfunctionSep 25, 2025Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON WHEN THE DEVICE LID WAS ACTIVATED. IN THIS STATE DEFIBRILLATION THERAPY MAY BE DELAYED IF NEEDED OR MAY LEAD TO USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23143129 ↗MalfunctionJul 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE¿S LID MAGNET WAS MISSING FROM THE LID. AS A RESULT, THE DEVICE MAY NOT AUTOMATICALLY POWER ON WHEN THE LID IS OPENED. THIS MAY LED TO DEFIBRILLATION THERAPY BEING DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22583438 ↗MalfunctionNov 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THERE WAS AN ISSUE WITH THEIR DEVICE'S LID MAGNET. THE LID MAGNET CONTROLS THE DEVICE AUTOMATICALLY POWERING ON WHEN THE LID IS OPENED. AS A RESULT, THE DEVICE MAY NOT AUTOMATICALLY POWER ON WHEN THE LID IS OPENED. THIS MAY LEAD TO DEFIBRILLATION THERAPY BEING DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20651903 ↗MalfunctionNov 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THERE WAS AN ISSUE WITH THEIR DEVICE'S LID MAGNET. THE LID MAGNET CONTROLS THE DEVICE AUTOMATICALLY POWERING ON WHEN THE LID IS OPENED. AS A RESULT, THE DEVICE MAY NOT AUTOMATICALLY POWER ON WHEN THE LID IS OPENED. THIS MAY LEAD TO DEFIBRILLATION THERAPY BEING DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20651901 ↗MalfunctionOct 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20416435 ↗MalfunctionOct 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS MISSING ITS LID MAGNET. IN THIS STATE, THE DEVICE WOULD NOT DETECT THE LID BEING OPENED OR CLOSED, AND THEREFORE WOULD NOT AUTOMATICALLY POWER ON OR OFF WHICH CAN LEAD TO A USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20415530 ↗MalfunctionOct 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON INSPECTION STRYKER OBSERVED THEIR DEVICE WAS MISSING THE MAGNETIC PIN IN THE LID. IN THIS STATE THE DEVICE WILL NOT TURN ON AUTOMATICALLY, WHICH CAN LEAD TO A USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20398282 ↗MalfunctionOct 4, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20377800 ↗MalfunctionOct 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS MISSING THE MAGNET FROM THE LID OF THE DEVICE. IN THIS STATE THE DEVICE WILL NOT TURN ON AUTOMATICALLY WHEN THE LID IS ACTIVATED, WHICH CAN CAUSE A DELAY IN DEFIBRILLATION THERAPY IF NEEDED OR MAY LEAD TO USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20364577 ↗MalfunctionOct 3, 2024Failure to Power Up; Detachment of Device or Device Component; Failure to Power Up; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT POWER ON AS EXPECTED WHEN THE LID IS OPENED, WHICH CAN LEAD TO A USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20359967 ↗MalfunctionOct 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20357668 ↗MalfunctionOct 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USERS HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20357441 ↗MalfunctionOct 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20349255 ↗MalfunctionOct 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION, STRYKER OBSERVED THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20343948 ↗MalfunctionSep 26, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
STRYKER WAS PERFORMING PREVENTATIVE MAINTENANCE ON THE CUSTOMER'S DEVICE AND IT WAS OBSERVED THAT THE DEVICE WILL NOT DELIVER A SHOCK. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20311700 ↗MalfunctionSep 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20287131 ↗MalfunctionSep 20, 2024Failure to Sense; Failure to Sense
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE BACKING FROM THE ELECTRODES STUCK TO THE ELECTRODES, THEREFORE THEY WERE UNABLE TO USE THEM ON A PATIENT. THE CUSTOMER ADVISED THAT A BACK UP DEVICE WAS AVAILABLE AND WAS USED TO CONTINUE PATIENT CARE. THE PATIENT ASSOCIATED WITH THE REPORTED EVENT DID NOT SURVIVE. THE CUSTOMER INFORMED THAT IT IS UNLIKELY THE ISSUE WITH THE ELECTRODES CONTRIBUTED TO THE PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20264617 ↗MalfunctionSep 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET IN THEIR DEVICE LID WAS LOOSE. IN THIS STATE THE DEVICE MAY NOT TURN ON AUTOMATICALLY WHEN THE LID IS ACTIVATED, WHICH CAN CAUSE A DELAY IN DEFIBRILLATION THERAPY IF NEEDED OR MAY LEAD TO USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20264493 ↗MalfunctionSep 19, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION IT WAS FOUND THAT THE DEVICE WOULD NOT POWER ON WHEN LID IS OPENED. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20257376 ↗MalfunctionSep 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USERS HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20257309 ↗MalfunctionSep 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE DOES NOT TURN ON AUTOMATICALLY WHEN THE LID IS OPENED. THIS CAN RESULT IN A USER ERROR PREVENTING OR DELAYING DEFIBRILLATION THERAPY. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20256614 ↗MalfunctionSep 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS MISSING ITS LID MAGNET. IN THIS STATE, THE DEVICE WOULD NOT DETECT THE LID BEING OPENED OR CLOSED, AND THEREFORE WOULD NOT AUTOMATICALLY POWER ON OR OFF WHICH CAN LEAD TO A USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20256344 ↗MalfunctionSep 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USERS HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20255865 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P170018. One submission often covers many models, so check the product and lot details in each record.
- Z-2499-2025 · Jul 22, 2025 · Open, ClassifiedLIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEPAK CR2 defibrillator is indicated for use on patients 1 year of age
- Z-1189-2021 · Jan 14, 2021 · Open, ClassifiedPHYSIO CONTROL LIFEPAK CR2 Defibrillator 3321605-xxx The LIFEPAK CR2 defibrillator is designed for indoor and outdoor use in a stationary environment. After the electrode pads a
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
- AED PLUS 20000000100011010Zoll Medical Corporation · Automated external defibrillators (non-wearable)16 deaths1,072 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026