Accu-Chek Guide Link 08109257001Recalls & reported problems

Roche Diagnostics Operations, Inc. · ACCU-CHEK GUIDE LINK MG/DL 118 (US) · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00365702702134), received between Jun 30, 2021 and Apr 15, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fire” (2 reports).

14 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fire2
  • Display Difficult to Read1
  • Melted1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 15, 2024Display Difficult to Read; Display Difficult to Read

    IT WAS REPORTED THAT THE DISPLAY OF THE BLOOD GLUCOSE MONITOR APPEARED TO BE BURNT. THE CALLER REPORTED THAT THERE IS A BLACK SPOT WITH A WHITE RING AROUND IT, IN THE MIDDLE OF THE DISPLAY, THAT APPEARS TO BE A BURN MARK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19106949 ↗
  • MalfunctionNov 16, 2023Fire; Fire

    THE CALLER REPORTED THAT WHEN HE WAS DRIVING, HIS BLOOD GLUCOSE MONITOR STARTED SMOKING AND CAUGHT ON FIRE. THE CUSTOMER REPORTED THAT HE THREW THE DEVICE OUT THE WINDOW OF THE CAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18146527 ↗
  • MalfunctionJun 30, 2021Fire; Melted; Fire; Melted

    THE CALLER REPORTED THAT THE BLOOD GLUCOSE METER CAUGHT ON FIRE WHILE IT WAS STILL IN THE BOX. THE CUSTOMER OBSERVED THE METER BOX IN FLAMES AND STATED FLAMES AND SMOKE WERE VISIBLE AND THAT SHE SMELLED THE METER BURNING. THE METER HOUSING NOW SHOWS VISIBLE SIGNS OF MELTING AND THE CUSTOMER'S TABLE WAS DAMAGED BY THE FIRE. THE METER WAS NEW IN BOX AND HAD NEVER BEEN USED. THE CUSTOMER STATED THAT THE METER WAS NOT EXPOSED TO ANY EXTERNAL HEAT SOURCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12092417 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2981-2026 · Jul 13, 2026 · Open, Classified
    MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
  • Z-1742-2026 · Feb 13, 2026 · Open, Classified
    MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1739-2026 · Feb 13, 2026 · Open, Classified
    MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
  • Z-1261-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
  • Z-1265-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1266-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
  • Z-0003-2025 · Jul 31, 2024 · Open, Classified
    MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G
  • Z-0193-2023 · Sep 20, 2022 · Open, Classified
    Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
  • Z-0855-2022 · Jan 27, 2022 · Open, Classified
    MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
  • Z-1771-2021 · Mar 15, 2021 · Open, Classified
    MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE SF M
  • Z-0516-2021 · Apr 30, 2020 · Open, Classified
    MiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
  • Z-0956-2020 · Nov 21, 2019 · Open, Classified
    Medtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
  • Z-2605-2017 · May 8, 2017 · Terminated
    MiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.
  • Z-2604-2017 · May 8, 2017 · Terminated
    MiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.

Other Automated Insulin Dosing Device System, Single Hormonal Control devices

All Automated Insulin Dosing Device System, Single Hormonal Control devices →