Skip to content

Alcon 00380650001399Recalls & reported problems

Alcon Laboratories, Inc. · ACTIVE FMS W/ 0.9 MM INFUSION SLEEVES, 0.9MM 45° ABS* INTREPID* Balanced Tip · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380650001399), received between Jan 3, 2023 and Oct 10, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Accessory Incompatible” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 28, 2023Class IIOpen, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2407-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Accessory Incompatible1
  • Fitting Problem1
  • Material Twisted/Bent1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2023Accessory Incompatible; Accessory Incompatible

    A CUSTOMER REPORTED THAT THE SLEEVES THAT WE HAD AND USED ON AN INFINITI MACHINE WERE ¿THICKER AND LESS PLIABLE¿ DURING CATARACT SURGERY (WITH INTRAOCULAR LENSES IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17903435 ↗
  • MalfunctionAug 3, 2023Fitting Problem; Fitting Problem

    A CUSTOMER REPORTED THAT THE SLEEVES ARE THINNER THAN USUAL, THE SLEEVES DO NOT FIT AS TIGHT ON THE TIPS AND THAT IS WHERE THE ¿EXTRA¿ FLUID WAS CAME FROM DURING CATARACT SURGERY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17450730 ↗
  • MalfunctionJan 3, 2023Material Twisted/Bent; Suction Failure; Material Twisted/Bent; Suction Failure

    A CUSTOMER REPORTED THEY HAD VACUUM ISSUES WHILE IN QUADRANT REMOVAL USING THE CASSETTE. IT WAS LATER REALIZED THAT THE FMS (FLUIDICS MANAGEMENT SYSTEM) WAS KINKED DUE TO WHICH THE ISSUE OCCURRED. THE PROCEDURE DETAILS AND PATIENT IMPACT WAS NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16086670 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alcon 00380650001399

Has the Alcon 00380650001399 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2407-2023 (Class II), began on Jun 28, 2023: Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Alcon 00380650001399?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jan 3, 2023 and Oct 10, 2023. In those reports, the most common specific device problems are “Accessory Incompatible” (1), “Fitting Problem” (1) and “Material Twisted/Bent” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alcon 00380650001399?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alcon 00380650001399, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Alcon, model 00380650001399). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page