Skip to content

Alcon 00380650001399Recalls & reported problems

Alcon Laboratories, Inc. · ACTIVE FMS W/ 0.9 MM INFUSION SLEEVES, 0.9MM 45° ABS* INTREPID* Balanced Tip · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380650001399), received between Jan 3, 2023 and Oct 10, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Accessory Incompatible” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 28, 2023Open, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2407-2023 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Accessory Incompatible1
  • Fitting Problem1
  • Material Twisted/Bent1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2023Accessory Incompatible; Accessory Incompatible

    A CUSTOMER REPORTED THAT THE SLEEVES THAT WE HAD AND USED ON AN INFINITI MACHINE WERE ¿THICKER AND LESS PLIABLE¿ DURING CATARACT SURGERY (WITH INTRAOCULAR LENSES IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17903435 ↗
  • MalfunctionAug 3, 2023Fitting Problem; Fitting Problem

    A CUSTOMER REPORTED THAT THE SLEEVES ARE THINNER THAN USUAL, THE SLEEVES DO NOT FIT AS TIGHT ON THE TIPS AND THAT IS WHERE THE ¿EXTRA¿ FLUID WAS CAME FROM DURING CATARACT SURGERY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17450730 ↗
  • MalfunctionJan 3, 2023Material Twisted/Bent; Suction Failure; Material Twisted/Bent; Suction Failure

    A CUSTOMER REPORTED THEY HAD VACUUM ISSUES WHILE IN QUADRANT REMOVAL USING THE CASSETTE. IT WAS LATER REALIZED THAT THE FMS (FLUIDICS MANAGEMENT SYSTEM) WAS KINKED DUE TO WHICH THE ISSUE OCCURRED. THE PROCEDURE DETAILS AND PATIENT IMPACT WAS NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16086670 ↗

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.