Skip to content

Alcon 8065752202Recalls & reported problems

Alcon Laboratories, Inc. · ACTIVE FMS W/ INTREPID* NANO SLEEVES, 0.9MM 30° ABS* INTREPID* Balanced Tip · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00380657522026), received between Jun 9, 2021 and Dec 11, 2023 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Inability to Irrigate” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 28, 2023Class IIOpen, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2407-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inability to Irrigate3
  • Air/Gas in Device2
  • Break2
  • Complete Blockage1
  • No Flow1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Burn(s)1
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2023Complete Blockage; Complete Blockage

    A PHYSICIAN REPORTED THAT THE CASSETTE TUBE WAS CLOGGED DURING IRRIGATION OR ASPIRATION SURGERY. THE PROCEDURE WAS COMPLETED WITH ANOTHER PACK. THE PATIENT HARM DETAILS WAS NOT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18303067 ↗
  • MalfunctionSep 11, 2023Inability to Irrigate; Inability to Irrigate

    A NON HEALTH CARE PROFESSIONAL REPORTED THAT THERE WAS NO IRRIGATION OF FLUIDICS MANAGEMENT SYSTEM (FMS) WHEN IT WAS IN IRRIGATION ASPIRATION MODE DURING CATARACT SURGERY. THERE WERE NO SYSTEM MESSAGES OR ERROR CODES. THEY HAD TO OPEN NEW PACKS TO COMPLETE THE PROCEDURES. THIS OCCURRED PRIOR TO PATIENT CONTACT BUT DURING SURGERY AFTER CORTEX. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17719733 ↗
  • MalfunctionSep 11, 2023Inability to Irrigate; Inability to Irrigate

    A NON HEALTH CARE PROFESSIONAL REPORTED THAT THERE WAS NO IRRIGATION OF FLUIDICS MANAGEMENT SYSTEM (FMS) WHEN IT WAS IN IRRIGATION ASPIRATION MODE DURING CATARACT SURGERY. THERE WERE NO SYSTEM MESSAGES OR ERROR CODES. THEY HAD TO OPEN NEW PACKS TO COMPLETE THE PROCEDURES. THIS OCCURRED PRIOR TO PATIENT CONTACT BUT DURING SURGERY AFTER CORTEX. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17719638 ↗
  • MalfunctionSep 11, 2023Inability to Irrigate; Inability to Irrigate

    A NON HEALTH CARE PROFESSIONAL REPORTED THAT THERE WAS NO IRRIGATION OF FLUIDICS MANAGEMENT SYSTEM (FMS) WHEN IT WAS IN IRRIGATION ASPIRATION MODE DURING CATARACT SURGERY. THERE WERE NO SYSTEM MESSAGES OR ERROR CODES. THEY HAD TO OPEN NEW PACKS TO COMPLETE THE PROCEDURES. THIS OCCURRED PRIOR TO PATIENT CONTACT BUT DURING SURGERY AFTER CORTEX. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17719634 ↗
  • MalfunctionApr 24, 2023Air/Gas in Device; Air/Gas in Device

    A PHARMACIST REPORTED DURING NUCLEUS REMOVAL/ASPIRATION, SURGEON NOTED BUBBLES CAME FROM HANDPIECE, THESE BUBBLES IMPEDED THE SURGEON'S VISION. THERE WAS NO IMPACT ON PATIENT. BUBBLES WERE ASPIRATED. ADDITIONAL INFORMATION WAS REQUESTED; HOWEVER, NONE HAS BEEN RECEIVED TO DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16793963 ↗
  • MalfunctionMar 3, 2023Air/Gas in Device; Air/Gas in Device

    A PHARMACIST REPORTED THAT BUBBLES WERE NOTED DURING SCULPTURE AND QUADRANT STEPS WHICH LED TO OBSTRUCTION IN THE SURGEON'S VIEW WHILE PERFORMING THE CATARACT PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16477791 ↗
  • MalfunctionJul 19, 2022Break; Break

    A PHYSICIAN REPORTED BROKEN PHACOEMULSIFICATION TIP BROKE OUT OF PATIENT'S EYE DURING SCULPTING PHASE OF THE CATARACT PROCEDURE; HOWEVER, THERE WAS NO PATIENT CONTACT WHEN THIS WAS OBSERVED. THE BROKEN TIP WAS REPLACED BY A MINI TIP FOR COMPLETING THE PROCEDURE. THERE WAS NO PATIENT HARM. THERE WERE TWO PRODUCTS REPORTED FROM THE SITE AS POTENTIALLY INVOLVED THE REPORTER COULD NOT PROVIDE CLARIFICATION REGARDING WHICH PRODUCT IS ASSOCIATED WITH THIS COMPLAINT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15051057 ↗
  • MalfunctionJul 19, 2022Break; Break

    A PHYSICIAN REPORTED BROKEN PHACOEMULSIFICATION TIP BROKE OUT OF PATIENT'S EYE DURING SCULPTING PHASE OF THE CATARACT PROCEDURE; HOWEVER, THERE WAS NO PATIENT CONTACT WHEN THIS WAS OBSERVED. THE BROKEN TIP WAS REPLACED BY A MINI TIP FOR COMPLETING THE PROCEDURE. THERE WAS NO PATIENT HARM. THERE WERE TWO PRODUCTS REPORTED FROM THE SITE AS POTENTIALLY INVOLVED THE REPORTER COULD NOT PROVIDE CLARIFICATION REGARDING WHICH PRODUCT IS ASSOCIATED WITH THIS COMPLAINT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15051053 ↗
  • InjuryJan 3, 2022No Flow; No Flow

    A PHYSICIAN REPORTED NO IRRIGATION FLOW THROUGH THE SLEEVE WOUND BURN OCCURRED SOMEWHERE AROUND SCULPTING AND MOVING TO QUADRANT REMOVAL IN THE LEFT EYE DURING A CATARACT SURGERY. THE WOUND WAS SUTURED AND CLOSED.

    Patient problems: Burn(s)

    MAUDE report 13141582 ↗
  • MalfunctionJun 9, 2021Suction Failure; Suction Failure

    A CUSTOMER REPORTED THAT THE CASSETTE DID NOT ASPIRATE DURING SURGERY. THE PRODUCT WAS REPLACED AND THE SURGERY WAS COMPLETED. THERE IS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11969146 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alcon 8065752202

Has the Alcon 8065752202 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2407-2023 (Class II), began on Jun 28, 2023: Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Alcon 8065752202?

The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jun 9, 2021 and Dec 11, 2023. In those reports, the most common specific device problems are “Inability to Irrigate” (3), “Air/Gas in Device” (2) and “Break” (2); the most common patient problem is “Burn(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alcon 8065752202?

None of the reports are classified as deaths, 1 as an injury and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alcon 8065752202, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Alcon, model 8065752202). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page