Skip to content

CENTURION®, INTREPID® 8065753135Recalls & reported problems

Alcon Laboratories, Inc. · CENTURION® Vision System Active FMS w/ INTREPID Ultra Sleeves, 0.9MM 45° ABS* INTERPID Hybrid Tip · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380657531356), received between Jul 10, 2024 and Aug 19, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Suction Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 28, 2023Open, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2407-2023 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Suction Failure2
  • Suction Problem2
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2025Suction Problem; Obstruction of Flow; Suction Problem; Obstruction of Flow

    ADDITIONAL INFORMATION RECEIVED CONFIRMING THE SUSPECTED ISSUE APPEARS TO BE CLOGGING IN THE FLUIDICS MANAGEMENT SYSTEM PACK, RATHER THAN A PROBLEM WITH THE PROBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22837441 ↗
  • MalfunctionJun 17, 2025Suction Problem; Suction Problem

    A PHYSICIAN REPORTED THAT ASPIRATION PRESSURE DID NOT INCREASE DURING CATARACT SURGERY (WITH INTRAOCULAR LENS IMPLANT). THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22231422 ↗
  • MalfunctionDec 9, 2024Suction Failure; Suction Failure

    A PHYSICIAN REPORTED THAT THE ASPIRATION FAILURE OCCURRED DURING ULTRA SOUND. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20876151 ↗
  • MalfunctionJul 10, 2024Suction Failure; Suction Failure

    A PHYSICIAN REPORTED THAT THE ASPIRATION FAILURE OCCURRED AND AN ERROR WAS DISPLAYED DURING ULTRA SOUND (US) MODE OF CATARACT SURGERY (WITH (IOL) INTRAOCULAR LENS IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19708429 ↗