ARCHITECT 7D71-22Recalls & reported problems
Abbott Laboratories · Phosphorus · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740012946), received between Sep 1, 2017 and Sep 1, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- High Test Results1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2017High Test Results; High Test Results
THE CUSTOMER REPORTS THAT ONE PATIENT (BEING TREATED FOR HYPERCALCEMIA AND RECEIVING CALCITONIN / HEMOFILTERED / 24 HOURS) SAMPLE GENERATED THE FOLLOWING CLIN CHEM PHOSPHORUS ASSAY RESULTS ON AN ARCHITECT C8000 ANALYZER: (B)(6): 3.87, 2.76, 3.27 AND 2.52 MMOL/L. THE SAMPLE WAS DRAWN WITH A HEPARIN COLLECTION TUBE. AN ADDITIONAL SAMPLE WAS DRAWN ON THIS PATIENT LATER IN THE DAY ((B)(6)) AND GENERATED RESULTS IN THE RANGE OF 1.35-1.83 MMOL/L (N=13) ALSO DRAWN IN A HEPARIN COLLECTION TUBE. THE CUSTOMER THEN RAN A SAMPLE FROM THIS SAME PATIENT THAT WAS DRAWN IN A RED TOP COLLECTION TUBE ((B)(6)). THIS SAMPLE GENERATED RESULTS IN THE RANGE OF 0.98-1.27 MMOL/L (N=10). THE CUSTOMER USES A NORMAL REFERENCE RANGE OF 0.74-1.52 MMOL/L. THE PATIENT IS IN RECOVERY. NO SUSPECT RESULTS WERE REPORTED FROM THE LAB. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6839718 ↗
Other Phosphomolybdate (colorimetric), inorganic phosphorus devices
- Synchron 467868Beckman Coulter, Inc. · Phosphomolybdate (Colorimetric), Inorganic Phosphorus1 recall0 reports
- VITROS 1513209Ortho-Clinical Diagnostics, Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 recall0 reports
- Phosphate (Inorganic) ver.2 03183793122Roche Diagnostics GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus11 reports
- Phosphate (Inorganic) ver.2 5171377190Roche Diagnostics GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus5 reports
- Phosphate (Inorganic) ver.2 08058610190Roche Diagnostics GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus4 reports
- ARCHITECT 7D71-23Abbott GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus2 reports
- Dimension Vista® Flex® reagent cartridge PHOS 10445144Siemens Healthcare Diagnostics Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 report
- Alinity 08P4030Abbott GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.