ARCHITECT 01G06-01Recalls & reported problems
Abbott Laboratories · ARCHITECT c8000 PROCESSING MODULE · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740023089), received between May 29, 2024 and Oct 29, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results3
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED POTASSIUM RESULTS GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): INITIAL REPORT = 6.5 MMOL/L, REPEAT = 4.8 MMOL/L. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23413712 ↗MalfunctionApr 21, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED A FALSELY DECREASED SODIUM RESULT GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE. RESULTS PROVIDED: 111 / 140 MMOL/L (NORMAL RANGE: 135-145 MMOL/L). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21871477 ↗MalfunctionSep 3, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED CREATININE ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 0.72 - 1.25 MG/DL): INITIAL CREATININE RESULT= 14 MG/DL; REPEAT RESULT= 0.71 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20127653 ↗MalfunctionMay 29, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR TWO PATIENTS. THE CUSTOMER IS USING A NON-ABBOTT MAGNESIUM REAGENT. THE FIELD SERVICE REPRESENTATIVE (FSR) WENT ONSITE TO INSPECT THE INSTRUMENT. THE SAMPLES WERE REPEATED AFTER THE REPAIR AND THE RESULTS WERE NORMAL. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S NORMAL RANGE: < 1.02 MMOL/L. SID (B)(6) INITIAL RESULT = >10 MMOL/L, REPEAT RESULT POST REPAIR = 0.76 MMOL/L. SID (B)(6) INITIAL RESULT = >2 MMOL/L, REPEAT RESULT POST REPAIR = 0.87 MMOL/L . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19418153 ↗
Other Analyzer, chemistry (photometric, discrete), for clinical use devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,197 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall911 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death694 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death646 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE685 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.