Abbott Laboratories · ARCHITECT c8000 PROCESSING MODULE · Class 1

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740023089), received between May 29, 2024 and Oct 29, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results3
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED POTASSIUM RESULTS GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): INITIAL REPORT = 6.5 MMOL/L, REPEAT = 4.8 MMOL/L. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413712 ↗
  • MalfunctionApr 21, 2025Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY DECREASED SODIUM RESULT GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE. RESULTS PROVIDED: 111 / 140 MMOL/L (NORMAL RANGE: 135-145 MMOL/L). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21871477 ↗
  • MalfunctionSep 3, 2024High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED CREATININE ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 0.72 - 1.25 MG/DL): INITIAL CREATININE RESULT= 14 MG/DL; REPEAT RESULT= 0.71 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20127653 ↗
  • MalfunctionMay 29, 2024High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR TWO PATIENTS. THE CUSTOMER IS USING A NON-ABBOTT MAGNESIUM REAGENT. THE FIELD SERVICE REPRESENTATIVE (FSR) WENT ONSITE TO INSPECT THE INSTRUMENT. THE SAMPLES WERE REPEATED AFTER THE REPAIR AND THE RESULTS WERE NORMAL. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S NORMAL RANGE: < 1.02 MMOL/L. SID (B)(6) INITIAL RESULT = >10 MMOL/L, REPEAT RESULT POST REPAIR = 0.76 MMOL/L. SID (B)(6) INITIAL RESULT = >2 MMOL/L, REPEAT RESULT POST REPAIR = 0.87 MMOL/L . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19418153 ↗

Other Analyzer, chemistry (photometric, discrete), for clinical use devices