ARCHITECT 3M74-98Recalls & reported problems
Abbott Laboratories · ARCHITECT i2000SR REFURBISHED PROCESSING MODULE · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00380740068851), received between Aug 1, 2016 and Oct 8, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Smoking2
- Thermal Decomposition of Device2
- False Negative Result1
- False Positive Result1
- High Test Results1
Patient problems
- No Consequences Or Impact To Patient6
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2025Thermal Decomposition of Device; Thermal Decomposition of Device
THE CUSTOMER REPORTED A BURNING SMELL EMANATING FROM THE ARCHITECT I2000SR ANALYZER. UPON INSPECTION BY ABBOTT FIELD SERVICE, BURN MARKS WERE FOUND ON THE ANTENNA BOARD OF THE SAMPLE HANDLER FOR STAT PIPETTORS. NO IMPACT TO USER SAFETY OR PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23242884 ↗MalfunctionJul 5, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE ARCHITECT SARS-COV-2 IGG RESULT FOR ONE PATIENT SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: ON (B)(6) 2023 SID (B)(6): INITIAL SARS-COV-2 IGG RESULT = 2.58 INDEX; REPEAT RESULT = 0.54 INDEX (REFERENCE RANGE: < 1.4 INDEX IS NEGATIVE, >/= 1.4 INDEX IS POSITIVE) THE CUSTOMER STATED QC FOR SARS-COV-2 IGG WAS WITHIN RANGE. THE CUSTOMER ALSO STATED THE QC WAS OUT OF RANGE FOR THEIR COV-2 IGG II NEGATIVE CONTROL HOWEVER WHEN REPEATED WAS WITHIN RANGE (INITIAL = 656 AU/ML; REPEAT = 5.8 AU/ML). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17263511 ↗MalfunctionMar 5, 2021False Negative Result; False Negative Result
THE CUSTOMER OBSERVED A FALSE NEGATIVE SARS-COV-2 IGG RESULT FOR ONE PATIENT ON AN ARCHITECT I2000SR ANALYZER. THE FOLLOWING DATA WAS PROVIDED (<1.4 INDEX IS NEGATIVE, >/=1.4 INDEX IS POSITIVE): SAMPLE ID (B)(6) INITIAL RESULT, ON (B)(6) 2021, WAS 2.16, REPEATS WERE 0.52 AND 2.02 INDEX. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11426234 ↗MalfunctionJul 23, 2020Smoking; Smoking
THE CUSTOMER OBSERVED SMOKE AND A BURNING SMELL COMING FROM THE SCC POWER SUPPLY OF THE ARCHITECT I2000SR WHEN TRYING TO POWER ON THE INSTRUMENT MULTIPLE TIMES. THE SMOKE WENT AWAY WHEN THE POWER WAS TURNED OFF. THERE WAS NO IMPACT TO PATIENT MANAGEMENT, USER SAFETY, OR FACILITY DAMAGE REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10317233 ↗MalfunctionNov 5, 2018Thermal Decomposition of Device; Thermal Decomposition of Device
THE CUSTOMER REPORTED THAT THE SCC SCREEN WENT BLACK AND WOULD NOT RESTART. THE CUSTOMER REPORTED A BURNING SMELL COMING FROM THE ARCHITECT ANALYZER. THERE WAS NO VISIBLE FIRE OR SMOKE OBSERVED. THE CUSTOMER FOUND THAT THE SCC POWER SUPPLY HAS A BURN MARK ON IT. THERE WAS NO IMPACT TO PATIENT MANAGEMENT, USER SAFETY, OR FACILITY DAMAGE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8038045 ↗MalfunctionApr 24, 2017High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSE POSITIVE B-HCG RESULT ON THE ARCHITECT I2000SR ANALYZER. THE FOLLOWING DATA WAS PROVIDED: INITIAL 66.7, REPEATED ON THIS INSTRUMENT AND ANOTHER INSTRUMENT, BOTH RESULTS CAME OUT AS < 2.0 RESULT UNIT IS MIU/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6516587 ↗MalfunctionNov 16, 2016Smoking; Smoking
THE CUSTOMER OBSERVED SMOKE FROM THE ARCHITECT I2000SR ANALYZER SYSTEM CONTROL CENTER (SCC). THE CUSTOMER STATED THAT THEY SMELLED SMOKE, AND WHEN THEY OPENED THE DOOR WHERE THE SCC IS LOCATED, THEY SAW VISIBLE SMOKE COMING FROM THE REAR OF THE SCC. NO FIRE OR OTHER VISIBLE DAMAGE WAS SEEN. THE ANALYZER AND THE SCC WERE POWERED OFF. NO FIRE OR INJURY WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 6107261 ↗MalfunctionAug 17, 2016Fluid/Blood Leak; Fluid/Blood Leak
FIELD SERVICE REPLACED THE INCORRECT TUBING ON THE ARCHITECT I2000SR ANALYZER PER THE MANDATORY TECHNICAL SERVICE BULLETIN. THERE WAS NO IMPACT TO PATIENT RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5884522 ↗MalfunctionAug 17, 2016Fluid/Blood Leak; Fluid/Blood Leak
FIELD SERVICE REPLACED THE INCORRECT TUBING ON THE ARCHITECT I2000SR ANALYZER PER THE MANDATORY TECHNICAL SERVICE BULLETIN. THERE WAS NO IMPACT TO PATIENT RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5884246 ↗MalfunctionAug 17, 2016Fluid/Blood Leak; Fluid/Blood Leak
FIELD SERVICE REPLACED THE INCORRECT TUBING ON THE ARCHITECT I2000SR ANALYZER PER THE MANDATORY TECHNICAL SERVICE BULLETIN. THERE WAS NO IMPACT TO PATIENT RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5884168 ↗MalfunctionAug 17, 2016Fluid/Blood Leak; Fluid/Blood Leak
FIELD SERVICE REPLACED THE INCORRECT TUBING ON THE ARCHITECT I2000SR ANALYZER PER THE MANDATORY TECHNICAL SERVICE BULLETIN. THERE WAS NO IMPACT TO PATIENT RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5882593 ↗MalfunctionAug 1, 2016Fluid/Blood Leak; Fluid/Blood Leak
FIELD SERVICE REPLACED THE INCORRECT TUBING ON THE ARCHITECT I2000SR ANALYZER PER THE MANDATORY TECHNICAL SERVICE BULLETIN. THERE WAS NO IMPACT TO PATIENT RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5838137 ↗
Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,197 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall911 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death694 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death646 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE685 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page