ARCHITECT 07K68-35Recalls & reported problems
Abbott Ireland Diagnostics · ARCHITECT CEA Reagent kit · Class 2
All models of ARCHITECT (23)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380740112424), received between Apr 12, 2018 and Jun 10, 2022 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Low Test Results7
- High Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient3
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2022Low Test Results; Low Test Results
THE CUSTOMER SUSPECTS FALSELY DEPRESSED ARCHITECT CEA AND ARCHITECT CA 19-9XR RESULTS FOR A PATIENT. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6)2021: CEA: 11.3 NG/ML / CA19-9 RESULT: 246.2 U/ML ON (B)(6)2022: CEA: 16.3 NG/ML / CA 19-9 RESULT: 166.6 U/ML ON (B)(6)2022 SID(B)(6): CEA RESULT: 4.38 NG/ML, CA 19-9 RESULT: 22.32 U/ML RETESTED ON (B)(6)2022: CEA RESULT: 4.42 NG/ML, CA 19-9 RESULT: 26.5 U/ML THE PATIENT¿S CLINICAL HISTORY IS UNKNOWN AND DOCTOR SUSPECTS MAY¿S RESULTS ARE LOW. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14658197 ↗MalfunctionApr 25, 2022Low Test Results; Low Test Results
THE CUSTOMER IDENTIFIED FALSELY DEPRESSED ARCHITECT CEA (CARCINOEMBRYONIC ANTIGEN) RESULTS FOR ONE PATIENT. THE CUSTOMER PROVIDED THE FOLLOWING INFORMATION: CEA (B)(6) 2022 SID: (B)(6) GENERATED WITH SERIAL NUMBER (B)(4) RESULT 80.25 NG/ML WITH REAGENT LOT 29180FN00; GENERATED WITH SERIAL NUMBER (B)(4) RESULT < 0.50 NG / ML WITH REAGENT LOT 29180FN00, RETEST 1:1 79.80 NG/ML AND 84.3 NG/ML. THE CUSTOMER REPORTED RESULTS AS 80.25 NG/ML. PATIENT HISTORY: (B)(6) 2022 12.3 NG/L. THE PATIENT IS A (B)(6) FEMALE WITH A MALIGNANT TUMOR; THE CUSTOMER IS NOT AWARE OF WHAT KIND OF CANCER THE PATIENT HAS BEEN DIAGNOSED WITH. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14191550 ↗MalfunctionDec 17, 2021Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT CEA (CARCINOEMBRYONIC ANTIGEN) RESULTS FOR A 67 YEAR OLD MALE DIAGNOSED WITH DIGESTION VISCERAL CANCER. THE FOLLOWING DATA WAS PROVIDED: 01DEC2021 SID (B)(6) INITIAL RESULT WAS >1500 NG/ML, REPEAT = 9.7 NG/ML, REPEAT WITH A DILUTION = 2114.7 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13028546 ↗MalfunctionJul 19, 2019Low Test Results; Low Test Results
THE CUSTOMER STATED THAT A FALSELY DECREASED ARCHITECT CEA RESULT WAS GENERATED FOR A PATIENT SAMPLE (ID (B)(6)). THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT: 316.60 NG/ML. RETEST 1: 0.5 NG/ML. RETEST 2: 315.32 NG/ML. RETEST 3: 316.72 NG/ML. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8809765 ↗InjuryJan 30, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSE ELEVATED ARCHITECT CEA RESULT. THE SAMPLE GENERATED A RESULT OF 50.3 NG/ML. THE PATIENT'S PREVIOUS RESULT WAS 28.1 NG/ML. THE CUSTOMER INDICATED THE PATIENT HAD A PET-CT, COLON FIBERSCOPE AND GASTROSCOPY DUE TO THE ELEVATED CEA RESULTS. THERE WAS NO ADVERSE IMPACT TO THE PATIENT DUE TO THE PROCEDURES.
Patient problems: No Code Available
MAUDE report 8289790 ↗MalfunctionDec 3, 2018Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED RESULTS WHILE USING ARCHITECT CEA (CARCINOEMBRYONIC ANTIGEN) REAGENTS. THE FOLLOWING DATA WAS PROVIDED FOR THE SAME PATIENT. SID (B)(6) (TESTED (B)(6) 2018) INITIAL 11.36 NG/ML, REPEAT 10.92 NG/ML. PREVIOUS RESULTS TESTED ON (B)(6) 2018 WERE LOWER, SID (B)(6) INITIAL 22.94 NG/ML, REPEAT 23.76 NG/ML. THE PATIENT WAS DIAGNOSED WITH CANCER; HOWEVER, THE SPECIFIC CANCER DIAGNOSIS WAS NOT PROVIDED. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8126684 ↗MalfunctionOct 24, 2018Low Test Results; Low Test Results
THE CUSTOMER OBSERVED POOR REPRODUCIBILITY AND FALSELY DEPRESSED RESULTS WHILE USING ARCHITECT CEA REAGENTS. THE FOLLOWING DATA WAS PROVIDED. PATIENT 1 SID (B)(6) REPEAT USING ANALYZER SN (B)(4) 4.9 IU/ML, INITIAL VALUE WAS HIGHER USING ANOTHER ANALYZER (SN (B)(4)) 937 IU/ML. PATIENT 2 SID (B)(6) REPEAT USING ANALYZER SN (B)(4) 0.5 IU/ML, INITIAL AND REPEAT VALUES WERE HIGHER USING ANOTHER ANALYZER (SN (B)(4)) 5.4, 5.63 IU/ML. PATIENT 3 SID (B)(6) REPEAT USING ANALYZER SN (B)(4) 0.5 IU/ML, INITIAL VALUE WAS HIGHER USING ANOTHER ANALYZER (SN (B)(4)) 4.9 IU/ML. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8003074 ↗MalfunctionApr 12, 2018Low Test Results; Low Test Results
THE CUSTOMER OBSERVED ERRATIC RESULTS WHILE USING THE ARCHITECT CEA ASSAY. THE CUSTOMER PROVIDED THE FOLLOWING RESULTS DATA: ON (B)(6) 2018 CEA INITIAL 0.5, RETEST 4.4 NG/ML THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED. NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7426437 ↗
FDA clearance and classification
The FDA regulates product code DHX (system, test, carcinoembryonic antigen) under 21 CFR 866.6010 as a Class II device. 21 CFR 866.6010 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ARCHITECT 07K68-35
Has the ARCHITECT 07K68-35 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00380740112424), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ARCHITECT 07K68-35?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Apr 12, 2018 and Jun 10, 2022. In those reports, the most common specific device problems are “Low Test Results” (7) and “High Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARCHITECT 07K68-35?
None of the reports are classified as deaths, 1 as an injury and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ARCHITECT 07K68-35, and is it still sold?
The FDA's device database lists Abbott Ireland Diagnostics as the company responsible for it (brand ARCHITECT, model 07K68-35). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN devices
- ADVIA Centaur® CEA 10309977Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN14 reports
- ARCHITECT 07K68-27Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN13 reports
- Elecsys CEA 07027079190Roche Diagnostics GmbH · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN13 reports
- Atellica IM CEA 10995523Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN11 reports
- ADVIA Centaur® CEA 10309978Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN10 reports
- IMMULITE® 2000 Systems CEA 10380995Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN9 reports
- Elecsys Anti-CCP 07251670190Roche Diagnostics GmbH · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN8 reports
- Alinity 07P6220Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page