IMMULITE® 2000 Systems CEA 10380995Recalls & reported problems
Siemens Healthcare Diagnostics Products Ltd · CEA - IMMULITE - RGT - 600 Tests · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00630414961125), received between Jun 2, 2023 and Jun 14, 2023 — classified as 9 malfunctions.
The most frequently reported specific device problem is “High Test Results” (9 reports).
1 other recall cover devices cleared under the same FDA submission (K980855); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results9
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17125060 ↗MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17125008 ↗MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17124954 ↗MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17124909 ↗MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17124888 ↗MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17124837 ↗MalfunctionJun 14, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. IT IS UNKNOWN IF THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17124772 ↗MalfunctionJun 2, 2023High Test Results; High Test Results
A FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULT WAS OBTAINED ON SAMPLE FROM ONE PATIENT ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17050258 ↗MalfunctionJun 2, 2023High Test Results; High Test Results
FALSELY ELEVATED CEA (CARCINOEMBRYONIC ANTIGEN) RESULTS WERE OBTAINED ON SAMPLES FROM TWO PATIENTS ON AN IMMULITE 2000 XPI INSTRUMENT AND CONSIDERED DISCORDANT WITH THE LOWER REPEAT RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CEA RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17050071 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K980855. One submission often covers many models, so check the product and lot details in each record.
- Z-1679-2018 · Jan 2, 2018 · TerminatedIMMULITE 2000/IMMULITE 2000 XPi CEA
Other SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN devices
- ADVIA Centaur® CEA 10309977Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN14 reports
- ARCHITECT 07K68-27Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN13 reports
- Elecsys CEA 07027079190Roche Diagnostics GmbH · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN13 reports
- Atellica IM CEA 10995523Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN11 reports
- ADVIA Centaur® CEA 10309978Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN10 reports
- Elecsys Anti-CCP 07251670190Roche Diagnostics GmbH · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN8 reports
- ARCHITECT 07K68-35Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN8 reports
- Alinity 07P6220Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page