Skip to content

Alinity 07P5030Recalls & reported problems

Abbott Ireland · Alinity i Estradiol Reagent Kit 1000 Tests · Class 1

All models of Alinity (22)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00380740129927), received between Jul 8, 2019 and Mar 11, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results6

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2024Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DECREASED ALINITY I ESTRADIOL RESULTS WHICH QUESTIONED BY THE PHYSICIAN ON MULTIPLE PATIENTS FROM MULTIPLE SPECIMENS. THE RESULTS PROVIDED WERE: ON (B)(6) 2024 PATIENT 1 SID (B)(6)=9596.4 PMOL/L /REPEATED= 11705.3 PMOL/L; SECOND SPECIMEN SID (B)(6)=1872.3 PMOL/L ; THIRD SPECIMEN SID (B)(6)=13230.6 PMOL/L PATIENT 4: SID (B)(6) INITIAL= 2259.9 PMOL/L /REPEATED ON (B)(6) 2024=4361.6 PMOL/L ON (B)(6) 2024 PATIENT 2 SID (B)(6) INITIAL= 4296 PMOL/L /REPEATED=4779.5 PMOL/L; SECOND SPECIMEN ON (B)(6) 2024 SID (B)(6) INITIAL=2773.1 PMOL/L /REPEATED=8476.2 PMOL/L ON (B)(6) 2024 PATIENT 3: SID (B)(6) INITIAL=6117.8 PMOL/L /REPEATED ON (B)(6) 2024=5476.4 AND 6625.4 PMOL/L; SECOND SPECIMEN ON (B)(6) 2024 SID (B)(6) INITIAL=734.5 PMOL/L /REPEATED=1077.5 PMOL/L AND 11930.5 PMOL/L THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18874348 ↗
  • MalfunctionSep 21, 2023Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY DEPRESSED ALINITY I ESTRADIOL RESULT FOR A PATIENT ON THE ALINITY I PROCESSING MODULE. THE SAMPLE WAS REPEATED AFTER THE PHYSICIAN QUESTIONED THE RESULT. THE FOLLOWING DATA WAS PROVIDED FOR SID (B)(6). INITIAL RESULT =414.5 PG/ML (1522 PMOL/L) AND THE REPEAT RESULT = 4087.8 PG/ML (15008 PMOL/L). THE CUSTOMER USES A REPORTABLE RANGE: FOLLICULAR: 77-921 MID-CYCLE:140-2380 LUTEAL:77-1145 THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17789091 ↗
  • MalfunctionJul 12, 2023Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED A FALSELY DECREASED ALINITY I ESTRADIOL RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: SAMPLE ID (B)(6) INITIAL RESULT, ON 22JUN2023, WAS 985, REPEAT, ON ANOTHER ANALYZER WAS >1000, REPEAT, WITH A 1:5 DILUTION, WAS 1817 PG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17308425 ↗
  • MalfunctionApr 24, 2023Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY DECREASED ALINITY I ESTRADIOL RESULT ON A FEMALE PATIENT. RESULTS PROVIDED: (B)(6) 2023, (B)(6) = 3017 / 3243 PG/ML (ALINITY SERIAL NUMBER (B)(6 ) (B)(6) 2023 (B)(6)= 565 PG/ML (ALINITY SERIAL NUMBER (B)(6) ) REPEATED BOTH SAMPLES ON (B)(6) 2023: (B)(6) = 3672 PG/ML; (B)(6)= >1000 / 3415 PG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16794867 ↗
  • MalfunctionSep 15, 2021Low Test Results; Low Test Results

    ON (B)(6) 2021, THE CUSTOMER PROVIDED A SECOND CASE: SID: (B)(6). INITIAL ESTRADIOL RESULT = 930 PG/ML WHICH DOES NOT CORRELATE WITH THE PATIENT¿S HISTORY. THE CUSTOMER RERAN THE SAMPLE ON ANOTHER INSTRUMENT AND GENERATED AN ESTRADIOL RESULT OF 1400 PG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12475651 ↗
  • MalfunctionJul 8, 2019Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY DECREASED ALINITY I ESTRADIOL RESULT OF 352.5 NG/ML ON A FEMALE PATIENT THAT WAS QUESTIONED BY THE PHYSICIAN. RESULTS PROVIDED: ON (B)(6) 2019 (PATIENT STIMULATED IN VITRO) = 1259.3 PG/ML; SID (B)(6), (B)(6) 2019 = 352.5 / >1000 / 2801.2 / 2914.9 PG/ML; (B)(6) 2019 = >3000 PG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8768998 ↗

FDA clearance and classification

The FDA regulates product code CHP (radioimmunoassay, estradiol) under 21 CFR 862.1260 as a Class I device. 21 CFR 862.1260 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alinity 07P5030

Has the Alinity 07P5030 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740129927), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Alinity 07P5030?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jul 8, 2019 and Mar 11, 2024. In those reports, the most common specific device problem is “Low Test Results” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alinity 07P5030?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Alinity 07P5030 FDA approved?

Devices of this type (product code CHP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott Ireland.

Who makes the Alinity 07P5030, and is it still sold?

The FDA's device database lists Abbott Ireland as the company responsible for it (brand Alinity, model 07P5030). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other RADIOIMMUNOASSAY, ESTRADIOL devices

All RADIOIMMUNOASSAY, ESTRADIOL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page