Alinity 07P6230Recalls & reported problems
Abbott Ireland Diagnostics · Alinity i CEA Reagent Kit 1200 Tests · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740130862), received between Aug 16, 2023 and Jun 17, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I CEA RESULT ON ONE 63 YEAR OLD MALE PATIENT DIAGNOSED WITH ASTHMA/COPD/PULMONARY FIBROSIS/HISTORY OF SMOKING. THE PATIENT HAS NOT BEEN DIAGNOSED WITH CANCER. THE FOLLOWING DATA WAS PROVIDED: ALINITY RESULT PROCESSED ON (B)(6) 2024 = 46.31 NG/ML HISTORICAL RESULTS: (B)(6) 2024 = 40.9 NG/ML (B)(6) 2024 = 31.4 NG/ML (B)(6) 2021 = 16.8 NG/ML THE CUSTOMER REPORTED THAT A CHEST/ABDOMINAL CT SCAN(WITHOUT CONTRAST) AND A UPPER/LOWER DIGESTIVE ENDOSCOPY WAS PERFORMED DUE TO THE FALSELY ELEVATED ALINITY I CEA RESULT OF 40.9 NG/ML, BUT NO MALIGNANT TUMOR WAS FOUND. THE CUSTOMER STATED THERE WAS NO HARM TO THE PATIENT. NO FURTHER IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19547939 ↗MalfunctionAug 16, 2023Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DECREASED ALINITY I CEA RESULT FOR A CANCER PATIENT. THE FOLLOWING DATA WAS PROVIDED: SAMPLE ID 406218372 INITIAL RESULT, ON (B)(6) 2023, WAS <2 UG/L. THE PATIENT WAS REDRAWN AND TESTED ON (B)(6) 2023, AND THE RESULT WAS 381 UG/L. THE PATIENT WAS REDRAWN AND TESTED ON (B)(6) 2023, AND THE RESULT WAS 321 UG/L. THE PATIENT¿S PREVIOUS RESULT, ON (B)(6) 2023, WAS 193 UG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17560295 ↗
Other SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN devices
- ADVIA Centaur® CEA 10309977Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN14 reports
- ARCHITECT 07K68-27Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN13 reports
- Elecsys CEA 07027079190Roche Diagnostics GmbH · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN13 reports
- Atellica IM CEA 10995523Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN11 reports
- ADVIA Centaur® CEA 10309978Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN10 reports
- IMMULITE® 2000 Systems CEA 10380995Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN9 reports
- ARCHITECT 07K68-35Abbott Ireland Diagnostics · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN8 reports
- Elecsys Anti-CCP 07251670190Roche Diagnostics GmbH · SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN8 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.