Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00380740145262), received between Jun 20, 2019 and Nov 8, 2019 — classified as 12 malfunctions.
The most frequently reported specific device problem is “High Test Results” (12 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- May 24, 2019Class IIITerminated
Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
Z-2079-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- High Test Results12
Patient problems
- No Patient Involvement12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 8, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED ABBOTT CONTROLS SHIFTING LOW WHILE USING ARCHITECT BNP CONTROL LOT 44K82818. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 9296060 ↗MalfunctionSep 20, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED ABBOTTT CONTROLS LOT 44K81118 AND BIORAD CONTROLS SHIFTING HIGH WHILE USING THE ARCHITECT BNP CALIBRATOR LOT 44K81018. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 9100659 ↗MalfunctionAug 7, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED A SHIFT IN CONTROLS WHILE USING ARCHITECT BNP CONTROLS LOT 44K79518. THE CUSTOMER INDICATED THEY MAY HAVE HAD FALSE RESULTS REPORTED HOWEVER HAVE HAD NO RESULTS QUESTIONED BY THE PHYSICIAN. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 8869427 ↗MalfunctionAug 7, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED A SHIFT IN CONTROLS WHILE USING ARCHITECT BNP CONTROLS LOT 44K81118. THE CUSTOMER INDICATED THEY MAY HAVE HAD FALSE RESULTS REPORTED HOWEVER HAVE HAD NO RESULTS QUESTIONED BY THE PHYSICIAN. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 8869425 ↗MalfunctionJul 16, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED ABBOTT CONTROLS OUT OF RANGE WHILE USING ARCHITECT BNP CONTROL LOT 44K82818. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8796754 ↗MalfunctionJul 16, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED ABBOTT CONTROLS OUT OF RANGE WHILE USING ARCHITECT BNP CONTROL LOT 44K79518. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8796753 ↗MalfunctionJul 16, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED ABBOTT CONTROLS SHIFT HIGH AND OUT OF RANGE WHILE USING ARCHITECT BNP CONTROL LOT 44K82818. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8796741 ↗MalfunctionJul 14, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED ABBOTT CONTROLS SHIFTING LOW AND OUT OF RANGE WHILE USING THE ARCHITECT BNP CONTROL LOT NUMBER 44K81118. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8787631 ↗MalfunctionJul 1, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED SURVEY RESULTS AND CONTROLS SHIFTED HIGH AND OUT OF RANGE WHILE USING ARCHITECT BNP CONTROL LOT 44K81118. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8749433 ↗MalfunctionJun 28, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED ABBOTT CONTROLS OUT OF RANGE WHILE USING ARCHITECT CONTROL LOT 44K81118. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8745483 ↗MalfunctionJun 28, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED BIORAD CONTROLS SHIFT HIGH ON THE ARCHITECT BNP ASSAY. THE CUSTOMER INDICATED THE ASSAY WAS CALIBRATED WITH CONTROL LOT 44K82018. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8743585 ↗MalfunctionJun 20, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED ABBOTT ARCHITECT BNP CONTROL LOT 44K82818 SHIFTED UP AFTER CALIBRATION. THERE WAS NO REPORT OF INCORRECT PATIENT RESULTS OR IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Patient Involvement
MAUDE report 8717455 ↗
FDA clearance and classification
The FDA regulates product code JJX (single (specified) analyte controls (assayed and unassayed)) under 21 CFR 862.1660 as a Class I device. 21 CFR 862.1660 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ARCHITECT 08K2812
Has the ARCHITECT 08K2812 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2079-2019 (Class III), began on May 24, 2019: Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ARCHITECT 08K2812?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Jun 20, 2019 and Nov 8, 2019. In those reports, the most common specific device problem is “High Test Results” (12). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARCHITECT 08K2812?
None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARCHITECT 08K2812 FDA approved?
Devices of this type (product code JJX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott GmbH.
Who makes the ARCHITECT 08K2812, and is it still sold?
The FDA's device database lists Abbott GmbH as the company responsible for it (brand ARCHITECT, model 08K2812). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Single (specified) analyte controls (assayed and unassayed) devices
- CONTOUR® NEXT 7314Ascensia Diabetes Care US Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)227 reports
- FastPack® IP SHBG FastPack® IP SHBG 30 ImmunoassayQualigen Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)2 recalls0 reports
- CONTOUR® NEXT 7377Ascensia Diabetes Care US Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)95 reports
- ARCHITECT 7K62-25Abbott Ireland · SINGLE (SPECIFIED) ANALYTE CONTROLS (ASSAYED AND UNASSAYED)94 reports
- CONTOUR® NEXT 9697Ascensia Diabetes Care US Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)56 reports
- ACE Control Set 1667001Sentinel Ch. SPA · Single (Specified) Analyte Controls (Assayed And Unassayed)1 recall0 reports
- Eurotrol GlucoTrol-AQ 180.013.002Eurotrol B.V. · Single (Specified) Analyte Controls (Assayed And Unassayed)1 recall0 reports
- Atellica IM CKMB MCM 10995534Siemens Healthcare Diagnostics Inc. · Single (specified) analyte controls (assayed and unassayed)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page