Skip to content

BD BBL™ Sensi-Disc™ Piperacillin/Tazobactam 100/10 µg 231691Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Sensi-Disc™ Piperacillin/Tazobactam 100/10 µg · Class 2

All models of BD BBL Sensi-Disc Piperacillin/Tazobactam 100/10 µg (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382902316916), received between Apr 24, 2024 and Apr 24, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Inaccurate Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 8, 2024Class IIOpen, Classified

    There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

    Z-1037-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inaccurate Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2024Inaccurate Information; Inaccurate Information

    IT WAS REPORTED THAT THE PRODUCT INSERT FOR BD BBL¿ SENSI-DISC¿ PIPERACILLIN/TAZOBACTAM - 100/10 G CONTAINED INCORRECT BREAKPOINT INFORMATION FOR PIPERACILLIN/TAZOBACTAM. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19173661 ↗

Other SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices

All SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page