Skip to content

BD BBL™ Sensi-Disc™ Ceftazidime 30 µg 231633Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Sensi-Disc™ Ceftazidime 30 µg · Class 2

All models of BD BBL Sensi-Disc Ceftazidime - 30 µg (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 30382902316337), received between May 9, 2025 and May 9, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 8, 2024Class IIOpen, Classified

    There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

    Z-1023-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED PRIOR TO USING THE BD BBL¿ SENSI-DISC¿ CEFTAZIDIME 30 ¿G, THE USER NOTED THAT THE LABEL OF THE CARTRIDGE CONTAINING THE PRODUCT (INTERNAL LABELING) WAS DIFFERENT FROM THE SHELF PACKAGE LABEL (EXTERNAL LABEL). NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21997189 ↗

Other SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices

All SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page