Intergard IGW0028-30Recalls & reported problems
Intervascular Sas · Intergard Woven Straight Collagen Coated Vascular Graft · Class 2 · Implant
All models of Intergard (41)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00384401000242), received between Nov 18, 2015 and Jun 21, 2024 — classified as 3 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Flaked” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K984294); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Flaked4
- Defective Device3
- Use of Device Problem2
- Inaccurate Information1
- Labelling, Instructions for Use or Training Problem1
- Manufacturing, Packaging or Shipping Problem1
- Product Quality Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- No Patient Involvement1
- Thrombosis1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2024Labelling, Instructions for Use or Training Problem; Inaccurate Information; Labelling, Instructions for Use or Training Problem; Inaccurate Information
COMPLAINT #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19584852 ↗MalfunctionMay 24, 2023Flaked; Product Quality Problem; Flaked; Product Quality Problem
COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16992485 ↗InjuryJun 29, 2022Use of Device Problem; Defective Device; Use of Device Problem; Defective Device
IT WAS REPORTED TO INTERVASCULAR (COMPLAINT #(B)(4)) THAT DURING SURGERY, A GRAFT (IGW0028-30) WAS OPENED BUT ITS LENGTH APPEARED SHORTER THAN 30 CM. THEREFORE, ANOTHER GRAFT OF THE SAME PRODUCT REFERENCE (IGW0028-30) WAS ADDED TO THE FIRST ONE TO COMPLETE THE SURGERY. THE SURGERY WAS SIGNIFICANTLY DELAYED. IT WAS SHARED THAT THERE WAS A NOTICEABLE SHORTNESS WHEN THE PRODUCTS WERE OPENED DURING THE SURGERY. THERE IS NO INFORMATION ABOUT LOT/SERIAL NUMBER OF THE GRAFT USED AND PATIENT INFORMATION. SINCE THE ALLEGATION OF SHORTNESS INVOLVES BOTH IMPLANTED PRODUCTS, ANOTHER MDR IS BEING SUBMITTED REGARDING THE SECOND DEVICE THAT WAS IMPLANTED. PLEASE NOTE THAT DUE TO THIS EVENT, THE STOCKS OF THE HOSPITAL WERE CHECKED AFTER THE OPERATION AND ANOTHER PRODUCT OF THE SAME PRODUCT REFERENCE, ALLEGEDLY SHORTER THAN 30 CM, WAS IDENTIFIED. A COMPLAINT HAS BEEN OPENED IN ORDER TO HANDLE THE EVENT (COMPLAINT #(B)(4)) AND A MDR IS BEING SUBMITTED FOR THIS CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14855812 ↗MalfunctionJun 24, 2022Defective Device; Defective Device
IT WAS REPORTED TO INTERVASCULAR (COMPLAINT (B)(4)) THAT DURING A PREVIOUS SURGERY, IT WAS OBSERVED THAT THE LENGTH OF TWO GRAFTS (MODEL IGW0028-30) WERE NOTICEABLY SHORTER THAN 30 CM. AFTER THE OPERATION, DUE TO THIS EVENT, THE HOSPITAL STOCKS WERE CHECKED AND IT WAS OBSERVED THAT A GRAFT OF THE SAME REFERENCE (IGW0028-30) , IN STOCK WAS ALSO SHORTER THAN 30 CM. TWO MDRS HAVE BEEN SUBMITTED REGARDING THE TWO FIRST GRAFTS IMPLANTED (COMPLAINT (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14801078 ↗InjuryJun 24, 2022Use of Device Problem; Defective Device; Use of Device Problem; Defective Device
IT WAS REPORTED TO INTERVASCULAR (COMPLAINT #(B)(4)) THAT DURING SURGERY, A GRAFT (IGW0028-30) WAS OPENED BUT ITS LENGTH APPEARED SHORTER THAN 30 CM. THEREFORE, ANOTHER GRAFT OF THE SAME PRODUCT REFERENCE (IGW0028-30) WAS ADDED TO THE FIRST ONE TO COMPLETE THE SURGERY. THE SURGERY WAS SIGNIFICANTLY DELAYED. IT WAS SHARED THAT THERE WAS A NOTICEABLE SHORTNESS WHEN THE PRODUCTS WERE OPENED DURING THE SURGERY. THERE IS NO INFORMATION ABOUT LOT/SERIAL NUMBER OF THE GRAFT USED AND PATIENT INFORMATION. SINCE THE ALLEGATION OF SHORTNESS INVOLVES BOTH IMPLANTED PRODUCTS, ANOTHER MDR IS BEING SUBMITTED REGARDING THE SECOND DEVICE THAT WAS IMPLANTED. PLEASE NOTE THAT DUE TO THIS EVENT, THE STOCKS OF THE HOSPITAL WERE CHECKED AFTER THE OPERATION AND ANOTHER PRODUCT OF THE SAME PRODUCT REFERENCE, ALLEGEDLY SHORTER THAN 30 CM, WAS IDENTIFIED. A COMPLAINT HAS BEEN OPENED IN ORDER TO HANDLE THE EVENT (COMPLAINT #(B)(4)) AND A MDR IS BEING SUBMITTED FOR THIS CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14801055 ↗MalfunctionDec 17, 2020Flaked; Flaked
SEE INITIAL MFR REPORT 1640201-2020-00027. COMPLAINT #(B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11029142 ↗MalfunctionFeb 20, 2020Flaked; Flaked
(B)(4). AFTER THE GRAFT OPENED ON THE TABLE DURING SURGERY, THE DOCTOR DETERMINED THAT THE COLLAGEN STRUCTURE WAS PEELING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9733412 ↗MalfunctionJun 7, 2018Flaked; Flaked
THE INTERIOR LUMEN OF THE VASCULAR GRAFT HAD SOME TYPE OF DEBRIS IN IT UPON OPENING THE GRAFT.
Patient problems: No Patient Involvement
MAUDE report 7578307 ↗InjuryNov 18, 2015Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT THE GRAFT WAS IMPLANTED ON (B)(6) 2015, ALONG WITH AN AORTIC VALVE REPLACEMENT. THE WEEK OF (B)(6) 2015, SYMPTOMS OF THROMBOSIS STARTED. ON (B)(6) 2015, THE PATIENT UNDERWENT AN ECHOCARDIOGRAM WHERE SIGN OF THROMBOSIS WERE FOUND, AND POSSIBLY CAUSED BY THE GRAFT. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Thrombosis
MAUDE report 5232134 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K984294. One submission often covers many models, so check the product and lot details in each record.
- Z-1282-2022 · Apr 29, 2022 · TerminatedINTERGARD Woven Vascular Graft is a woven polyester graft coated with a bovine collagen. It has an external velour surface and non-velour inner surface. The device is available in
Other Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices
- Cardioroot HEWROOT0030Intervascular Sas · Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 recall2 deaths13 reports
- FLIXENE SLIDER GDS 25058Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall13 reports
- ADVANTA SLIDER GDS 22114Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall11 reports
- Cardioroot HEWROOT0032Intervascular Sas · Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 recall2 deaths10 reports
- FLIXENE SLIDER GDS 25129Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall9 reports
- FLIXENE SLIDER GDS 25137Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall8 reports
- ADVANTA SLIDER GDS 22222Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall8 reports
- FLIXENE SLIDER GDS 25138Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page