Skip to content

Intergard IGK0008-20Recalls & reported problems

Intervascular Sas · Intergard Knitted Straight Collagen Coated Vascular Graft · Class 2 · Implant

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00384401007586), received between Feb 26, 2026 and Mar 2, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Product Quality Problem” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K964625); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Product Quality Problem2

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 2, 2026Product Quality Problem

    (B)(4).

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24480598 ↗
  • InjuryFeb 26, 2026Product Quality Problem

    IT WAS REPORTED TO INTERVASCULAR THAT DURING IMPLANTATION OF INTERGARD STANDARD GRAFT BLEEDING OCCURRED. THE SURGEON HAD TO SUTURE THE GRAFT AT SEVERAL POINTS WHICH RESULTS IN THE BLEEDING STOPPING. COMPLAINT#(B)(4).

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24454616 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964625. One submission often covers many models, so check the product and lot details in each record.

  • Z-1427-2010 · Dec 9, 2009 · Terminated
    InterGard Knitted Collagen Coated Vascular prostheses, Catalog number: IGK1206. InterVascular SAS, France. Indicated for surgical repair, bypass or replacement of arteries in th
  • Z-1439-2010 · Dec 9, 2009 · Terminated
    InterGard Knitted Collagen Coated Vascular prostheses marketed under 510(k) # K964625; InterGard Woven Collagen Coated Vascular prostheses marketed under 510(k} #K970843 and K9842

Other Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices

All Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.