Skip to content

Cardioroot HEWROOT0034Recalls & reported problems

Intervascular Sas · Cardioroot Woven Collagen Coated Vascular Graft · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00384401013938), received between Jun 13, 2016 and Apr 14, 2017 — classified as 1 death, 1 malfunction.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Aug 12, 2026Open, Classified

    Product failed to meet the specified acceptance criteria for Integral Water Permeability (IWP).

    Z-3258-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1
  • Material Frayed1

Patient problems

  • Blood Loss1
  • Death1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2017Material Frayed; Material Frayed

    THE HOSPITAL REPORTED THAT DURING A CARDIOROOT GRAFT PROCEDURE, SURGEON EXPERIENCED SOME GRAFT FRAYING AT THE POINT WHERE HE USED SCISSORS TO TRIM THE GRAFT FOR LENGTH. NO NEGATIVE PATIENT OUTCOME WAS REPORTED. THE GRAFT WAS IMPLANTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6493157 ↗
  • DeathJun 13, 2016Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    DURING A TIRON-DAVID VASCULAR SURGERY PERFORMED FOR AORTIC DISSECTION, SUTURE HOLE BLEEDING STARTED AFTER DECLAMPING THE AORTA. THE STAFF TRIED TO STOP THE BLEEDING, AND AFTER THE PATIENT WAS TRANSFERRED IN INTENSIVE CARE UNIT. IT WAS REPORTED THAT THE PATIENT DIED IN ICU DUE TO HEMORRHAGE. THE GRAFT REMAINED IMPLANTED. NO INFORMATION REGARDING THE DATES OF IMPLANTATION AND PATIENT DEATH IS AVAILABLE AT THE DATE OF THIS REPORT.

    Patient problems: Blood Loss; Death; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 5719796 ↗

Other Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices

All Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page