Skip to content

Hemagard HGKAX1010Recalls & reported problems

Intervascular Sas · Hemagard Knitted Axillo-bifemoral Collagen Coated Vascular Graft · Class 2 · Implant

All models of Hemagard (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00384401014478), received between Aug 18, 2023 and Aug 22, 2023 — classified as 2 deaths.

The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K964625); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Puncture/Hole2
  • Off-Label Use2

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathAug 22, 2023Off-Label Use; Material Puncture/Hole; Off-Label Use; Material Puncture/Hole

    COMPLAINT #(B)(4).

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17597084 ↗
  • DeathAug 18, 2023Off-Label Use; Material Puncture/Hole; Off-Label Use; Material Puncture/Hole

    IT WAS REPORTED TO INTERVASCULAR BY THE HOSPITAL THAT A PATIENT PRESENTED WITH ST SEGMENT ELEVATION MYOCARDIAL INFARCTION (STEMI) AND HAD AN INTRA-AORTIC BALLOON PUMP (IABP) PLACED. PLAN WAS TO INITIATE VENO-ARTERIAL EXTRACORPOREAL MEMBRANE OXYGENATION (VA ECMO) THEN HAVE CORONARY STENT PLACEMENT. PROCEDURES PERFORMED IN CATHERIZATION LABORATORY. RIGHT GROIN WITH 8FR CATHETER AND IABP BUT WAS OCCLUSIVE DUE TO SEVERE PERIPHERAL VASCULAR DISEASE (PVD). FEMORAL VENOUS ACCESS OBTAINED FOR ECMO AND LEFT AXILLARY CUTDOWN PERFORMED FOR ARTERIAL ACCESS. END TO SIDE ANASTOMOSIS PERFORMED TO LEFT AXILLARY ARTERY AND HEMAGARD GRAFT. IMPELLA INSERTED INTO 1 LIMB WITH 3/8-3/8 CONNECTOR TO OTHER LIMB AND THEN TO ECMO. FLOW RATE WAS 3.5 L/MIN WITH SYSTEM PRESSURE OF 220-230 MMHG. PINHOLE BLEEDING OCCURRED ALONG GRAFT, SUTURE LINES, AND SEAM OF BRANCHED GRAFT. HEMOSTATIC AGENTS USED WITHOUT DESIRED RESULTS WHICH INCLUDED EVARREST, EVASEAL, NU-KNIT, AND QUICK CLOT. ACTIVATED CLOTTING TIME (ACT) WAS REPORTED TO BE 140'S AND NO HEPARIN WAS GIVEN DURING THE PROCEDURE. PATIENT WITH HISTORY OF TAKING ELLIQUIS FOR ATRIAL FIBRILLATION. IMPELLA WAS REMOVED WITH CLAMPING OF GRAFT LIMB WITHOUT RESOLUTION OF

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17571637 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964625. One submission often covers many models, so check the product and lot details in each record.

  • Z-1427-2010 · Dec 9, 2009 · Terminated
    InterGard Knitted Collagen Coated Vascular prostheses, Catalog number: IGK1206. InterVascular SAS, France. Indicated for surgical repair, bypass or replacement of arteries in th
  • Z-1439-2010 · Dec 9, 2009 · Terminated
    InterGard Knitted Collagen Coated Vascular prostheses marketed under 510(k) # K964625; InterGard Woven Collagen Coated Vascular prostheses marketed under 510(k} #K970843 and K9842

Other Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices

All Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page