Skip to content

Dexcom G5 Mobile IOS App SW10611Recalls & reported problems

Dexcom, Inc. · G5 Mobile IOS App Stand Alone Software · Class 3

Quick summary

This is the Dexcom G5 mobile app for iOS. Its reports mostly describe lost signal between the app and the transmitter, unexpected app shutdowns and crashes, and the device not working as expected. There are 54,544 reports, with no deaths and 103 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54,544
Deaths reported
0
Injuries reported
103
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54,544 adverse event reports that list this exact device (DI 00386270000002), received between Oct 27, 2015 and Apr 21, 2021 — classified as 103 injuries, 54,441 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (49,322 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Wireless Communication Problem49,322
  • Use of Device Problem10,417
  • Device Inoperable1,893
  • Device Operates Differently Than Expected1,057
  • Unintended Application Program Shut Down986
  • Display or Visual Feedback Problem283
  • Connection Problem201
  • No Audible Alarm201

Patient problems

  • No Consequences Or Impact To Patient54,025
  • No Known Impact Or Consequence To Patient425
  • Hypoglycemia63
  • Loss of consciousness23
  • Seizures16
  • Hyperglycemia8
  • Injury7
  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 25 most recent of 54,544 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2021Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT AN APP CRASH OCCURRED. NO DATA OR PRODUCT WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11702415 ↗
  • MalfunctionDec 22, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. IT WAS DETERMINED THAT THE LOSS OF CONNECTION WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOSS OF CONNECTION WAS FOUND WITHIN THE INVESTIGATION WINDOW. HOWEVER, THE DEVICE OPERATED WITHIN SPECIFICATION. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11057524 ↗
  • MalfunctionDec 18, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE THE MOBILE DEVICE LOST POWER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11042195 ↗
  • MalfunctionDec 8, 2020Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT AN UNEXPECTED CGM APP SHUT-OFF OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE POTENTIAL PATIENT MISUSE AS THE APP WAS MANUALLY CLOSED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10973344 ↗
  • MalfunctionDec 1, 2020Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT AN APP CRASH ALERT OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE AN SQL ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10925725 ↗
  • MalfunctionNov 13, 2020Operating System Version or Upgrade Problem; Operating System Version or Upgrade Problem

    IT WAS REPORTED THAT MOBILE APP INTERRUPTION DUE TO OS UPGRADE ON SMART DEVICE OCCURRED. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10840933 ↗
  • MalfunctionNov 3, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10780204 ↗
  • MalfunctionNov 3, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10779318 ↗
  • MalfunctionSep 21, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE SYSTEM DIDN'T ALERT ON LOW OCCURRED ON APP. NO DATA OR PRODUCT WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE COMPLAINT WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 10561514 ↗
  • MalfunctionSep 14, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A BLUETOOTH CONNECTION ISSUE BETWEEN TRANSMITTER AND MOBILE APP OCCURRED.IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS RETURNED BUT WILL NOT CONFIRM THE ISSUE OR DETERMINE A PROBABLE CAUSE. THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10532193 ↗
  • MalfunctionSep 1, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10475877 ↗
  • MalfunctionAug 24, 2020Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT AN APP CRASH OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10446071 ↗
  • MalfunctionAug 17, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE LOSS OF CONNECTION WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOSS OF CONNECTION WAS FOUND WITHIN THE INVESTIGATION WINDOW. HOWEVER, THE DEVICE OPERATED WITHIN SPECIFICATION. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10418206 ↗
  • MalfunctionAug 17, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE LOSS OF CONNECTION WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOSS OF CONNECTION WAS FOUND WITHIN THE INVESTIGATION WINDOW. HOWEVER, THE DEVICE OPERATED WITHIN SPECIFICATION. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10415118 ↗
  • MalfunctionAug 17, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE PATIENT CLOSED THE MOBILE APP. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10413063 ↗
  • MalfunctionAug 4, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE PATIENT CLOSED THE MOBILE APP. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10364596 ↗
  • MalfunctionJul 14, 2020Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT THE APP SHUTTING DOWN OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE THE APP UNEXPECTEDLY SHUTTING OFF, OF WHICH THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10271850 ↗
  • MalfunctionJul 10, 2020No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT THERE WAS NO AUDIO ALERT ON THE MOBILE APP OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE DEVICE FUNCTIONED WITHIN SPECIFICATIONS AND PATIENT SETTINGS. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10261260 ↗
  • MalfunctionJul 10, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT AN UNKNOWN DURATION OF SIGNAL LOSS ON THE MOBILE APP OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO COMPLETE A DATA TRANSFER. THE REPORTED EVENT OF AN UNKNOWN DURATION OF SIGNAL LOSS IS REPORTABLE BASED ON THE FINDING OF SIGNAL LOSS OVER ONE HOUR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10261158 ↗
  • MalfunctionJul 7, 2020Alarm Not Visible; Alarm Not Visible

    IT WAS REPORTED THAT NO LOW ALERT ON THE MOBILE APP OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10240853 ↗
  • MalfunctionJul 7, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE LOSS OF CONNECTION WAS RELATED TO THE MOBILE APPLICATION. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOSS OF CONNECTION WAS FOUND WITHIN THE INVESTIGATION WINDOW. HOWEVER, THE DEVICE OPERATED WITHIN SPECIFICATION. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10240413 ↗
  • MalfunctionJul 7, 2020No Audible Alarm; Alarm Not Visible; No Audible Alarm; Alarm Not Visible

    IT WAS REPORTED THAT NO AUDIO ON THE MOBILE APP OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE DEVICE FUNCTIONED WITHIN SPECIFICATIONS AND PATIENT SETTINGS. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10240333 ↗
  • MalfunctionJul 1, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE LOSS OF CONNECTION WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOSS OF CONNECTION WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10218285 ↗
  • MalfunctionJun 25, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10197695 ↗
  • MalfunctionJun 24, 2020Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT AN UNEXPECTED CGM APP SHUT-OFF OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10192457 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.