Dexcom G4 Platinum Sensor 9500-27Recalls & reported problems
Dexcom, Inc. · Sensor Applicator Assembly · Class 3
Quick summary
This is a Dexcom G4 Platinum sensor. Its reports mostly describe imprecise readings, early sensor expiration, sensor wires coming loose and glucose values that differ from blood glucose meter readings. There are 99,426 reports, including 2 deaths and 2,966 injuries. Both recalls are closed: they concern a cancer drug, hydroxyurea, which may make glucose readings falsely high.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 99,426
- Deaths reported
- 2
- Injuries reported
- 2,966
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 99,426 adverse event reports that list this exact device (DI 00386270000019), received between Aug 10, 2015 and Aug 23, 2021 — classified as 2 deaths, 2,966 injuries, 96,456 malfunctions.
The most frequently reported specific device problem is “Imprecision” (48,033 reports).
2 FDA recall records name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Feb 3, 2020Terminated
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
Z-1560-2020 · Root cause: Under Investigation by firm · FDA record ↗
- Feb 3, 2020Terminated
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
Z-1561-2020 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Imprecision48,033
- Invalid Sensing29,596
- Use of Device Problem24,696
- Device Operates Differently Than Expected10,491
- Detachment of Device or Device Component2,531
- Detachment Of Device Component2,307
- Component Missing2,273
- Adverse Event Without Identified Device or Use Problem1,792
Patient problems
- No Consequences Or Impact To Patient93,354
- No Known Impact Or Consequence To Patient3,045
- Reaction991
- Erythema969
- Hypoglycemia674
- Itching Sensation672
- Rash506
- Loss of consciousness317
Recent adverse event reports
The 25 most recent of 99,426 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 23, 2021Material Separation; Material Separation
IT WAS REPORTED THAT THE SENSOR WIRE BECAME DETACHED FROM THE SENSOR. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT SENSOR WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12355392 ↗MalfunctionMay 1, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE COUNTS ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11758628 ↗MalfunctionApr 25, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE COUNTS ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11721999 ↗MalfunctionApr 25, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE COUNTS ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11721967 ↗MalfunctionApr 23, 2021Imprecision; Imprecision
IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6)2021. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE B ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11721385 ↗MalfunctionApr 20, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT MISSING SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2021. NO PRODUCT WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11692941 ↗MalfunctionApr 19, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE COUNTS ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11686572 ↗MalfunctionApr 8, 2021Imprecision; Imprecision
IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6), 2021. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE C ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11638323 ↗MalfunctionApr 5, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DETACHED SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN ON (B)(6) 2021. NO PRODUCT WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11614247 ↗MalfunctionApr 1, 2021Failure to Sense; Failure to Sense
IT WAS REPORTED THAT A SENSOR EXPIRATION ALERT OCCURRED PREMATURELY. THERE WAS NO REPORTED ADVERSE IMPACT. THE SENSOR WAS REPLACED TO RESOLVE THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11605421 ↗MalfunctionMar 18, 2021Imprecision; Imprecision
IT WAS REPORTED THAT THERE WAS AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) READING AND THE BLOOD GLUCOSE (BG) METER READING. BG AND CGM VALUES WERE NOT PROVIDED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. MULTIPLE ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER ON THE REPORTED ISSUE, BUT NO RESPONSE WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11517545 ↗MalfunctionMar 15, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE POTENTIAL PATIENT MISUSE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11490281 ↗InjuryMar 9, 2021Imprecision; Imprecision
THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE B ZONE OF THE PARKES ERROR GRID.
Patient problems: Hypoglycemia; Loss of consciousness
MAUDE report 11445169 ↗MalfunctionFeb 18, 2021Imprecision; Imprecision
IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2021. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE B ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11345544 ↗MalfunctionFeb 16, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE RESIDUAL ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11335804 ↗MalfunctionFeb 12, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE POTENTIAL PATIENT MISUSE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11325671 ↗MalfunctionFeb 8, 2021Imprecision; Imprecision
IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ARM ON (B)(6) 2021. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALLS WITHIN THE B ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11297316 ↗MalfunctionFeb 3, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE COUNTS ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11276099 ↗MalfunctionFeb 2, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DETACHED SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN ON (B)(6) 2021. NO PRODUCT WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11266316 ↗MalfunctionFeb 1, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DETACHED SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE UPPER BUTTOCKS ON (B)(6) 2021. NO PRODUCT WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11262984 ↗MalfunctionFeb 1, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DETACHED SENSOR WIRE OCCURRED. THE SENSOR WAS INSERTED INTO THE UPPER BUTTOCKS ON (B)(6) 2021. NO PRODUCT WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11262943 ↗MalfunctionJan 29, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE POTENTIAL PATIENT MISUSE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11255510 ↗MalfunctionJan 25, 2021Imprecision; Imprecision
IT WAS REPORTED THAT THERE WAS AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) READING AND THE BLOOD GLUCOSE (BG) METER READING. REPORTEDLY, THE CGM BG READING WAS 40 MG/DL, AND THE METER BG READING WAS 353 MG/DL. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT SENSOR WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11224232 ↗MalfunctionJan 24, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE COUNTS ABERRATION VIA DATA. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11220987 ↗MalfunctionJan 20, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11204785 ↗
Other Sensor, Glucose, Invasive devices
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
- Dexcom G5 Mobile Receiver MT22719-PNKDexcom, Inc. · Sensor, Glucose, Invasive1 death6,048 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.