Dexcom G5 Transmitter Kit STT-GF-001Recalls & reported problems
Dexcom, Inc. · G5 Transmitter Kit · Class 3
Quick summary
This is a Dexcom G5 transmitter kit. Its reports mostly describe transmitter failure errors, some traced to a low battery, along with lost wireless signal and no device output. There are 58,932 reports, including 1 death and 60 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 58,932
- Deaths reported
- 1
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 58,932 adverse event reports that list this exact device (DI 10386270000221), received between Jan 21, 2016 and Jan 26, 2023 — classified as 1 death, 60 injuries, 58,871 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (26,789 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Wireless Communication Problem26,789
- Device Displays Incorrect Message17,049
- No Device Output7,828
- Premature End-of-Life Indicator2,171
- Low Battery1,592
- Communication or Transmission Problem1,428
- Device Operates Differently Than Expected681
- Imprecision449
Patient problems
- No Consequences Or Impact To Patient53,012
- No Known Impact Or Consequence To Patient5,280
- No Clinical Signs, Symptoms or Conditions607
- Hypoglycemia27
- Loss of consciousness16
- Seizures7
- Hyperglycemia6
- Malaise6
Recent adverse event reports
The 25 most recent of 58,932 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2023No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS OFF-LABEL USAGE OF THE DEVICE. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16247002 ↗MalfunctionJan 5, 2023No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16106066 ↗MalfunctionDec 30, 2022No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16080857 ↗MalfunctionDec 14, 2022No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15980170 ↗MalfunctionOct 26, 2022Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15677602 ↗MalfunctionSep 16, 2022Connection Problem; Connection Problem
IT WAS REPORTED THAT A PAIRING FAILURE OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE SIGNAL LOSS. IN ADDITION, THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. THE REPORTED EVENT OF A PAIRING FAILURE IS REPORTABLE BASED ON THE FINDING OF THE SIGNAL LOSS OVER ONE HOUR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15437916 ↗MalfunctionAug 26, 2022No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15307439 ↗MalfunctionAug 18, 2022Connection Problem; Connection Problem
IT WAS REPORTED THAT A PAIRING FAILURE OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A FAILED TRANSMITTER ERROR. THE REPORTED EVENT OF A PAIRING FAILURE IS REPORTABLE BASED ON THE FINDING OF A TRANSMITTER FAILED ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15254438 ↗MalfunctionApr 14, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14112840 ↗MalfunctionMar 17, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13802413 ↗MalfunctionMar 11, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13731354 ↗MalfunctionMar 10, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13723623 ↗MalfunctionMar 1, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13633007 ↗MalfunctionFeb 15, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13536584 ↗MalfunctionFeb 11, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13515063 ↗MalfunctionFeb 2, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13423931 ↗MalfunctionJan 12, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13236165 ↗MalfunctionJan 10, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13205952 ↗MalfunctionDec 22, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13064192 ↗MalfunctionDec 13, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12986595 ↗MalfunctionNov 23, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12862786 ↗MalfunctionNov 9, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12780769 ↗MalfunctionSep 9, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12448747 ↗MalfunctionSep 8, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN INVALID TRANSMITTER ID ALERT. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE STARTED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12439816 ↗MalfunctionAug 27, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12378404 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
- Dexcom G5 Mobile Receiver MT22719-PNKDexcom, Inc. · Sensor, Glucose, Invasive1 death6,048 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.