Skip to content

Trilogy U1040005BRecalls & reported problems

Respironics, Inc. · Trilogy 200 Ventilator with Bluetooth, Recertified, USA · Class 2

All models of Philips Trilogy Ventilator (90) · Trilogy (46)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00606959028876), received between Apr 30, 2025 and Feb 19, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Software Problem1
  • Failure to Calibrate1
  • Failure to Power Up1
  • Improper Flow or Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2026Computer Software Problem

    THE TRILOGY 200 VENTILATOR WAS RETURNED TO THE MANUFACTURER SERVICE CENTER FOR RECERTIFICATION AND PREVENTATIVE MAINTENANCE. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE OCCURRENCE. DURING THE EVALUATION, IT WAS NOTED THAT THE DEVICE RETURNED TO THE TECHNICIAN FOR ADDITIONAL EVALUATION DUE TO A FAILED REPAIR TEST (POS. FLOW VERIFY). THE DEVICE WAS RECALIBRATED TO RESOLVE THE ISSUE. THERE WERE NO SIGNIFICANT EVENT(S) IN THE ERROR LOG(S). ADDITIONALLY, DURING THE SERVICE OF THE DEVICE, COMPONENTS WERE ONLY REPLACED AS PART OF MAINTENANCE OF THE DEVICE. THE DEVICE PASSED ALL TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24388200 ↗
  • MalfunctionFeb 4, 2026Failure to Calibrate; Improper Flow or Infusion

    A TRILOGY 200 VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO SERIOUS PATIENT HARM OR INJURY THAT OCCURRED OR WAS REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED THE POSITIVE FLOW VERIFY TEST STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROLLER WAS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, UNRELATED TO THE TEST FAILURE, DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE WAS FOUND TO HAVE MISSING/DISPLACED RUBBER FEET. FEET WERE REPLACED TO ADDRESS THE ISSUE. ALSO, DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED THE NON-REPORTABLE INTERNAL BATTERY CAPACITY TEST STEP DURING TESTING. THE INTERNAL BATTERY PACK WAS CHARGED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24260202 ↗
  • MalfunctionApr 30, 2025Failure to Power Up; Failure to Power Up

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A TRILOGY 200 FAILED INITIAL POWER UP. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER¿S COMPLAINT WAS CONFIRMED. THE DEVICE'S POWER MANAGEMENT BOARD REQUIRED REPLACEMENT TO ADDRESS THE ISSUE. HOWEVER, NO REPAIRS WERE COMPLETED BECAUSE THE DEVICE WAS NOT NEEDED FOR INVENTORY AND WAS SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21935872 ↗

FDA clearance and classification

The FDA regulates product code NOU (continuous, ventilator, home use) under 21 CFR 868.5895 as a Class II device. 21 CFR 868.5895 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

FDA definition of this device type: This product code was needed for the home use indication. CBK is not a tracked device and NOU is a tracked device.

Common questions about the Trilogy U1040005B

Has the Trilogy U1040005B been recalled?

No FDA recall record in our data names this exact device identifier (DI 00606959028876), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Trilogy U1040005B?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Apr 30, 2025 and Feb 19, 2026. In those reports, the most common specific device problems are “Computer Software Problem” (1), “Failure to Calibrate” (1) and “Failure to Power Up” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Trilogy U1040005B?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Trilogy U1040005B, and is it still sold?

The FDA's device database lists Respironics, Inc. as the company responsible for it (brand Trilogy, model U1040005B). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Continuous, ventilator, home use devices

All Continuous, ventilator, home use devices →