Trilogy 100 Trilogy 100 Ventilator, Bluetooth, AustraliaRecalls & reported problems
Respironics, Inc. · Trilogy 100 Ventilator, Bluetooth, Australia · Class 2
All models of Philips Trilogy Ventilator (90) · Trilogy 100 (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00606959038226), received between Aug 11, 2021 and Jul 3, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Failure to Calibrate” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (K083526); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Calibrate2
- Device Sensing Problem1
- Electrical /Electronic Property Problem1
- Failure to Charge1
- Pressure Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2025Device Sensing Problem; Device Sensing Problem
A VENTILATOR WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22396855 ↗MalfunctionMay 8, 2024Failure to Calibrate; Failure to Calibrate
THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE DEVICE HANGS UP DURING THE DISPLAYS REMOTE CALIBRATION CAUSING A FAILED SYSTEM TEST HAD OCCURRED ON THE DEVICE. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19270988 ↗MalfunctionMar 13, 2024Failure to Calibrate; Pressure Problem; Failure to Calibrate; Pressure Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A TRILOGY 100 VENTILATOR FAILED PVT AT THE HOSPITAL BIOMED WORK BENCH. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18893435 ↗MalfunctionJan 29, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A TRILOGY 100 VENTILATOR INOPERATIVE CONDITION OCCURRED. THE DEVICE REPORTEDLY BECAME NOISY ABOVE USUAL OPERATIONAL LEVEL, AND THE SCREEN WAS BLANK. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18594281 ↗MalfunctionAug 11, 2021Failure to Charge; Failure to Charge
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ERROR CODE RELATED TO THE CHARGING OF THE INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12307400 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083526. One submission often covers many models, so check the product and lot details in each record.
- Z-0882-2023 · Dec 7, 2022 · Open, ClassifiedPhilips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B, R1054260, RSL1054260, CA1054096, R1054260B, R1054655
- Z-0884-2023 · Dec 7, 2022 · Open, ClassifiedPhilips Garbin Ventilator, Model Number 1058180B
- Z-1956-2021 · Jun 14, 2021 · Open, ClassifiedTrilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.
- Z-1065-2016 · Jan 18, 2016 · TerminatedTrilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory su
- Z-1143-2014 · Feb 11, 2014 · TerminatedTrilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilog
- Z-2024-2012 · Apr 27, 2012 · TerminatedRespironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of individuals who require m
- Z-0296-2012 · Oct 11, 2011 · TerminatedTrilogy 100, 200 and 202 Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used fo
- Z-1896-2010 · Jun 10, 2010 · TerminatedTrilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for
Other CONTINUOUS, VENTILATOR, HOME USE devices
- Astral 150Resmed Pty Ltd · Continuous, Ventilator, Home Use1 recall17 deaths2,361 reports
- Trilogy EV300, USA DS2200X11BRespironics, Inc. · Continuous, ventilator, home use5 recalls2,071 reports
- Trilogy 00Respironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE15 deaths2,059 reports
- Trilogy Evo, Japan JP2110X16BRespironics, Inc. · Continuous, ventilator, home use5 recalls27 deaths1,112 reports
- Astral 100Resmed Pty Ltd · Continuous, Ventilator, Home Use9 deaths865 reports
- Trilogy Trilogy Evo Ventilator, IntlRespironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE4 recalls3 deaths648 reports
- Life2000 MS-01-0118Welch Allyn, Inc. · Continuous, ventilator, home use10 recalls2 reports
- Trilogy Evo GB2110X15BRespironics, Inc. · Continuous, ventilator, home use5 recalls1 death141 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page