Capsure®SP Novus 559445Recalls & reported problems

Medtronic, Inc. · LEAD 559445 FPA BI SIL ATR STER US · Class 3 · Implant

All models of CapSureSP Novus (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Capsure SP Novus pacemaker lead is an implanted lead that carries pacing signals to the heart. Many of the 124 reports describe the lead failing to sense heartbeats or needing more energy to pace the heart, and some describe pocket infections. The reports include 1 death and 54 injuries. Reports fell to 6 in the last 12 months from 11 in the previous 12. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
126
Deaths reported
1
Injuries reported
53
Recalls naming it
0

Summary

The FDA's MAUDE database contains 126 adverse event reports that list this exact device (DI 00613994250230), received between Dec 9, 2015 and Sep 28, 2026 — classified as 1 death, 53 injuries, 72 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (54 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Under-Sensing54
  • High Capture Threshold24
  • Adverse Event Without Identified Device or Use Problem19
  • Over-Sensing15
  • Failure to Capture9
  • Low impedance9
  • Pacing Problem7
  • Decreased Sensitivity6

Patient problems

  • No Clinical Signs, Symptoms or Conditions65
  • No Known Impact Or Consequence To Patient30
  • Unspecified Infection16
  • No Consequences Or Impact To Patient6
  • Endocarditis2
  • Erosion2
  • Erythema2
  • Arrhythmia1

Recent adverse event reports

The 25 most recent of 126 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911330 ↗
  • MalfunctionSep 23, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited under and over sensing resulting in inappropriate pacing. It was also reported that the RA lead exhibited intermittent undersensing noted on stored episodes which resulted in false termination of Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes. Low P waves were also observed Both leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865937 ↗
  • MalfunctionSep 11, 2026Under-Sensing; High Capture Threshold

    It was reported that the Right Atrial (RA) lead exhibited high capture thresholds. It was also noted that atrial undersensing occurr ed during Atrial Tachycardia/Atrial Fibrillation (AT/AF). No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50725476 ↗
  • MalfunctionMay 25, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDER SENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD DISPLAYED UNDER SENSING. BOTH THE LEADS REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25268318 ↗
  • MalfunctionApr 16, 2026Failure to Capture; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POSSIBLE LOSS OF CAPTURE IN REFRACTORY PERIOD AND UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24902937 ↗
  • MalfunctionMar 24, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24681678 ↗
  • InjuryMar 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) PATIENT EXPERIENCED A BACTEREMIA INFECTION. THE DEVICE SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 24507553 ↗
  • MalfunctionFeb 9, 2026Pacing Problem; Under-Sensing

    IT WAS FURTHER REPORTED VIA FOLLOW-UP WITH THE CLINIC THAT THE CRTD THERAPIES WERE DEEMED APPROPRIATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24290287 ↗
  • InjurySep 19, 2025High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION DUE TO HIGH RIGHT ATRIAL (RA) LEAD THRESHOLDS. THE IPG WAS EXPLANTED AND REPLACED. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23094108 ↗
  • InjurySep 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED A POCKET INFECTION. THE PATIENT EXPERIENCED DISCOMFORT, EROSION, PAIN, AND REDNESS. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort; Erosion; Erythema; Medical device site infection; Pain

    MAUDE report 23032377 ↗
  • MalfunctionAug 18, 2025Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22833825 ↗
  • MalfunctionJul 25, 2025Decreased Sensitivity; Capturing Problem; Decreased Sensitivity; Capturing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A SUDDEN DECREASE IN LONGEVITY DUE TO AN INCREASE IN RIGHT VENTRICULAR (RV) LEAD THRESHOLD. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN INCREASE IN THRESHOLD AND DECREASE IN P WAVE MEASUREMENT. IT WAS NOTED THAT THE IPG EXHIBITED A FALSE LOW IMPEDANCE WARNING ON THE RA LEAD. THE IPG AND THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22608173 ↗
  • MalfunctionJun 16, 2025Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22209654 ↗
  • MalfunctionMay 30, 2025Under-Sensing; Decreased Sensitivity; Under-Sensing; Decreased Sensitivity

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POSSIBLE UNDERSENSING ON STORED ELECTROGRAMS (EGM). THE RA LEAD REMAINS IN USE. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROGRAMMED VVIR. IT WAS NOTED THAT THE RA LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22113731 ↗
  • MalfunctionMar 18, 2025High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD REACHED END OF SERVICE (EOS) AND STARTED ALERTING FOR CHARGE CIRCUIT TIMEOUT. IT WAS NOTED PER TRANSMISSION REPORT THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH ATRIAL THRESHOLD. THE LEAD REMAINS IN USE AND THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633602 ↗
  • MalfunctionJan 30, 2025Electrical /Electronic Property Problem; Under-Sensing; Electrical /Electronic Property Problem; Under-Sensing

    IT WAS FURTHER REPORTED THAT RA LEAD EXHIBITED POTENTIAL ATRIAL UNDER SENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF AT/AF EVENT(S) AT TIMES.

    Patient problems: Chest Pain; Dizziness; Presyncope

    MAUDE report 21273477 ↗
  • InjuryJan 17, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A CONFIRMED FRACTURE AND THE LEAD TIP BROKE OFF AND COMPLETELY SEPARATED FROM LEAD BODY. DUE TO POTENTIAL INADEQUATE LEAD SLACK, TENSION WAS PLACED ON THE LEAD WHEN THE PATIENT STOOD, PLACING TENSION ON LEAD BODY WHERE IT ATTACHED TO LEAD TIP AND OVER TIME CAUSED FRACTURE. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21177759 ↗
  • MalfunctionNov 22, 2024Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED GROSS UNDERSENSING NOTED ON STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20757304 ↗
  • MalfunctionNov 20, 2024Failure to Capture; Failure to Capture

    IT WAS REPORTED THAT THERE WAS POSSIBLE RIGHT ATRIAL (RA) LEAD CAPTURE ISSUES NOTED ON A STORED ELECTROGRAM (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20736406 ↗
  • MalfunctionOct 30, 2024Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT UNDERSENSING ON STORED ELECTROGRAMS (EGM) WITH CLASSIFIED TERMINATION OF ONGOING ATRIAL ARRHYTHMIA. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20565679 ↗
  • InjuryOct 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO AN INFECTION. A NEW SYSTEM WAS SUBSEQUENTLY IMPLANTED APPROXIMATELY ONE WEEK LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 20515694 ↗
  • MalfunctionOct 17, 2024Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING APPEARS TO RESULT IN INAPPROPRIATE ATRIAL PACING ON CURRENT ELECTROGRAMS (EGM). THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20471413 ↗
  • MalfunctionOct 8, 2024Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THERE WAS POSSIBLE RIGHT ATRIAL (RA) LEAD OVERSENSING OF FAR FIELD R-WAVE (FFRW) CAUSING FALSE DETECTION OF ATRIAL ARRHYTHMIA. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20402826 ↗
  • InjurySep 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A POSSIBLE INFECTION LESS THAN TWO WEEKS POST GENERATOR CHANGE. THERE WAS LOTS OF SWELLING AT THE DEVICE SITE AND IT WAS TENDER TO TOUCH. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 20123282 ↗
  • MalfunctionApr 11, 2024Low impedance; Low impedance

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD TRIGGERED AN ALERT DUE TO LOW OUT OF RANGE (OOR) PACING IMPEDANCE. IT WAS NOTED THAT THE MEASUREMENT AT THE TIME OF THE ALERT WAS 0 OHMS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19093277 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page