Capsure®SP Novus 559445Recalls & reported problems
Medtronic, Inc. · LEAD 559445 FPA BI SIL ATR STER US · Class 3 · Implant
All models of CapSureSP Novus (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Capsure SP Novus pacemaker lead is an implanted lead that carries pacing signals to the heart. Many of the 124 reports describe the lead failing to sense heartbeats or needing more energy to pace the heart, and some describe pocket infections. The reports include 1 death and 54 injuries. Reports fell to 6 in the last 12 months from 11 in the previous 12. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 126
- Deaths reported
- 1
- Injuries reported
- 53
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 126 adverse event reports that list this exact device (DI 00613994250230), received between Dec 9, 2015 and Sep 28, 2026 — classified as 1 death, 53 injuries, 72 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (54 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Under-Sensing54
- High Capture Threshold24
- Adverse Event Without Identified Device or Use Problem19
- Over-Sensing15
- Failure to Capture9
- Low impedance9
- Pacing Problem7
- Decreased Sensitivity6
Patient problems
- No Clinical Signs, Symptoms or Conditions65
- No Known Impact Or Consequence To Patient30
- Unspecified Infection16
- No Consequences Or Impact To Patient6
- Endocarditis2
- Erosion2
- Erythema2
- Arrhythmia1
Recent adverse event reports
The 25 most recent of 126 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Pacing Problem; Under-Sensing
It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50911330 ↗MalfunctionSep 23, 2026Under-Sensing
It was reported that the Right Ventricular (RV) lead exhibited under and over sensing resulting in inappropriate pacing. It was also reported that the RA lead exhibited intermittent undersensing noted on stored episodes which resulted in false termination of Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes. Low P waves were also observed Both leads remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865937 ↗MalfunctionSep 11, 2026Under-Sensing; High Capture Threshold
It was reported that the Right Atrial (RA) lead exhibited high capture thresholds. It was also noted that atrial undersensing occurr ed during Atrial Tachycardia/Atrial Fibrillation (AT/AF). No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50725476 ↗MalfunctionMay 25, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDER SENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD DISPLAYED UNDER SENSING. BOTH THE LEADS REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268318 ↗MalfunctionApr 16, 2026Failure to Capture; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POSSIBLE LOSS OF CAPTURE IN REFRACTORY PERIOD AND UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24902937 ↗MalfunctionMar 24, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24681678 ↗InjuryMar 4, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) PATIENT EXPERIENCED A BACTEREMIA INFECTION. THE DEVICE SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Unspecified Infection
MAUDE report 24507553 ↗MalfunctionFeb 9, 2026Pacing Problem; Under-Sensing
IT WAS FURTHER REPORTED VIA FOLLOW-UP WITH THE CLINIC THAT THE CRTD THERAPIES WERE DEEMED APPROPRIATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24290287 ↗InjurySep 19, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION DUE TO HIGH RIGHT ATRIAL (RA) LEAD THRESHOLDS. THE IPG WAS EXPLANTED AND REPLACED. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23094108 ↗InjurySep 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED A POCKET INFECTION. THE PATIENT EXPERIENCED DISCOMFORT, EROSION, PAIN, AND REDNESS. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Discomfort; Erosion; Erythema; Medical device site infection; Pain
MAUDE report 23032377 ↗MalfunctionAug 18, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833825 ↗MalfunctionJul 25, 2025Decreased Sensitivity; Capturing Problem; Decreased Sensitivity; Capturing Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A SUDDEN DECREASE IN LONGEVITY DUE TO AN INCREASE IN RIGHT VENTRICULAR (RV) LEAD THRESHOLD. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN INCREASE IN THRESHOLD AND DECREASE IN P WAVE MEASUREMENT. IT WAS NOTED THAT THE IPG EXHIBITED A FALSE LOW IMPEDANCE WARNING ON THE RA LEAD. THE IPG AND THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608173 ↗MalfunctionJun 16, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22209654 ↗MalfunctionMay 30, 2025Under-Sensing; Decreased Sensitivity; Under-Sensing; Decreased Sensitivity
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POSSIBLE UNDERSENSING ON STORED ELECTROGRAMS (EGM). THE RA LEAD REMAINS IN USE. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROGRAMMED VVIR. IT WAS NOTED THAT THE RA LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22113731 ↗MalfunctionMar 18, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD REACHED END OF SERVICE (EOS) AND STARTED ALERTING FOR CHARGE CIRCUIT TIMEOUT. IT WAS NOTED PER TRANSMISSION REPORT THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH ATRIAL THRESHOLD. THE LEAD REMAINS IN USE AND THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21633602 ↗MalfunctionJan 30, 2025Electrical /Electronic Property Problem; Under-Sensing; Electrical /Electronic Property Problem; Under-Sensing
IT WAS FURTHER REPORTED THAT RA LEAD EXHIBITED POTENTIAL ATRIAL UNDER SENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF AT/AF EVENT(S) AT TIMES.
Patient problems: Chest Pain; Dizziness; Presyncope
MAUDE report 21273477 ↗InjuryJan 17, 2025Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A CONFIRMED FRACTURE AND THE LEAD TIP BROKE OFF AND COMPLETELY SEPARATED FROM LEAD BODY. DUE TO POTENTIAL INADEQUATE LEAD SLACK, TENSION WAS PLACED ON THE LEAD WHEN THE PATIENT STOOD, PLACING TENSION ON LEAD BODY WHERE IT ATTACHED TO LEAD TIP AND OVER TIME CAUSED FRACTURE. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21177759 ↗MalfunctionNov 22, 2024Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED GROSS UNDERSENSING NOTED ON STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20757304 ↗MalfunctionNov 20, 2024Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THERE WAS POSSIBLE RIGHT ATRIAL (RA) LEAD CAPTURE ISSUES NOTED ON A STORED ELECTROGRAM (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20736406 ↗MalfunctionOct 30, 2024Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT UNDERSENSING ON STORED ELECTROGRAMS (EGM) WITH CLASSIFIED TERMINATION OF ONGOING ATRIAL ARRHYTHMIA. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20565679 ↗InjuryOct 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO AN INFECTION. A NEW SYSTEM WAS SUBSEQUENTLY IMPLANTED APPROXIMATELY ONE WEEK LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20515694 ↗MalfunctionOct 17, 2024Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING APPEARS TO RESULT IN INAPPROPRIATE ATRIAL PACING ON CURRENT ELECTROGRAMS (EGM). THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20471413 ↗MalfunctionOct 8, 2024Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THERE WAS POSSIBLE RIGHT ATRIAL (RA) LEAD OVERSENSING OF FAR FIELD R-WAVE (FFRW) CAUSING FALSE DETECTION OF ATRIAL ARRHYTHMIA. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20402826 ↗InjurySep 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A POSSIBLE INFECTION LESS THAN TWO WEEKS POST GENERATOR CHANGE. THERE WAS LOTS OF SWELLING AT THE DEVICE SITE AND IT WAS TENDER TO TOUCH. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20123282 ↗MalfunctionApr 11, 2024Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD TRIGGERED AN ALERT DUE TO LOW OUT OF RANGE (OOR) PACING IMPEDANCE. IT WAS NOTED THAT THE MEASUREMENT AT THE TIME OF THE ALERT WAS 0 OHMS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19093277 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page