CapSureSP NovusRecalls & reported problems across all models
Medtronic, Inc. · 14 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2,107
- Deaths reported
- 18
- Injuries reported
- 1,067
- Recalls
- 0
We found 14 models of CapSureSP Novus by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 2,107 adverse event reports in the FDA's MAUDE database, received between Dec 9, 2015 and Sep 28, 2026 — classified as 18 deaths, 1,067 injuries, 1,022 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (488 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CapSureSP Novus problems
Device problems
- Under-Sensing488
- Over-Sensing476
- High Capture Threshold440
- Adverse Event Without Identified Device or Use Problem306
- Signal Artifact/Noise184
- High impedance181
- Failure to Capture180
- Low impedance141
Patient problems
- No Clinical Signs, Symptoms or Conditions896
- No Known Impact Or Consequence To Patient522
- Unspecified Infection238
- No Consequences Or Impact To Patient89
- Bradycardia43
- Dizziness40
- Dyspnea38
- Syncope/Fainting38
CapSureSP Novus models
14 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Capsure®SP Novus 5092-58Permanent defibrillator electrodes3 deaths350 reports
- CapSure®SP Novus 4092-52permanent pacemaker Electrode1 death307 reports
- Capsure®SP Novus 5092-52Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes4 deaths294 reports
- CapSure®SP Novus 5592-45permanent pacemaker Electrode246 reports
- CapSure®SP Novus 4592-45Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes2 deaths237 reports
- CapSure®SP Novus 4092-58Permanent defibrillator electrodes3 deaths185 reports
- CapSure®SP Novus 4592-53permanent pacemaker Electrode1 death165 reports
- CapSure®SP Novus 5592-53Permanent defibrillator electrodes3 deaths140 reports
- Capsure®SP Novus 559445permanent pacemaker Electrode1 death126 reports
- Capsure®SP Novus 559453Permanent defibrillator electrodes48 reports
- CAPSURE®SP NOVUS 5592-45permanent pacemaker Electrode3 reports
- CAPSURE®SP NOVUS 5592-53Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes3 reports
- CAPSURE®SP NOVUS 5092-52Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes2 reports
- CAPSURE®SP NOVUS 5092-58permanent pacemaker Electrode1 report
Recent CapSureSP Novus adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Pacing Problem; Under-Sensing
It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50911330 ↗MalfunctionSep 23, 2026High impedance; High Capture Threshold
It was reported that the Right Ventricular (RV) lead exhibited high thresholds and out of range high impedance. The lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869117 ↗MalfunctionSep 23, 2026Under-Sensing
It was reported that the Right Ventricular (RV) lead exhibited under and over sensing resulting in inappropriate pacing. It was also reported that the RA lead exhibited intermittent undersensing noted on stored episodes which resulted in false termination of Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes. Low P waves were also observed Both leads remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865937 ↗MalfunctionSep 23, 2026Over-Sensing; Pacing Problem; Under-Sensing
It was reported that the Right Ventricular (RV) lead exhibited under and over sensing resulting in inappropriate pacing. It was also reported that the RA lead exhibited intermittent undersensing noted on stored episodes which resulted in false termination of Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes. Low P waves were also observed Both leads remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865909 ↗MalfunctionSep 18, 2026Signal Artifact/Noise
It was reported that the Right Atrial (RA) lead exhibited noise. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50815245 ↗MalfunctionSep 16, 2026Failure to Capture; High Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50794928 ↗MalfunctionSep 11, 2026Low impedance
It was reported that the Right Atrial (RA) lead exhibited low impedance. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50731455 ↗InjurySep 11, 2026Signal Artifact/Noise; Failure to Capture; Electrical /Electronic Property Problem; Fracture; High impedance; Over-Sensing; Pacing Problem
It was reported that the Right Ventricular (RV) lead exhibited loss of capture, a lead fracture, oversensing resulting in inhibition of pacing and misclassification of episode type, noise, a polarity switch and high impedance. The RV lead was capped and was replaced by a leadless Implantable Pulse Generator (IPG). No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50729766 ↗MalfunctionSep 11, 2026Under-Sensing; High Capture Threshold
It was reported that the Right Atrial (RA) lead exhibited high capture thresholds. It was also noted that atrial undersensing occurr ed during Atrial Tachycardia/Atrial Fibrillation (AT/AF). No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50725476 ↗MalfunctionSep 10, 2026Over-Sensing
It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50709137 ↗MalfunctionSep 8, 2026High Capture Threshold
It was reported that the Right Ventricular (RV) lead exhibited high thresholds. Additionally, it was noted that the patient experienced palpitations. The RV lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Palpitations
MAUDE report 50673461 ↗InjurySep 8, 2026Failure to Capture
It was reported that the patient experienced syncope. Upon device assessment it was noted that the Right Ventricular (RV) lead exhibited loss of capture. The RV lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Syncope/Fainting
MAUDE report 50663379 ↗MalfunctionSep 2, 2026Signal Artifact/Noise; Failure to Capture; Over-Sensing; Pacing Problem; Under-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50563085 ↗InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the Implantable Pulse Generator (IPG) system was explanted due to an infection. No further patient complications have been reported as a result of this event.
Patient problems: Unspecified Infection
MAUDE report 50451158 ↗InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced bacteremia. The Implantable Pulse Generator (IPG) system and previously capped Right Ventricular (RV) lead were explanted and replaced with a leadless IPG. No further patient complications have been reported as a result of this event.
Patient problems: Bacteremia
MAUDE report 50432878 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page