Skip to content

INTERSTIM 3093-28Recalls & reported problems

Medtronic, Inc. · LEAD 3093-28 TINED QUAD EXT ELECT SNS US · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994601926), received between Jan 8, 2019 and Jun 30, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

5 other recalls cover devices cleared under the same FDA submission (P970004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Difficult to Remove1

Patient problems

  • Device Embedded In Tissue or Plaque2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2025Difficult to Remove; Appropriate Device Problem Term/Code Not Available; Difficult to Remove; Appropriate Device Problem Term/Code Not Available

    INFORMATION WAS RECEIVED FROM A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT THEY THINK LEAD FRAGMENT THAT REMAINED FROM THEIR PREVIOUS SYSTEM HAD MOVED FROM WHERE IT WAS BECAUSE THEY FOUND IT AFTER TAKING A SHOWER. THEY NEED AN MREI AND DON'T KNOW WHAT TO DO BECAUSE THE ABANDONED LEAD IS NOT MRI COMPATIBLE. AGENT REVIEWED UPDATED MRI GUIDELINES FOR ABANDONED PRODUCT WITH PATIENT, PATIENT STATED THAT THEIR ABANDONED FRAGMENT IS ABOUT AN INCH. THE PATIENT WAS REDIRECTED TO THEIR HEALTHCARE PROVIDER TO FURTHER ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22356648 ↗
  • MalfunctionDec 10, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION. THE CALLER REPORTED THAT THE PATIENT HAD THE INS REMOVED, BUT THE DOCTOR WAS UNABLE TO REMOVE THE LEAD. THEY HAD AN ABANDONED LEAD STILL IMPLANTED. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED OR ANTICIPATED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9443733 ↗
  • MalfunctionJan 8, 2019Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION. IT WAS REPORTED BY A HEALTH CARE PROFESSIONAL (HCP) WHO HAD CALLED FOR MRI GUIDELINES THAT THE PATIENT HAD THE DEVICE EXPLANTED BUT THE TIP OF THE PLATINUM BASED ELECTRODE LEAD TIP WAS STUCK IN THE S3 FORAMEN. THERE WERE NO SYMPTOMS REPORTED AND IT WAS UNKNOWN WHEN THIS OCCURRED. THERE WERE NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8228528 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P970004. One submission often covers many models, so check the product and lot details in each record.

  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-0998-2022 · Feb 28, 2022 · Terminated
    MEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA,
  • Z-1147-2015 · Jan 9, 2015 · Terminated
    The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
  • Z-0040-2008 · Jun 25, 2007 · Terminated
    Medtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim
  • Z-0037-2008 · Jun 25, 2007 · Terminated
    Medtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implanta

Other Stimulator, electrical, implantable, for incontinence devices

All Stimulator, electrical, implantable, for incontinence devices →

Sources and updates