Advisa DR MRI™ SureScan® A2DR01 (UDI 00613994713650)Recalls & reported problems
Medtronic, Inc. · IPG A2DR01 ADVISA DR MRI US MR · Class 3 · Implant
All models of Advisa DR MRI SureScan (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Advisa DR MRI pacemaker is an implanted device that paces the heart. Of its 1,757 reports, the reports describe pacing problems, batteries running down early and the device failing to respond to checks, and include 41 deaths and 949 injuries. Some mention slow heart rates, infection and fainting. Reports fell from 109 in the prior 12 months to 58 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,757
- Deaths reported
- 41
- Injuries reported
- 949
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,757 adverse event reports that list this exact device (DI 00613994713650), received between Dec 9, 2013 and Sep 29, 2026 — classified as 41 deaths, 949 injuries, 767 malfunctions.
The most frequently reported specific device problem is “Pacing Problem” (326 reports).
16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem476
- Pacing Problem326
- Appropriate Device Problem Term/Code Not Available321
- Premature Discharge of Battery146
- Unable to Obtain Readings75
- Failure to Interrogate72
- Over-Sensing62
- Electromagnetic Interference50
Patient problems
- No Clinical Signs, Symptoms or Conditions434
- No Known Impact Or Consequence To Patient332
- Unspecified Infection308
- Bradycardia154
- Syncope/Fainting64
- Dizziness61
- No Consequences Or Impact To Patient43
- Pain43
Recent adverse event reports
The 25 most recent of 1,757 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Pacing Problem
The patient reported that after an Magnetic Resonance Imaging (MRI) they felt unwell. They reported that during the MRI there heart was pounding, then after they experienced lightheaded-ness, dizziness, fatigue and heart rates in the forties. The Implantable Pulse Generator (IPG) remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Bradycardia; Dizziness; Fatigue; Palpitations
MAUDE report 50932243 ↗MalfunctionSep 24, 2026Pacing Problem
It was reported that the patient experienced weakness and the Implantable Pulse Generator (IPG) was pacing below the lower set rate. The IPG remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Muscle Weakness/Atrophy
MAUDE report 50883354 ↗InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced endocarditis. The Implantable Pulse Generator (IPG) was explanted and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Endocarditis
MAUDE report 50751043 ↗InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient felt the pacing and pulsing in their neck and it wakes them up at night. It was also reported that the Right Ventricular (RV) lead exhibited steadily increasing impedance resulting in high impedance causing them to be unable to have Magnetic Resonance Imaging (MRI) scan. Both the Implantable Pulse Generator (IPG) and the RV lead remains in use. No further patient complications have been reported as a result of this event. It was further reported that the RV lead exhibited polarity switch. The RV lead was reprogrammed and the IPG was explanted and replaced.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50688933 ↗InjurySep 1, 2026Electrical /Electronic Property Problem
It was further reported that the Implantable Pulse Generator (IPG) was explanted and replaced.
Patient problems: Chest Pain
MAUDE report 50542949 ↗MalfunctionAug 12, 2026Pacing Problem
Patient problems: Bradycardia
MAUDE report 50222462 ↗MalfunctionAug 10, 2026Pacing Problem
It was reported that the Right Ventricular (RV) lead exhibited high thresholds. It was also stated that "Looked like device wasn't pacing correctly on telemetry". Both the RV lead and the Implantable Pulse Generator (IPG) remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50192839 ↗InjuryAug 7, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced pain at the site of the implant, the Implantable Pulse Generator (IPG) also reached Elective Replacement Indicator (ERI). The IPG was explanted and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Pain
MAUDE report 50170061 ↗InjuryJul 29, 2026Appropriate Device Problem Term/Code Not Available
It was reported that Implantable Pulse Generator (IPG) was non-Prophylactically replaced. The IPG was explanted and replaced. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50077620 ↗MalfunctionJul 9, 2026Electromagnetic Interference
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED ELECTROMAGNETIC INTERFERENCE (EMI) AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE. THE RA LEAD AND IPG REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25763014 ↗MalfunctionJul 2, 2026Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. THE HEART RATE WAS PACING BELOW THE LOWER RATE LIMIT. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDER-SENSING ON STORED ELECTROGRAMS (EGM) DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENTS IN PROGRESS AND MISCLASSIFICATION OF EPISODE TYPES. THE IMPLANTABLE PULSE GENERATOR (IPG) AND RA LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 25687954 ↗InjuryMay 12, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT ATTENDED THE EMERGENCY ROOM AS THEY EXPERIENCED SYNCOPE. UPON DEVICE ASSESSMENT IT WAS NOTED THAT T HE IMPLANTABLE PULSE GENERATOR (IPG) HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE RIGHT ATRIAL (RA) LEAD WAS ALSO UNDER SENSING, DURING ATRIAL TACHYCARDIA (AT)/AF STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE IPG AND RA LEAD BOTH REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 25148081 ↗MalfunctionApr 28, 2026Device Sensing Problem
IT WAS REPORTED THAT DURING INTERROGATION MIRROR IMAGE ARTIFACT WAS NOTED IN A FEW EPISODES AND INTERMITTENTLY. SENSING ISSUES, SPECIFICALLY NOISE, WERE IDENTIFIED ON BOTH THE RIGHT ATRIAL (RA) LEAD AND THE RIGHT VENTRICULAR (RV) LEAD. ADDITIONALLY, A NON-PHYSIOLOGIC SENSING WAS IDENTIFIED ON THE RV AND RA LEAD. DEVICE INTERROGATION WAS PERFORMED, AND REPROGRAMMING WAS CONDUCTED AS AN INTERVENTION. THE RA AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A DETECTION/SENSING ISSUE. THE RA LEAD AND THE RV LEAD EXHIBITED INSULATION BREACH. THE IPG SYSTEM REMAINS IN USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25014825 ↗MalfunctionApr 24, 2026False Alarm
IT WAS REPORTED THAT THERE WAS A FALSE LEAD IMPEDANCE ON THE IMPLANTABLE PULSE GENERATOR (IPG). IT WAS ALSO NOTED THERE WAS AN OUT OF RANGE (OOR) LEAD IMPEDANCE ON THE RIGHT ATRIAL (RA) LEAD. IT WAS ALSO NOTED THERE WAS A POLARITY SWITCH AND THERE WERE HIGH SHORT V-V INTERVALS ON THE SENSING INTEGRITY COUNTER (SIC) AND NOISE. THE IPG SYSTEMA REMAINS USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24979486 ↗InjuryApr 24, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION AND EXPERIENCED PURULENT DISCHARGE. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED WITH NO REPLACEMENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 24977274 ↗MalfunctionApr 22, 2026Unable to Obtain Readings; Communication or Transmission Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) LEAD IMPEDANCE TREND DATA WAS NOT AVAILABLE. IT WAS ALSO REPORTED AN IN CLINIC MANUAL LEAD IMPEDANCE MEASUREMENT WAS ATTEMPTED BUT WAS UNSUCCESSFUL. THE PACING SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24953463 ↗MalfunctionApr 17, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THERE WAS TELEMETRY ISSUE AND THE RIGHT ATRIAL (RA) LEAD HAD A LOSS OF CAPTURE. THE RA LEAD AND IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24912043 ↗InjuryApr 15, 2026Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIZZINESS WHEN WALKING, UPON INTERROGATION, IT WAS NOTED THAT THE RIGHT ATRIAL (RA) LEAD ATRIAL LEAD HAD A POLARITY SWITCH DUE TO LOW IMPEDANCE MEASUREMENTS, NOISE AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED SHORT V-V, AND NOISE. THE DEVICE WAS INITIALLY REPROGRAMMED TO MINIMIZE NOISE EPISODES CAUSING PACING INHIBITION. THE LEADS WERE THEN CAPPED AND THE IMPLANTABLE PULSE GENERATOR (IPG) EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness
MAUDE report 24884235 ↗MalfunctionApr 14, 2026Failure to Interrogate
IT WAS REPORTED THAT A PATIENT WITH AN IMPLANTABLE PULSE GENERATOR (IPG) EXPERIENCED A FAILED TELEMETRY TRANSMISSION USING THE MYCARELINK MONITOR WHEN ATTEMPTING TO INITIATE A TRANSMISSION BY PLACING THE MONITOR OVER THE DEVICE. THE PATIENT STATED THE TELEMETRY PROCESS DID NOT START, THE MONITOR REPEATEDLY RETURNED TO THE START SCREEN, AND NO PROGRESS BAR APPEARED DURING MULTIPLE ATTEMPTS. THE PATIENT ADJUSTED THE MONITOR POSITION SEVERAL TIMES, CORRECT MONITOR PLACEMENT WAS VERIFIED, AND THE PATIENT CONFIRMED THERE WAS NO NEARBY EQUIPMENT THAT COULD INTERFERE WITH THE SIGNAL. NO SUCCESSFUL TELEMETRY TRANSMISSION OCCURRED, THE PATIENT WAS NOT CONNECTED TO THE DEVICE, AND THE PATIENT WAS REFERRED TO A DOCTOR¿S OFFICE/CLINIC FOR AN IN-PERSON DEVICE CHECK. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24869405 ↗MalfunctionApr 10, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THERE WAS A TELEMETRY ISSUE WITH THE IMPLANTABLE PULSE GENERATOR (IPG) WITH NO SUCCESSFUL TRANSMISSIONS.THE SMART MONITORING APPLICATION WAS UNABLE TO SEND A MANUAL TRANSMISSION, WITH THE PROGRESS BAR STOPPING DURING TELEMETRY AND MULTIPLE UNSUCCESSFUL DATA TRANSMISSION ATTEMPTS DESPITE TROUBLESHOOTING STEPS. THE PATIENT DID NOT HAVE A GOOGLE PIXEL OR ZTE SMARTPHONE, THE READER BATTERIES WERE REPLACED, A DRY WASHCLOTH WAS PLACED BETWEEN THE READER AND THE IMPLANT, AND THE PATIENT CONFIRMED THERE WERE NO INTERFERENCE DEVICES NEARBY. THE READER BATTERIES WERE REMOVED AND REINSERTED, BUT THE PROGRESS BAR CONTINUED TO STOP DURING TELEMETRY AND NO ERROR CODE APPEARED ON THE APPLICATION. THE PATIENT WAS REFERRED TO THEIR CLINIC AND ADVISED TO TAKE THE MONITOR TO THE APPOINTMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24838801 ↗MalfunctionApr 9, 2026Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA WITH A HEART RATE OF FORTY BEATS PER MINUTES WHICH IS BELOW THE LOWER RATE OF THE IMPLANTABLE PULSE GENERATOR (IPG). THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Arrhythmia; Bradycardia
MAUDE report 24830118 ↗MalfunctionApr 1, 2026Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THERE WAS NO PERFORMANCE ISSUE WITH THE IPG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24754114 ↗MalfunctionMar 31, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEY HAVE BEEN "FEELING SHITTY" AND THERE IS A " PROBLEM WITH THEIR PACEMAKER". THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 24742366 ↗InjuryMar 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY A PATIENT REPRESENTATIVE THAT THE PATIENT EXPERIENCED AN INFECTION ONE MONTH POST IMPLANT OF AN IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM. THE IPG SYSTEM WAS REMOVED AND LATER REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 24700466 ↗InjuryMar 23, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED EARLY BATTERY DEPLETION. THE IPG WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660368 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P980035 on Jan 29, 1999. Submitted by Medtronic, Inc. as “Medtronic Kappa 700/600 Series Pulse Generators And Model 9953 Software”. The FDA review took 157 days. The FDA has since approved 824 supplements (changes to the device or its labeling), most recently on Nov 11, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
The FDA lists this type of device as life-sustaining or life-supporting.
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-0806-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (e) Mo
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0811-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
…and 1 more.
Common questions about the Advisa DR MRI™ SureScan® A2DR01
Has the Advisa DR MRI™ SureScan® A2DR01 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00613994713650), based on the FDA recall file dated Nov 27, 2026. However, 16 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Advisa DR MRI™ SureScan® A2DR01?
The FDA's MAUDE database lists 1,757 adverse event reports for this device, received between Dec 9, 2013 and Sep 29, 2026. In those reports, the most common specific device problems are “Pacing Problem” (326), “Premature Discharge of Battery” (146) and “Unable to Obtain Readings” (75); the most common patient problems are “Unspecified Infection” (308), “Bradycardia” (154) and “Syncope/Fainting” (64). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Advisa DR MRI™ SureScan® A2DR01?
41 reports are classified as deaths, 949 as injuries and 767 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Advisa DR MRI™ SureScan® A2DR01 FDA approved?
It was approved through premarket approval (PMA) P980035 on Jan 29, 1999.
Is the Advisa DR MRI™ SureScan® A2DR01 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Advisa DR MRI™ SureScan® A2DR01, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Advisa DR MRI™ SureScan®, model A2DR01). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall262 deaths16,938 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths8,017 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,625 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,297 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,966 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,511 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,911 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,978 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page