Skip to content

Paradigm® REAL-Time Revel™ MMT-723NAS (UDI 00613994904355)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-723NAS PRDGM INS V2.4 SK EN PR · Class 3

All models of Paradigm REAL-Time Revel (180)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic MiniMed Paradigm REAL-Time Revel insulin pump delivers insulin. Its 104 reports, 16 of them injuries, often describe alarms and error messages, the pump failing to cycle or program properly, and battery and cracked-casing problems, with some reports describing high or low blood sugar. There were no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
104
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 104 adverse event reports that list this exact device (DI 00613994904355), received between Jun 2, 2017 and Jul 3, 2020 — classified as 16 injuries, 88 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (19 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message19
  • Failure to Cycle19
  • Device Difficult to Program or Calibrate17
  • Mechanical Problem16
  • Detachment of Device or Device Component11
  • Obstruction of Flow10
  • Patient Device Interaction Problem8
  • Power Problem8

Patient problems

  • No Consequences Or Impact To Patient79
  • Hyperglycemia21
  • Hypoglycemia7
  • No Known Impact Or Consequence To Patient1
  • Reaction, Injection Site1

Recent adverse event reports

The 25 most recent of 104 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2020Mechanical Problem; Power Problem; Mechanical Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND. CUSTOMER STATED THAT DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER DID NOT REPORT MOTOR POSITION ENCODER ERROR ALARM. CUSTOMER STATED THAT INSULIN PUMP HAD FAILED BATTERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10229666 ↗
  • MalfunctionApr 27, 2020Obstruction of Flow; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED CODE AND NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10002178 ↗
  • MalfunctionApr 14, 2020Obstruction of Flow; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD RAINBOW EFFECT PREVENTING THE CUSTOMER FROM SEEING. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9956578 ↗
  • MalfunctionMar 20, 2020Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM. THE CUSTOMER STATED THE DEVICE WAS SQUIRTING INSULIN OUT DURING MANUAL PRIME PROCESS. THE DEVICE WAS NOT BUMPED OR DROPPED PRIOR TO THE ALARM AND THE DRIVE SUPPORT CAP APPEARS TO BE STICKING OUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9858729 ↗
  • MalfunctionMar 4, 2020Premature Discharge of Battery; Battery Problem; Power Problem; Premature Discharge of Battery; Battery Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A UNEXPECTED RESTART AND A COLD RESET WAS PERFORMED ALARM. CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP RECEIVED POOR BATTERY LIFE, REJECTED NEW BATTERY AND PREMATURE BATTERY DISCHARGE. CUSTOMER STATED THAT TROUBLESHOOTING WAS NOT DONE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9784490 ↗
  • MalfunctionFeb 25, 2020Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM DUE TO THE FORCE SENSOR DID NOT DETECT THE RESERVOIR DURING SEATING. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9749551 ↗
  • MalfunctionJan 10, 2020Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER STATED THEY ARE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9573707 ↗
  • MalfunctionDec 25, 2019Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem

    CUSTOMER REPORTED VIA PHONE CALL THAT THE BACK OF THE INSULIN PUMP HAD CRACK AT BATTERY AND RESERVOIR COMPARTMENT AND INSULIN PUMP BUTTONS WERE DIFFICULT TO BE PRESSED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF INCIDENT. CUSTOMER STATED THAT THEIR WAS NO PHYSICAL DAMAGE TO RESERVOIR LIP RING. CUSTOMER STATED THAT TIME WAS ADVANCING. ADVISED THE INSULIN PUMP WILL NEED TO BE REPLACED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9517898 ↗
  • MalfunctionDec 3, 2019Device Difficult to Program or Calibrate; Obstruction of Flow; Device Difficult to Program or Calibrate; Obstruction of Flow

    CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD NO DELIVERY AND BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER ALSO STATED THAT THE INSULIN PUMP HAD BUTTON ERROR EVEN THOUGH THE BUTTONS WERE WORKING. TROUBLE SHOOTING WAS PERFORMED FOR NO DELIVERY ALARM AND EVENT WAS RESOLVED BY CHANGING COMPLETE SET. IT WAS ALSO REPORTED THAT THE TIME ON INSULIN PUMP WAS ADVANCES. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9400016 ↗
  • MalfunctionNov 28, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM DURING TOURING. CUSTOMER¿S BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 397 MG/DL. CUSTOMER REPORTED THAT THE INSULIN WAS SQUIRTING OUT DURING THE MANUAL PRIME PROCESS. CUSTOMER STATED INSULIN CONTINUES TO DRIP AFTER MOTOR STOPS DURING THE MANUAL PRIME PROCESS. CUSTOMER STATED THE DRIVE SUPPORT CAP WAS STICKING OUT. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9389884 ↗
  • MalfunctionNov 26, 2019Premature Discharge of Battery; Loss of Data; Premature Discharge of Battery; Loss of Data

    CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD PUMP ALARM. CUSTOMER'S BLOOD GLUCOSE VALUE WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER REPORTS THEY WERE ABLE TO CLEAR THE ALARM. CUSTOMER ABLE TO COMPLETE REWIND. CUSTOMER STATES THE ALARM OCCURRED SHORTLY AFTER INSERTING A NEW BATTERY. THE CURRENT BATTERY FROM THE PUMP AND INSERT A NEW BATTER. CUSTOMER STATED THE INSULIN PUMP WAS NOT STORED OR NOT IN USE FOR AN EXTENDED PERIOD. THE DEVICE WILL BE RETURN FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9373880 ↗
  • MalfunctionNov 22, 2019Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Power Problem; Patient Device Interaction Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Power Problem; Patient Device Interaction Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP RECEIVED MOTOR ERROR. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER REPORTED THAT THEY RECEIVED A MOTOR ERROR ALARM AND ALSO HAD REJECTED BATTERIES. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9362168 ↗
  • MalfunctionNov 6, 2019Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA THAT THE INSULIN PUMP ALARMED FOR COMPROMISED FORCE SENSOR SYSTEM AND INSULIN PUMP WAS STUCK IN PRIMING. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 150 MG/DL. INSULIN WAS SQUIRTING OUT DURING MANUAL PRIME. CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS NORMAL. FORCE SENSOR DID NOT DETECT THE RESERVOIR DURING SEATING. CUSTOMER WAS NOT ABLE TO REWIND THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEED TO BE REPLACED AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9288910 ↗
  • MalfunctionNov 1, 2019Failure to Cycle; Obstruction of Flow; Failure to Cycle; Obstruction of Flow

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP SHOT INSULIN DURING PRIMING. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER ALSO REPORTED NO DELIVERY ALARMS RANDOMLY. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9265106 ↗
  • MalfunctionOct 31, 2019Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEIR INSULIN PUMP HAD A MOTOR ERROR ALARMS. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9259559 ↗
  • MalfunctionOct 30, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP RECEIVED COMPROMISED FORCE SENSOR SYSTEM ALARM DURING PRIMING OF TUBING. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 175 MG/DL AT THE TIME OF THE INCIDENT. CUSTOMER STATED THAT THE INSULIN PUMP NEITHER DROPPED NOR BUMPED PRIOR TO ALARM. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO THE BACKUP PLAN. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9254323 ↗
  • MalfunctionOct 23, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE ALL BUTTONS WAS STOP WORKING AND HAD BUTTON NOT RESPONDING ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER CONFIRMED THAT THE TIME CLOCK DID ADVANCED. THE CUSTOMER WAS ADVISED TO DISCONTINUED THE INSULIN PUMP AND REVERT TO BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9224406 ↗
  • MalfunctionSep 25, 2019Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEIR INSULIN PUMP BUTTONS HAD TO PRESSED MORE THAN ONCE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. INSULIN PUMP HAD SHORT BATTERY LIFE, UNEXPLAINED NO DELIVERY ALARMS AND HAD TO REWIND MORE THAN ONCE PER PRIME. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9114562 ↗
  • MalfunctionSep 24, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    THE CUSTOMER'S WIFE REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD BUTTON ERROR MESSAGE. CUSTOMER'S BLOOD GLUCOSE VALUE WAS 80 MG/DL. CUSTOMER'S WIFE STATED THAT INSULIN PUMP WAS STARTED BEEPING WHILE HE WAS ASLEEP. CUSTOMER STATED THAT WHILE ON THE CALL THEY TOOK THE BATTERY OUT NOW AND PUT ANOTHER NEW BATTERY. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE DEVICE WILL BE RETURN FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9108966 ↗
  • MalfunctionSep 11, 2019Device Difficult to Program or Calibrate; Obstruction of Flow; Device Difficult to Program or Calibrate; Obstruction of Flow

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE VALUE WAS UNKNOWN. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER REPORTED THAT THE TIME WAS ADVANCING. THE CUSTOMER WAS ADVISED TO REPLACE AND TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8987899 ↗
  • InjurySep 10, 2019Mechanical Problem; Obstruction of Flow; Detachment of Device or Device Component; Mechanical Problem; Obstruction of Flow; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED THE HIGH BLOOD GLUCOSE AND THE INSULIN PUMP HAD MOTOR ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS UP INTO THE 400 MG/DL. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS FLUSH. THE TROUBLESHOOTING WAS PERFORMED FOR THE MOTOR ERROR AND NOT MENTIONED FOR THE NO DELIVERY ALARMS. THE CUSTOMER STATED THAT HE WAS IN THE HOSPITAL DUE TO THE FLU, UNRELATED TO THE INSULIN PUMP. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP, REVERT TO A BACKUP PLAN AND THE INSULIN PUMP WILL NEED TO BE REPLACED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 8979549 ↗
  • MalfunctionAug 14, 2019Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 384 MG/DL. CUSTOMER REPORTED THAT THEY WERE ABLE TO CLEAR THE ALARM AND WAS ABLE TO REWIND THE INSULIN PUMP. THE CUSTOMER WAS ADVISED TO DISCONTINUED THE INSULIN PUMP AND REVERT TO BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8891536 ↗
  • MalfunctionAug 6, 2019Priming Problem; Priming Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT IN THE PROCESS OF CHANGING THE RESERVOIR OF THE INSULIN PUMP BUT WAS STUCK IN MANUAL PRIME. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER STATES THEY ARE STUCK IN REWIND COMPLETE OR BOLUS DELIVERY LOOP. CUSTOMER STATES THEY DID NOT RECEIVE 2ND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. THE CUSTOMER WAS ADVISED TO REVERT TO BACK UP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8862963 ↗
  • MalfunctionJul 23, 2019Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A COMPROMISED FORCED SENSOR SYSTEM ERROR. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 129 MG/DL. THE CUSTOMER DECLINED TROUBLESHOOTING. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8815447 ↗
  • MalfunctionJun 20, 2019Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. CUSTOMER STATED THEY WERE ABLE TO REWIND THE PUMP. FILL CANNULA WAS RECORDED IN DAILY HISTORY. CUSTOMER WAS ADVISED THE INSULIN PUMP WILL NEED TO BE REPLACED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8715419 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page