Dimension® CHEM I CAL 10716280Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Chemistry I Calibrator, 2x3x2mL · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414948492), received between Sep 8, 2017 and Jun 28, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Low Test Results3
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2019Low Test Results; Low Test Results
DISCORDANT FALSELY DEPRESSED CREATININE (CRE2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON A DIMENSION EXL 200 INSTRUMENT. THE DISCORDANT RESULTS WERE REPORTED TO THE PHYSICIAN(S) AND QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON THE SAME INSTRUMENT AND HIGHER RESULTS WERE OBTAINED. CORRECTED REPORTS WERE ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY DEPRESSED CREATININE (CRE2) RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8745043 ↗MalfunctionJun 28, 2019Low Test Results; Low Test Results
DISCORDANT FALSELY DEPRESSED CREATININE (CRE2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON A DIMENSION EXL WITH LM INSTRUMENT. THE DISCORDANT RESULTS WERE REPORTED TO THE PHYSICIAN(S) AND QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON AN ALTERNATE INSTRUMENT AND HIGHER RESULTS WERE OBTAINED. CORRECTED REPORTS WERE ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY DEPRESSED CREATININE (CRE2) RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8745035 ↗MalfunctionSep 8, 2017Low Test Results; Low Test Results
DISCORDANT FALSELY LOW GLUCOSE (GLU) RESULTS WERE OBTAINED ON THE DIMENSION XPAND PLUS. BIASED LOW RESULTS WERE OBTAINED WHILE THE CUSTOMER ACCEPTED A BAD CALIBRATION DUE TO UNDER-RECONSTITUTION OF THE CALIBRATOR. ONE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN FOR PATIENT SAMPLE IDENTIFICATION (B)(6) AND NOT QUESTIONED. NO OTHER PATIENT RESULTS WERE REPORTED AND A CORRECTED REPORT WAS ISSUED FOR THE INITIAL RESULT REPORTED. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW GLUCOSE RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6851958 ↗
Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,197 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall911 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death694 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death646 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE685 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page