Viva™ S CRT-D DTBB1D4 (UDI 00643169007178)Recalls & reported problems
Medtronic, Inc. · CRT-D DTBB1D4 VIVA S US IS1/DF4 · Class 3 · Implant
All models of Viva S CRT-D (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Viva S CRT-D is an implanted defibrillator that also coordinates the heart's rhythm. The 64 reports mostly describe batteries draining sooner than expected, along with sensing problems and shocks or pacing delivered when not needed; 51 injuries and no deaths were reported. Two recalls name this device: an open one from February 2021 for shortened battery life, and a closed one from December 2017 about possible prevention of therapy.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 51
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 00643169007178), received between Feb 10, 2015 and May 17, 2024 — classified as 51 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (23 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 3, 2021Class IOpen, Classified
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Z-1306-2021 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 19, 2017Class ITerminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Z-0584-2018 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Premature Discharge of Battery23
- Adverse Event Without Identified Device or Use Problem13
- Device Sensing Problem7
- Battery Problem6
- Electrical /Electronic Property Problem6
- Appropriate Device Problem Term/Code Not Available4
- Inappropriate/Inadequate Shock/Stimulation4
- Invalid Sensing3
Patient problems
- No Known Impact Or Consequence To Patient29
- No Clinical Signs, Symptoms or Conditions7
- Unspecified Infection7
- No Patient Involvement5
- Shock from Patient Lead(s)4
- Dizziness2
- Endocarditis2
- Erosion2
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
InjuryMay 17, 2024Battery Problem; Appropriate Device Problem Term/Code Not Available; Battery Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAS NOT WORKED IN TWO YEARS AND THE BATTERY IS DEAD. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19342093 ↗InjuryJan 30, 2024Pacing Problem; Pacing Problem
IT WAS FURTHER REPORTED THAT BOTH THE ATRIAL AND RIGHT VENTRICULAR (RV) LEADS HAD VEGETATION. THE DEVICE, THE ATRIAL, RV AND LEFT VENTRICULAR (LV) LEADS WERE EXPLANTED AND REPLACED ONE WEEK LATER DUE TO THE ENDOCARDITIS INFECTION.
Patient problems: Bradycardia; Endocarditis; Heart Failure/Congestive Heart Failure; Sepsis
MAUDE report 18607718 ↗InjuryApr 4, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE CRT-D WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13995149 ↗InjuryJan 12, 2022Battery Problem; Appropriate Device Problem Term/Code Not Available; Battery Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT WAS CONCERNED ABOUT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PROGRAMMING AFTER RECEIVED AN APPROPRIATE SHOCK. THE PATIENT NOTED THAT THEY HAD FELT DIFFERENT BEFORE THE SHOCK. THE CRT-D WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 13237146 ↗InjuryJan 5, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS ALSO REPORTED BY THE PATIENT THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DELIVERED HIGH VOLTAGE SHOCK THERAPY. IT WAS DISCLOSED BY THE CLINIC THAT SHOCK WAS INAPPROPRIATE THERAPY AS IT OCCURRED DURING EPISODE OF ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE. ADJUSTMENT WAS MADE TO MEDICATION TREATMENT. NO ACTION WAS TAKEN WITH THE DEVICE. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 13172233 ↗MalfunctionNov 12, 2021Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) MAY HAVE DELIVERED INAPPROPRIATE PACING. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12803797 ↗InjurySep 10, 2021Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED SHORTER THAN EXPECTED LONGEVITY ESTIMATE DUE TO A PROGRAMMER SOFTWARE ESTIMATOR ERROR. IT WAS ALSO REPORTED THE CRT-D EXHIBITED PREMATURE BATTERY DEPLETION. THE PROGRAMMER REMAINS IN USE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12453085 ↗InjuryJun 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) POCKET HAD ERODED OVER TIME WITH THE PATIENT'S WEIGHT LOSS BEING A CONTRIBUTING FACTOR. THE SYSTEM WAS REMOVED AND A NEW SYSTEM WAS IMPLANTED ON THE PATIENT'S OPPOSITE SIDE APPROXIMATELY TWO WEEKS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Pocket Erosion
MAUDE report 11957863 ↗InjuryMay 4, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS LATER REPORTED THAT THERE WERE NO KNOWN PRODUCT PERFORMANCE ISSUES THAT MAY HAVE CONTRIBUTED TO THE PATIENT PASSING OUT.
Patient problems: Syncope/Fainting
MAUDE report 11773016 ↗MalfunctionMay 4, 2021Battery Problem; Battery Problem
THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS REPLACED DUE TO REACHING RECOMMENDED REPLACEMENT TIME (RRT). THE CRT-D WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT-OF-SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OFTHIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11768821 ↗MalfunctionJan 28, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING A REPLACEMENT PROCEDURE THE SET SCREW FELL OUT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND THE LEAD COULD NOT BE REMOVED FROM THE HEADER. A SECOND SCREW WAS FOUND AND WHEN THE SECOND SCREW FELL OUT, THE LEAD WAS SUCCESSFULLY REMOVED. THE CRT-D WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11243207 ↗InjuryJan 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 11214705 ↗InjuryNov 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED POCKET EROSION WITH POSSIBLE INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND A NEW SYSTEM WAS IMPLANTED A FEW DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10885136 ↗InjuryOct 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN (B)(6) INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10619451 ↗InjuryAug 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RE-SYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO A SYSTEMIC INFECTION. A NEW CARDIAC RE-SYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM WAS IMPLANTED SIX DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE ORGANISM WAS IDENTIFIED TO BE METHICILLIN-SENSITIVE STAPHYLOCOCCUS AUREUS BACTEREMIA AND RIGHT VENTRICULAR (RV) LEAD VEGETATION WAS NOTED.
Patient problems: Unspecified Infection
MAUDE report 10420485 ↗InjuryMay 18, 2020Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED A RADIATION TREATMENT THAT FAR EXCEEDED THE RECOMMENDED DOSE OF RADIATION, CAUSING THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) TO FUNCTION INAPPROPRIATELY. ABNORMAL ELECTROGRAMS DEMONSTRATED NO SENSING OF RHYTHM. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10065982 ↗InjuryApr 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED BECAUSE THE SYSTEM WAS BLOCKING A VEIN WHICH THEN LEAD TO THE PATIENT'S ARM SWELLING. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Occlusion; Swelling
MAUDE report 9917822 ↗InjuryFeb 10, 2020Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED HIGH VOLTAGE THERAPY FOR SUSPECTED ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 9689065 ↗InjuryJan 28, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THERE WAS PREMATURE BATTERY DEPLETION ON THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9636477 ↗InjuryDec 3, 2019Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THERE WAS A LEAD INTEGRITY ALERT (LIA) ON THE RIGHT VENTRICULAR (RV) LEAD FOR HIGH RATE NON-SUSTAINED EPISODES AND ELEVATED SHORT VENTRICULAR INTERVALS. THERE WAS ALSO A NOISE WARNING, OVERSENSING, AND LOW PACING IMPEDANCE ON THE RV LEAD, WHICH WAS NOTED TO BE FRACTURED. THE PATIENT WAS INAPPROPRIATELY SHOCKED DUE TO THE RV LEAD ISSUE. THE THERAPY PORTION OF THE RV LEAD WAS PROGRAMMED OFF. IT WAS NOTED THAT THIS LEAD TO PREMATURE BATTERY DEPLETION ON THE DEVICE. THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9405025 ↗InjuryOct 16, 2019Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED RAPID BATTERY DEPLETION. THE CRT-D WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9200372 ↗InjuryOct 16, 2019Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH PACING IMPEDANCE VALUES AND OVERSENSING DUE TO FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED AND REPLACED DUE TO PREMATURE BATTERY DEPLETION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9199722 ↗InjurySep 27, 2019Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THERAPY WAS INAPPROPRIATELY WITHHELD FOR AN EPISODE OF VENTRICULAR TACHYCARDIA (VT). THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Code Available
MAUDE report 9129427 ↗InjuryJul 15, 2019Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) EARLY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8790014 ↗InjuryMay 23, 2019Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED ONE SHOCK FOR AN EPISODE IN THE VENTRICULAR FIBRILLATION (VF) ZONE THAT HAD AN ATRIAL ARRHYTHMIA IN-PROGRESS AT THE TIME OF THERAPY. THERE WAS SUSPECTED THERAPY FOR RAPID VENTRICULAR RATE DURING AN ATRIAL ARRHYTHMIA. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 8635002 ↗
Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices
- Gallant™ CDHFA500QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls58 deaths8,401 reports
- Quartet™ 1458Q-86St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)207 deaths6,631 reports
- QUADRA ASSURA MP™ CD3369-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)103 deaths6,084 reports
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- Quadra Assura™ CD3365-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)53 deaths2,056 reports
- Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls40 deaths1,639 reports
- Unify Assura™ CD3357-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)54 deaths1,784 reports
- Unify Assura™ CD3357-40CSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)56 deaths1,588 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page