Viva™ Quad S CRT-D DTBB1Q1 (UDI 00643169017993)Recalls & reported problems
Medtronic, Inc. · CRT-D DTBB1Q1 VIVA QUAD S US IS4/DF1 · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Viva Quad S CRT-D is an implanted defibrillator that delivers pacing and shock therapy for heart rhythm problems. The 43 reports mostly describe batteries running down sooner than expected, shocks or pacing delivered when not needed, and sensing or mechanical issues; some also mention infection at the implant site. There is 1 death and 26 injuries. Two recalls name this device: one open recall from February 2021 for batteries reaching end of service earlier than expected, and one closed recall from December 2017.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 1
- Injuries reported
- 26
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00643169017993), received between Dec 9, 2016 and Jun 20, 2025 — classified as 1 death, 26 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (8 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 3, 2021Open, Classified
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Z-1306-2021 · Root cause: Device Design · FDA record ↗
- Dec 19, 2017Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Z-0584-2018 · Root cause: Process design · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
- Premature Discharge of Battery8
- Inappropriate/Inadequate Shock/Stimulation6
- Device Sensing Problem5
- Mechanical Problem5
- Pacing Problem4
- Electrical /Electronic Property Problem3
- Electromagnetic Interference3
Patient problems
- No Known Impact Or Consequence To Patient15
- Unspecified Infection9
- No Clinical Signs, Symptoms or Conditions7
- Shock from Patient Lead(s)6
- No Consequences Or Impact To Patient2
- No Patient Involvement2
- Bradycardia1
- Death1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 20, 2025Mechanical Problem; Mechanical Problem
THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NORMAL ERI AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22301790 ↗InjuryJun 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. THE CRT-D WAS REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 17074795 ↗InjuryNov 10, 2022Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
IT WAS REPORTED THAT AN INTERNAL REVIEW OF DATA TRANSMITTED FROM THE DEVICE SHOWED THAT SHORT CIRCUIT PROTECTION OCCURRED RESULTING IN LESS THAN PROGRAMMED HIGH VOLTAGE ENERGY DELIVERY WITH LOW IMPEDANCE DURING DEFIBRILLATION THRESHOLD TESTING, CONDUCTED AS PART OF THE IMPLANT PROCEDURE. DURING AN EPISODE OF VENTRICULAR FIBRILLATION (VF) EXTERNAL RESCUE WAS REQUIRED TO TERMINATE THE EPISODE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Ventricular Fibrillation
MAUDE report 15771839 ↗MalfunctionOct 11, 2022Communication or Transmission Problem; Communication or Transmission Problem
ADDITIONAL INFORMATION RECEIVED INDICATED THAT NO TELEMETRY-RELATED OR OTHER PRODUCT PERFORMANCE ISSUE HAD BEEN IDENTIFIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15574329 ↗InjuryAug 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO A POCKET INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 15298003 ↗InjuryJul 13, 2022Failure to Interrogate; Pacing Problem; Battery Problem; Failure to Interrogate; Pacing Problem; Battery Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA AND ATTEMPTS TO INTERROGATE THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WERE UNSUCCESSFUL. A MAGNET WAS PLACED OVER THE DEVICE AND NO AUDIBLE TONE WAS EMITTED AND PACING WAS NOT VISIBLE ON ELECTROCARDIOGRAM (ECG). IT WAS SUSPECTED THAT THE DEVICE WAS AT END OF SERVICE (EOS) OR WITHOUT BATTERY POWER. THE CRT-D WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15005257 ↗MalfunctionJun 23, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED UNEXPECTED BATTERY DEPLETION. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14789091 ↗MalfunctionJun 8, 2022Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED ANTI-TACHYCARDIA PACING THERAPY FOR ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) WITH RAPID VENTRICULAR RESPONSE (RVR). IN ADDITION, R WAVE MEASUREMENTS DROPPED SINCE THE AT/AF EPISODES. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REACHED RECOMMENDED REPLACEMENT TIME (RRT) AND HAS SINCE BEEN EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14632953 ↗MalfunctionJan 10, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN REPLACEMENT PROCEDURE FOR NORMAL BATTERY DEPLETION THERE WAS DIFFICULTY RETRACTING THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SETSCREW FOR THE LEFT VENTRICULAR (LV) LEAD PORT. MULTIPLE WRENCHES WERE USED BUT THE USER STRUGGLED TO ENGAGE THE DRIVE OF THE SETSCREW. THE LEAD WAS ABLE TO BE REMOVED FROM THE DEVICE HEADER UNDAMAGED AND OPERATIONAL. THE CRT-D WAS EXPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13209319 ↗InjuryAug 24, 2021Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND NOISE. THE CAUSE OF THE NOISE COULD NOT BE DETERMINED. IT WAS ALSO REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE LEAD WAS CAPPED AND REPLACED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12359440 ↗InjuryJun 22, 2021Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THERE WAS UNDERSENSING ON THE RIGHT ATRIAL (RA) LEAD DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF). IT WAS ALSO NOTED THAT THE PATIENT MAY HAVE RECEIVED AN INAPPROPRIATE SHOCK FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE DEVICE CLASSIFIED AS VENTRICULAR TACHYCARDIA (VT)/ VENTRICULAR FIBRILLATION (VF). THE LEAD AND DEVICE REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 12041907 ↗InjuryJun 4, 2021Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING HEART RATES OF APPROXIMATELY FORTY BEATS PER MINUTE, WHICH WAS BELOW THE LOWER PROGRAMMED RATE OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS EXHIBITING POSSIBLE INTERMITTENT NON-CAPTURE. THE CRT-D AND LV LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 11938796 ↗InjuryJun 1, 2021Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) MAY HAVE DELIVERED SHOCKS FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS VENTRICULAR FIBRILLATION (VF) AND FAST VENTRICULAR TACHYCARDIA (FVT). THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 11913579 ↗InjuryOct 13, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) BATTERY WAS DEPLETING FASTER THAN EXPECTED, POTENTIALLY DUE TO SENSING OF FAST ATRIAL FIBRILLATION (AF). THE DEVICE WAS REPROGRAMMED TO DISCONTINUE AF SENSING AND OUTPUTS WERE ADJUSTED DOWNWARD TO EXTEND LONGEVITY. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10669547 ↗InjurySep 28, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THERE WAS POSSIBLE SHORTER THAN EXPECTED BATTERY LIFE ON THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10592793 ↗InjuryJul 10, 2020Signal Artifact/Noise; Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT WAS IN A SWIMMING POOL AND RECEIVED A SUSPECTED INAPPROPRIATE SHOCK. IT WAS NOTED A LEAD INTEGRITY ALERT (LIA) TRIGGERED AND ELECTROMAGNETIC INTERFERENCE/NOISE WAS OBSERVED ON ALL ELECTROGRAMS (EGM) OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 10258068 ↗InjuryMay 12, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING A ROUTINE DEVICE REPLACEMENT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) AND LEADS WERE EXPLANTED. THE SYSTEM WAS REPLACED TWO WEEKS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10047608 ↗MalfunctionApr 15, 2020Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) MAY HAVE BEEN OVERSENSING ELECTROMAGNETIC INTERFERENCE. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9962305 ↗InjuryMar 30, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) BATTERY LIFE WAS DECLINING EARLIER THAN EXPECTED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9901014 ↗InjuryMar 4, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9787579 ↗InjuryDec 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT VEGETATION WAS PRESENT ON ONE OF THE IMPLANTED LEADS AND THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 9519818 ↗InjuryOct 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 9203790 ↗DeathApr 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT¿S SPOUSE CALLED TO REPORT THE PATIENT ¿HAS A STAPH INFECTION ON THE LEADS". CALLER ALSO REPORTED THAT THE PATIENT ALERT TONE WAS GOING OFF MULTIPLE TIMES A DAY. IT WAS FURTHER REPORTED THE PATIENT SUBSEQUENTLY DIED. FOLLOW UP INDICATED THE PATIENT WAS PLACED IN HOSPICE AND HAD THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) TURNED OFF. THERE WAS NO OTHER INFORMATION WAS MADE AVAILABLE.
Patient problems: Death; Unspecified Infection
MAUDE report 8519153 ↗InjuryFeb 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN INFECTION. THE DEVICE AND LEADS WERE EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 8362720 ↗InjuryJan 28, 2019Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE DEVICE HAD REACHED BATTERY DEPLETION PREMATURELY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8282351 ↗
Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices
- Gallant™ CDHFA500QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls58 deaths8,401 reports
- Quartet™ 1458Q-86St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)207 deaths6,631 reports
- QUADRA ASSURA MP™ CD3369-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)103 deaths6,084 reports
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- Quadra Assura™ CD3365-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)53 deaths2,056 reports
- Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls40 deaths1,639 reports
- Unify Assura™ CD3357-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)54 deaths1,784 reports
- Unify Assura™ CD3357-40CSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)56 deaths1,588 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page