Paradigm® REAL-Time Revel™ MMT-523LNAS (UDI 00643169151314)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-523LNAS PRDGM INS V2.5 SK EN LN · Class 3
All models of Paradigm REAL-Time Revel (180)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic MiniMed Paradigm REAL-Time Revel is an insulin pump. Of its 274 reports, 85 are injury reports and 189 are malfunction reports, with no deaths. Many describe alarms, incorrect messages and unresponsive buttons, and some mention cracked or broken parts. High blood sugar is listed in 84 reports. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 274
- Deaths reported
- 0
- Injuries reported
- 85
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 274 adverse event reports that list this exact device (DI 00643169151314), received between May 12, 2017 and Jan 25, 2021 — classified as 85 injuries, 189 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (53 reports).
14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message53
- Device Difficult to Program or Calibrate42
- Patient Device Interaction Problem36
- Mechanical Problem27
- Adverse Event Without Identified Device or Use Problem21
- Power Problem20
- Break16
- No Apparent Adverse Event16
Patient problems
- No Consequences Or Impact To Patient173
- Hyperglycemia84
- Hypoglycemia24
- Elevated Ketones/ Diabetic Ketoacidosis7
- Vomiting3
- Nausea2
- Blood Loss1
- Dyspnea1
Recent adverse event reports
The 25 most recent of 274 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 25, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP'S RESERVOIR COMPARTMENT WAS CHIPPED AND THERE WAS MISSING PIECES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11222102 ↗InjuryAug 14, 2020No Apparent Adverse Event; Patient Device Interaction Problem; No Apparent Adverse Event; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY HAD A HIGH BLOOD GLUCOSE VALUE OF 400 MG/DL. CUSTOMER STATED THAT THE BATTERY COMPARTMENT WAS DAMAGE AND CRACK BATTERY. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10405940 ↗MalfunctionJul 31, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR ALARM. THE CUSTOMER STATED THAT TIME WAS ADVANCING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10348632 ↗MalfunctionJul 28, 2020Premature Discharge of Battery; Loss of Data; Premature Discharge of Battery; Loss of Data
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MULTIPLE PUMP ERROR ALARMS. CUSTOMER WAS ABLE TO CLEAR THE ALARM AND WAS NOT ABLE TO COMPLETE REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10331074 ↗MalfunctionJul 8, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE TIME CLOCK ADVANCES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10243784 ↗MalfunctionJun 18, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Battery Problem; Power Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Battery Problem; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT CUSTOMER HAD TO HOLD ACT BUTTON DOWN TO RESPOND. THE CUSTOMER ALSO STATED THAT THE INSULIN PUMP HAD BATTERY LIFE WAS SIX DAYS OR LESS, REJECTED NEW BATTERIES AND REWINDS SLOWER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10166367 ↗InjuryMay 29, 2020Mechanical Problem; Mechanical Problem
CUSTOMER'S FAMILY REPORTED VIA PHONE CALL THAT THEY HAD EXPERIENCED HYPERGLYCEMIA WITH BLOOD GLUCOSE LEVEL OF 456 MG/DL. CUSTOMER DECLINED TROUBLESHOOTING. CUSTOMER STATED THAT INSULIN PUMP HAD MOTOR ERROR. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10100189 ↗InjuryMay 19, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 514 MG/DL. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10073491 ↗InjuryMay 19, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 563 MG/DL. CUSTOMER REPORTED NO DELIVERY ALARM. CUSTOMER DECLINED TO TROUBLESHOOT FOR HIGH BLOOD GLUCOSE LEVEL. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10069646 ↗MalfunctionMay 15, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BOTTOM OF THE INSULIN PUMP WAS COMING OFF. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10060554 ↗MalfunctionMay 11, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTON ISSUE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10042549 ↗MalfunctionMay 7, 2020Premature Discharge of Battery; Loss of Data; Premature Discharge of Battery; Loss of Data
CUSTOMER CALLED AND REPORTED THAT THEIR INSULIN PUMP ALARMED WITH UNEXPECTED RESTART. A COLD RESET WAS PERFORMED AND EXTERNAL RAM CRC ERROR. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER WAS ABLE TO COMPLETE REWIND FOR UNEXPECTED RESTART. IT WAS CONFIRMED CUSTOMER WAS GETTING ALARMS WHEN CHANGING THE BATTERIES. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10031803 ↗MalfunctionMay 6, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS OF INSULIN PUMP WERE NOT RESPONDING AND TIME CLOCK DID NOT ADVANCES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10028821 ↗MalfunctionApr 21, 2020Display or Visual Feedback Problem; Mechanical Problem; Power Problem; Patient Device Interaction Problem; Display or Visual Feedback Problem; Mechanical Problem; Power Problem; Patient Device Interaction Problem
THE CUSTOMER REPORTED REWINDING ANOMALY. CUSTOMER BLOOD GLUCOSE READING WAS UNKNOWN. TROUBLESHOOTING WAS NOT PERFORMED. INSULIN PUMP WILL NOT BE RETURNING FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9984096 ↗MalfunctionMar 25, 2020Device Difficult to Program or Calibrate; Power Problem; Device Difficult to Program or Calibrate; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ANOMALY. INSULIN PUMP HAD BATTERY OUT LIMIT ALARM AND ALSO REPORTED THAT ACT BUTTON WAS NOT WORKING. TIME WAS ADVANCING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9876486 ↗InjuryMar 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 500 MG/DL. CUSTOMER STATED THAT SELF TEST WAS PASSED. THE CUSTOMER DECLINED TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 9864671 ↗MalfunctionMar 20, 2020Mechanical Problem; Patient Device Interaction Problem; Mechanical Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THERE WERE CRACKS WHERE THEY PUT THE RESERVOIR AND ON THE RESERVOIR CHAMBER, PIECE OF RUBBER CAME OFF OF RUBBER RING AND A LITTLE SLOWER TO REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9858456 ↗MalfunctionMar 19, 2020Priming Problem; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS STUCK ON PRIMING. CUSTOMER STATED THAT THEY WERE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. CUSTOMER DID NOT RECEIVE SECOND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9851965 ↗MalfunctionMar 16, 2020Material Integrity Problem; Patient Device Interaction Problem; Material Integrity Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT BATTERY CAP POPPED OFF AND IT DIDN'T STAY IN ANY LONGER. THE CUSTOMER STATED THAT THE PIECE WAS MISSING AND THE TOP OF THE BATTERY COMPARTMENT POPPED OFF. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9834398 ↗MalfunctionMar 16, 2020Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM ERROR. CUSTOMER STATED THAT THE INSULIN WAS SQUIRTING OUT DURING THE MANUAL PRIME PROCESS. CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9833808 ↗MalfunctionMar 13, 2020Detachment of Device or Device Component; Patient Device Interaction Problem; Detachment of Device or Device Component; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. CUSTOMER STATED THE INSULIN PUMP WAS DAMAGED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9832077 ↗MalfunctionFeb 10, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR. CUSTOMER CONFIRMED THAT THE TIME ADVANCED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9685924 ↗MalfunctionJan 30, 2020Display or Visual Feedback Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Display or Visual Feedback Problem; Mechanical Problem; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTONS WERE LESS RESPONSIVE AND BUTTONS WERE HARDER TO PRESS AND SOMETIMES HAD TO PRESS BUTTONS TWICE. THE CUSTOMER ALSO STATED THAT THE INSULIN PUMP WAS SLOWER AND LOUDER DURING REWIND AND DISPLAY LIGHT WAS DIMMER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9646533 ↗MalfunctionJan 19, 2020Material Integrity Problem; Patient Device Interaction Problem; Material Integrity Problem; Patient Device Interaction Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THERE WAS CRACK ON THE INSULIN PUMP¿S SCREEN AND IT WAS REALLY DIFFICULT TO READ. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER ALSO STATED THAT THE BATTERY CAP WAS STRIPPED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9605015 ↗MalfunctionJan 6, 2020Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem; Priming Problem; Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem; Priming Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WAS STUCK IN REWIND. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER STATED THAT THEY DID NOT RECEIVE 2ND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. CUSTOMER STATED THAT THE INSULIN PUMP HAD NO DELIVERY ALARM. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9550352 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.
- Z-1749-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Insulin Pump (MMT-522, MMT-722)
- Z-1750-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1256-2025 · Jan 31, 2025 · Open, ClassifiedParadigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
- Z-1589-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522K
- Z-1590-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722K
- Z-1591-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523
- Z-1593-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723K
- Z-1587-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522
- Z-1594-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723
- Z-1588-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722
- Z-1592-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523K
- Z-2442-2015 · Jul 22, 2015 · TerminatedGuardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
- Z-2179-2014 · Jul 21, 2014 · TerminatedMedtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Sensor, glucose, invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page